AquaBplus AquaBplus B2 2500Recalls & reported problems

VIVONIC GmbH · Additional membrane filtration unit for AquaBplus Water Treatment Systems. AquaBplus B2 Unit for: AquaBplus & AquaBplus B2 2500 (Primary DI No.: 04251285200111) AquaBplus & AquaBplus B2 & AquaBplus HF 2500 (Primary DI No.: 04251285200227) · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Aug 3, 2021Terminated

    AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation

    Z-1364-2022 · Root cause: Labeling Change Control · Recall details · FDA record ↗