AmeriWater MROSRecalls & reported problems
Ameriwater, LLC · MROS, Single Patient Portable RO, 730 GPD, 115V · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00817927020000), received between Feb 13, 2025 and Feb 13, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Product Quality Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K111740); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Product Quality Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 13, 2025Product Quality Problem; Product Quality Problem
AMERIWATER IS SELF-REPORTING AN ISSUE EXPERIENCED WITH THE PENTAIR CHLROPLUS20 CARBON BLOCK FILTER USED AS PRETREATMENT FOR THE AMERIWATER MODEL MROS AND AMERIWATER CENTURION, MODEL MROC. ON (B)(6) 2025, AMERIWATER RECEIVED NOTICE FROM A LOCAL CUSTOMER THAT THE CARBON BLOCK FILTERS (AMERIWATER PART NUMBER: 20-5012) USED IN THE AMERIWATER MODEL MROS PORTABLE REVERSE OSMOSIS SYSTEM WERE EXPERIENCING PREMATURE CHLORINE BREAKTHROUGH. NO ADVERSE EVENTS, INJURY, ILLNESS, OR PATIENT EXPOSURE WAS REPORTED. THE PREMATURE CHLORINE BREAKTHROUGH WAS BELIEVED TO BE THE RESULT OF WATER CONDITIONS AT THE FACILITY. AMERIWATER REPLACED THE FILTERS AND RETURNED THE ALLEGEDLY NONCONFORMING FILTER CARTRIDGES TO OUR FACILITY FOR TESTING. CHLORINE BREAKTHROUGH WAS EXPERIENCED AFTER LESS THAN A WEEK OF USE, WHEN THE FILTER CAPACITY NORMALLY LASTED LONGER THAN 1 MONTH. ON (B)(6) 2025 AMERIWATER REPLACED THE NONCONFORMING FILTERS WITH FILTERS FROM A KNOWN GOOD LOT. 16-JAN-2025, A SECOND REPORT WAS RECEIVED FROM AN END USER IN (B)(6). NO ADVERSE EVENTS, INJURY, ILLNESS, OR PATIENT EXPOSURE WAS REPORTED. AMERIWATER HALTED SHIPMENT OF THE 20-5012 FILTERS, QUARANTINED THE FILTERS, AND BEGAN TESTING SAMPLES FRO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21379303 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111740. One submission often covers many models, so check the product and lot details in each record.
- Z-2321-2012 · Aug 9, 2012 · TerminatedAmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com. The AmeriWater MRO Portable Reverse Osmosis Systems are wat
Other Subsystem, Water Purification devices
- Tablo(R) Hemodialysis System PN-0003000Outset Medical, Inc. · Subsystem, Water Purification3 recalls34 deaths83 reports
- AquaC UNO HVIVONIC GmbH · Subsystem, Water Purification4 recalls0 reports
- AquaBplus AquaBplus B2 2500VIVONIC GmbH · Subsystem, Water Purification1 recall0 reports
- AquaBplus & AquaBplus B2 & AquaBplus HF 2500VIVONIC GmbH · Subsystem, Water Purification1 recall0 reports
- TabloCart™ with Prefiltration Drawer PN-0006813Outset Medical, Inc. · Subsystem, Water Purification1 recall0 reports
- AquaA AquaA 900HVIVONIC GmbH · Subsystem, Water Purification1 recall0 reports
- Ultrafilter U 9000 112062Gambro Dialysatoren GmbH · Subsystem, water purification6 reports
- Millenium HX RO,MILLENIUM HX,750GPD,115~,60HZMar Cor Purification, Inc. · Subsystem, Water Purification3 reports