Ultrasound Probe M0068407160Recalls & reported problems
Boston Scientific Corporation · Ultrasound Probe · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729516606), received between Dec 27, 2017 and Jan 16, 2018 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
Patient problems
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJan 16, 2018Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2017 THAT AN ULTRASOUND PROBE WAS USED IN A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE ULTRASOUND PROBE BROKE ABOUT 4CM FROM THE PROXIMAL END. HALF OF THE BROKEN PIECE FELL INSIDE THE PATIENT AND WAS RETRIEVED WITH THE USE OF FORCEPS. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7193258 ↗InjuryDec 27, 2017Break; Break
NOTE: THIS MANUFACTURER REPORT PERTAINS TO ONE OF TWO DEVICES USED IN THE SAME PROCEDURE. REFER TO THE ASSOCIATED MANUFACTURER REPORT 3005099803-2017-03909 FOR THE OTHER ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2017 THAT TWO ULTRASOUND PROBES WERE USED IN A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, TEN MINUTES INTO THE PROCEDURE, THE FIRST ULTRASOUND PROBE (THE SUBJECT OF THIS MFR REPORT) BROKE IN HALF WHEN THE LITHOCLAST WAS AT LEVEL 90. THE SAME ISSUE HAPPENED WITH THE SECOND ULTRASOUND PROBE (MFR REPORT #3005099803-2017-03909). THE BROKEN PIECES WERE RETRIEVED FROM INSIDE THE PATIENT WITH THE USE OF A GRASPER. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING ANOTHER OF THE SAME DEVICE AND ALSO A HOLMIUM LASER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7148066 ↗InjuryDec 27, 2017Break; Break
NOTE: THIS MANUFACTURER REPORT PERTAINS TO ONE OF TWO DEVICES USED IN THE SAME PROCEDURE. REFER TO THE ASSOCIATED MANUFACTURER REPORT 3005099803-2017-03908 FOR THE OTHER ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2017 THAT TWO ULTRASOUND PROBES WERE USED IN A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, TEN MINUTES INTO THE PROCEDURE, THE FIRST ULTRASOUND PROBE (MFR REPORT # 3005099803-2017-03908) BROKE IN HALF WHEN THE LITHOCLAST WAS AT LEVEL 90. THE SAME ISSUE HAPPENED WITH THE SECOND ULTRASOUND PROBE (THE SUBJECT OF THIS MFR REPORT). THE BROKEN PIECES WERE RETRIEVED FROM INSIDE THE PATIENT WITH THE USE OF A GRASPER. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING ANOTHER OF THE SAME DEVICE AND ALSO A HOLMIUM LASER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7148058 ↗
Other Lithotriptor, electro-hydraulic devices
- Lithotripsy Probe 1.9FR 375cm DS M00546620Northgate Technologies Inc. · Lithotriptor, Electro-Hydraulic1 recall1 death19 reports
- Autolith Touch M005466800Northgate Technologies Inc. · Lithotriptor, Electro-Hydraulic1 recall12 reports
- Ultrasound Probe M0068407170Boston Scientific Corporation · Lithotriptor, electro-hydraulic15 reports
- CyberWand USL Ultrasonic Lithotripsy System CW-USLSACybersonics Inc · Lithotriptor, Electro-Hydraulic4 reports
- CyberWand USL Advanced Transducer CW-USLTACybersonics Inc · Lithotriptor, Electro-Hydraulic2 reports
- CyberWand USL NoseCone Kit CW-USLNCCybersonics Inc · Lithotriptor, Electro-Hydraulic1 report
- CyberWand USL Advanced Transducer CW-USLTACybersonics Inc · Lithotriptor, Electro-Hydraulic1 report
- StoneBreaker G52605Cook Incorporated · Lithotriptor, electro-hydraulic1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page