Skip to content

XScribe XSCRIBE-6AA-AFAXBRecalls & reported problems

Mortara Instrument, Inc. · Cardiac Stress Testing System · Class 2

All models of XScribe (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K022618, K152944); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Oct 11, 2024Class IIOpen, Classified

    There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.

    Z-0325-2025 · Root cause: Software design · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K152944 on Jan 6, 2016 (decision: Substantially Equivalent). Submitted by Mortara Instrument, Inc. as “Xscribe Stress Exercise Testing Systems, Q-Stress Stress Exercise Testing Systems”. The FDA review took 92 days. FDA record ↗

Cleared through Special 510(k) K022618 on Sep 6, 2002 (decision: Substantially Equivalent). Submitted by Mortara Instrument, Inc. as “X-Scribe Ii”. The FDA review took 30 days. FDA record ↗

The FDA regulates product code DPS (electrocardiograph) under 21 CFR 870.2340 as a Class II device. 21 CFR 870.2340 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K022618, K152944. One submission often covers many models, so check the product and lot details in each record.

  • Z-0321-2025 · Oct 11, 2024 · Open, Classified
    Hillrom Welch Allyn Q-Stress Cardiac Stress Testing System; cardiac stress test
  • Z-0706-2024 · Dec 8, 2023 · Open, Classified
    XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-BXXXX; b) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-DXXXX; c) XS6
  • Z-0705-2024 · Dec 8, 2023 · Open, Classified
    Q-Stress Cardiac Stress Testing System, version 6, with below product name / product codes: a) QS6 AM12Q KIT AHA XML / QS6-BXXCS: b) QS6 AM12Q KIT BCRT IEC / QS-6CC-DGCXX: c) QS6
  • Z-1817-2020 · Mar 18, 2020 · Terminated
    Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wire

Common questions about the XScribe XSCRIBE-6AA-AFAXB

Has the XScribe XSCRIBE-6AA-AFAXB been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0325-2025 (Class II), began on Oct 11, 2024: There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the XScribe XSCRIBE-6AA-AFAXB FDA approved?

It was cleared through Special 510(k) K022618 on Sep 6, 2002; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the XScribe XSCRIBE-6AA-AFAXB, and is it still sold?

The FDA's device database lists Mortara Instrument, Inc. as the company responsible for it (brand XScribe, model XSCRIBE-6AA-AFAXB). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Electrocardiograph devices

All Electrocardiograph devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page