NOA Medical Industries, Inc. 1055050GRYRecalls & reported problems
Hoffmann NOA Medical Industries, LLC · BED - Psych NS Hosp (Gray) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00817492012745), received between Jan 7, 2025 and Jan 7, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Nov 5, 2024Class IIOpen, Classified
Red button used to engage side rails can break or become stuck making it difficult to engage the side rail.
Z-0838-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 7, 2025Mechanical Problem; Mechanical Problem
(B)(4), CUSTOMER REPORTED ISSUE WITH THE RIGHT-SIDE HEAD SIDE RAIL. WHEN PRESSING THE RELEASE BUTTON, THE SIDE RAIL GETS STUCK. THERE IS A METAL PIECE THAT NEEDS TO RETRACT WHEN PUSHING THE RED BUTTON, BUT IT DOES NOT RETRACT COMPLETELY, AND THAT IS WHY THE SIDERAIL GETS STUCK. DEVICE WAS NOT RETURNED FOR INVESTIGATION. SIDE RAIL IS PART OF THE BED PACKAGE. THERE ARE NO PROBLEMS WITH THE BED ITSELF. REPORT IS BEING FILED AFTER THE 30 DAY TIMEFRAME DUE TO CONFUSION REGARDING THE FILING OF MDRS WITH LITTLE INFORMATION. WE ARE REPORTING THE INFORMATION REASONABLY KNOWN TO US OUT OF AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21092230 ↗MalfunctionJan 7, 2025Mechanical Problem; Mechanical Problem
(B)(4), ON TUE, DEC 5, 2023 AT 9:41AM WE RECEIVED AN EMAIL ALLEGING THAT THE RED BUTTON ON THE SIDE RAIL THAT RELEASES THE RAIL TO DROP DOWN WAS STUCK INSIDE AND DOESN'T POP OUT AS IT SHOULD TO LOCK THE RAIL IN PLACE. THE RAIL ISN'T LOCKING INTO PLACE, WHICH ALLOWED HIM TO FALL OUT OF HIS BED DUE TO THE RAIL DROPPING FROM THE UPRIGHT POSITION. FORTUNATELY, HE IS OKAY BUT THE RAIL IS NOT. INCIDENT WAS REPORTED FROM A USER VIA A DISTRIBUTOR, NO OTHER INFORMATION WAS PROVIDED. WE SENT A REPLACEMENT WITH A RETURN LABEL SO THE DEVICE COULD RETURN FOR INVESTIGATION. WE REACHED OUT AGAIN LATER WHEN THE PART HAD NOT RETURNED OFFERING TO SEND ANOTHER RETURN LABEL BUT NEVER HEARD BACK FROM THE CUSTOMER. WE COULD NOT DETERMINE THAT A MALFUNCTION OCCURED. REPORT IS BEING FILED AFTER THE 30 DAY TIMELINE DUE TO CONFUSION OVER THE FILING OF MDRS WITH LITTLE INFORMATION. WE ARE REPORTING WITH THE INFORMATION REASONABLY KNOWN TO US OUT OF AN ABUNDANCE OF CAUTION. SIDE RAIL IS PART OF THE BED PACKAGE. THERE ARE NO PROBLEMS ASSOCIATED WITH THE BED ITSELF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21091763 ↗
Other Bed, Ac-Powered Adjustable Hospital devices
- Centrella™ Bed P7900BHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL2 recalls5 deaths1,937 reports
- Progressa® Bed System P7500AHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death1,377 reports
- VersaCare® Med Surg Bed P3200KHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL3 deaths1,193 reports
- Medline MDR107002EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death374 reports
- Medline MDR107003EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall3 deaths362 reports
- InTouch 2141Stryker Corporation · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths361 reports
- Citadel Plus FX811B3B4AMABBArjohuntleigh AB · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths358 reports
- TotalCare® Bed P1900QHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL317 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page