Skip to content

NOA Medical Industries, Inc. 1055050GRYRecalls & reported problems

Hoffmann NOA Medical Industries, LLC · BED - Psych NS Hosp (Gray) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00817492012745), received between Jan 7, 2025 and Jan 7, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 5, 2024Class IIOpen, Classified

    Red button used to engage side rails can break or become stuck making it difficult to engage the side rail.

    Z-0838-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Mechanical Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 7, 2025Mechanical Problem; Mechanical Problem

    (B)(4), CUSTOMER REPORTED ISSUE WITH THE RIGHT-SIDE HEAD SIDE RAIL. WHEN PRESSING THE RELEASE BUTTON, THE SIDE RAIL GETS STUCK. THERE IS A METAL PIECE THAT NEEDS TO RETRACT WHEN PUSHING THE RED BUTTON, BUT IT DOES NOT RETRACT COMPLETELY, AND THAT IS WHY THE SIDERAIL GETS STUCK. DEVICE WAS NOT RETURNED FOR INVESTIGATION. SIDE RAIL IS PART OF THE BED PACKAGE. THERE ARE NO PROBLEMS WITH THE BED ITSELF. REPORT IS BEING FILED AFTER THE 30 DAY TIMEFRAME DUE TO CONFUSION REGARDING THE FILING OF MDRS WITH LITTLE INFORMATION. WE ARE REPORTING THE INFORMATION REASONABLY KNOWN TO US OUT OF AN ABUNDANCE OF CAUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21092230 ↗
  • MalfunctionJan 7, 2025Mechanical Problem; Mechanical Problem

    (B)(4), ON TUE, DEC 5, 2023 AT 9:41AM WE RECEIVED AN EMAIL ALLEGING THAT THE RED BUTTON ON THE SIDE RAIL THAT RELEASES THE RAIL TO DROP DOWN WAS STUCK INSIDE AND DOESN'T POP OUT AS IT SHOULD TO LOCK THE RAIL IN PLACE. THE RAIL ISN'T LOCKING INTO PLACE, WHICH ALLOWED HIM TO FALL OUT OF HIS BED DUE TO THE RAIL DROPPING FROM THE UPRIGHT POSITION. FORTUNATELY, HE IS OKAY BUT THE RAIL IS NOT. INCIDENT WAS REPORTED FROM A USER VIA A DISTRIBUTOR, NO OTHER INFORMATION WAS PROVIDED. WE SENT A REPLACEMENT WITH A RETURN LABEL SO THE DEVICE COULD RETURN FOR INVESTIGATION. WE REACHED OUT AGAIN LATER WHEN THE PART HAD NOT RETURNED OFFERING TO SEND ANOTHER RETURN LABEL BUT NEVER HEARD BACK FROM THE CUSTOMER. WE COULD NOT DETERMINE THAT A MALFUNCTION OCCURED. REPORT IS BEING FILED AFTER THE 30 DAY TIMELINE DUE TO CONFUSION OVER THE FILING OF MDRS WITH LITTLE INFORMATION. WE ARE REPORTING WITH THE INFORMATION REASONABLY KNOWN TO US OUT OF AN ABUNDANCE OF CAUTION. SIDE RAIL IS PART OF THE BED PACKAGE. THERE ARE NO PROBLEMS ASSOCIATED WITH THE BED ITSELF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21091763 ↗

Other Bed, Ac-Powered Adjustable Hospital devices

All Bed, Ac-Powered Adjustable Hospital devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page