Skip to content

AmeriWater MROSRecalls & reported problems

Ameriwater, LLC · MROS, Single Patient Portable RO, 730 GPD, 115V · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00817927020000), received between Feb 13, 2025 and Feb 13, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Product Quality Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K111740); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2025Product Quality Problem; Product Quality Problem

    AMERIWATER IS SELF-REPORTING AN ISSUE EXPERIENCED WITH THE PENTAIR CHLROPLUS20 CARBON BLOCK FILTER USED AS PRETREATMENT FOR THE AMERIWATER MODEL MROS AND AMERIWATER CENTURION, MODEL MROC. ON (B)(6) 2025, AMERIWATER RECEIVED NOTICE FROM A LOCAL CUSTOMER THAT THE CARBON BLOCK FILTERS (AMERIWATER PART NUMBER: 20-5012) USED IN THE AMERIWATER MODEL MROS PORTABLE REVERSE OSMOSIS SYSTEM WERE EXPERIENCING PREMATURE CHLORINE BREAKTHROUGH. NO ADVERSE EVENTS, INJURY, ILLNESS, OR PATIENT EXPOSURE WAS REPORTED. THE PREMATURE CHLORINE BREAKTHROUGH WAS BELIEVED TO BE THE RESULT OF WATER CONDITIONS AT THE FACILITY. AMERIWATER REPLACED THE FILTERS AND RETURNED THE ALLEGEDLY NONCONFORMING FILTER CARTRIDGES TO OUR FACILITY FOR TESTING. CHLORINE BREAKTHROUGH WAS EXPERIENCED AFTER LESS THAN A WEEK OF USE, WHEN THE FILTER CAPACITY NORMALLY LASTED LONGER THAN 1 MONTH. ON (B)(6) 2025 AMERIWATER REPLACED THE NONCONFORMING FILTERS WITH FILTERS FROM A KNOWN GOOD LOT. 16-JAN-2025, A SECOND REPORT WAS RECEIVED FROM AN END USER IN (B)(6). NO ADVERSE EVENTS, INJURY, ILLNESS, OR PATIENT EXPOSURE WAS REPORTED. AMERIWATER HALTED SHIPMENT OF THE 20-5012 FILTERS, QUARANTINED THE FILTERS, AND BEGAN TESTING SAMPLES FRO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21379303 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111740. One submission often covers many models, so check the product and lot details in each record.

  • Z-2321-2012 · Aug 9, 2012 · Terminated
    AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com. The AmeriWater MRO Portable Reverse Osmosis Systems are wat

Other Subsystem, Water Purification devices

All Subsystem, Water Purification devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.