Skip to content

PRECICE® Intramedullary Limb Lengthening System LSC5-075Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · 5.0 mm Locking Screw 75 mm Length · Class 2

All models of PRECICE Intramedullary Limb Lengthening System (551)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
4

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00812258021627), received between Dec 2, 2022 and Aug 3, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Oct 29, 2021Class IIOpen, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Class IIOpen, Classified

    Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.

    Z-1349-2021 · Root cause: Device Design · Recall details · FDA record ↗

  3. Feb 12, 2021Class IIOpen, Classified

    Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.

    Z-1351-2021 · Root cause: Device Design · Recall details · FDA record ↗

  4. Feb 12, 2021Class IIOpen, Classified

    Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.

    Z-1350-2021 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Migration or Expulsion of Device2

Patient problems

  • Failure of Implant1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 3, 2023Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT DURING A CONSULT, A LOCKING BOLT WAS EVIDENT THROUGH THE SKIN. A REVISION PROCEDURE WAS PERFORMED WHERE THEY REARRANGED THE BOLT. AS PER THE REPORTER, THE PATIENT HAS POOR BONE QUALITY WHICH MAY HAVE CONTRIBUTED TO THE PULL OUT. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 17451628 ↗
  • MalfunctionDec 2, 2022Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED THAT ONE OF THE LOCKING SCREWS IN THE IMPLANTED NAIL BACKED OUT. THE PATIENT WAS INSTRUCTED TO CONTINUE LENGTHENING AND LATER DURING A CLINIC THE SCREW WAS NOTED TO BE COMPLETELY BACKED OUT. A REVISION PROCEDURE IS PLANNED ON AN UNKNOWN DATE TO RE-ADJUST THE SCREW. THE PATIENT WAS REPORTED TO HAVE PRESENTED WITH PAIN ADJACENT TO THE BACKED OUT SCREW.

    Patient problems: Pain

    MAUDE report 15906945 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page