Skip to content

Autolith Touch M0054668R0Recalls & reported problems

Northgate Technologies Inc. · Lithotripter, Biliary, BSC, Refurbished · Class 2

All models of Autolith Touch (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00817183020523), received between Nov 10, 2023 and Nov 10, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

4 other recalls cover devices cleared under the same FDA submission (K130368); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    ON 09/21/2023 NORTHGATE TECHNOLOGIES WAS MADE AWARE OF AN ALLEGED MALFUNCTION OF THE DEVICE THAT PREVENT THE DEVICES FROM BEING USED (BLANK SCREEN). THE DEVICE WAS NOT RETURNED FOR EVALUATION UNTIL 09/28/2023. DURING THE SERVICE EVALUATION IT WAS IDENTIFIED THAT THE INSULATED WIRES FROM THE HIGH VOLTAGE TRANSFORMER WERE DISCONNECTED FROM THE HIGH VOLTAGE PCB WITHIN THE DEVICE AMONG A NUMBER OF OTHER INSTANCES OF PHYSICAL DAMAGE NOTED TO THE EXTERIOR AND INTERIOR OF THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18112272 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K130368. One submission often covers many models, so check the product and lot details in each record.

  • Z-1235-2022 · Apr 13, 2022 · Terminated
    AUTOLITH Touch 1.9F 375cm Biliary Electrohydraulic Lithotripter Probe
  • Z-0333-2022 · Oct 25, 2021 · Open, Classified
    Boston Scientific AUTOLITH TOUCH Bipolar Electrohydraulic Lithotripter, REF M005466800. The device is designed to be used with Northgate Technologies Inc. bipolar disposable EHL p
  • Z-1884-2020 · Mar 10, 2020 · Terminated
    AUTOLITH URO-TCH 9FR PROBE 54CM; NTI Catalog Number 9-203-0543 - Product Usage: Probes are used for the fragmentation of urinary/renal and biliary calculi.
  • Z-1883-2020 · Mar 10, 2020 · Terminated
    Probe, 9 FR AUTOLITH; NTI Catalog Number 9-900-54 - Product Usage: Probes are used for the fragmentation of urinary/renal and biliary calculi.

Other Lithotriptor, Electro-Hydraulic devices

All Lithotriptor, Electro-Hydraulic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page