Barricade Coil System 900106Recalls & reported problems
Balt USA LLC · Barricade Coil System (BCS) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The BCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The BCS consists of an implantable coil attached to a delivery pusher. The delivery pusher is powered by the Blockade Detachment Controller (BDC) designed specifically for the BCS. The device comes pre-loaded with one 9-volt battery. A BlockWire Connecting Cable or Handheld Detachment Cable, comprised of a red and black cable leads are required to detach Barricade Detachable Coils and must be used in conjunction with a return electrode. · Class 2 · Implant
All models of Barricade Coil System (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00818053021053), received between Sep 10, 2018 and Feb 16, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Stretched” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Stretched3
- Unraveled Material2
- Detachment of Device or Device Component1
- Device Slipped1
Patient problems
- Aneurysm2
- Patient Problem/Medical Problem2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 16, 2021Device Slipped; Stretched; Detachment of Device or Device Component; Device Slipped; Stretched; Detachment of Device or Device Component
IT WAS REPORTED THAT: "THE CASE WAS AN ACOM ANUERYSM MEASURING 3MMX2.5MM AND HAVING A NECK OF 2.3 MM. SINCE IT WAS A RUPTURED ANUERYSM THE PHYSICIAN PLANNED TO DO A BALLOON ASSISTED COILING. THE SUPPORT HARDWARE WAS NEUROMAX ,FARGOMAX 6F, ECLISPE 6X9 ,VASCO 10 MP WITH HYBRID 12/14 GUIDE WIRE. THE FIRST COIL WAS COMPLEX FINISH 3MMX4MM. THE COIL WAS DEPLOYED WITH BALLOON INFLATION AND THE SECOND COIL WAS COMPLEX FINISH 2MMX3MM AND WHEN HALF A CM OF COIL WAS REMAINING TO BE PUSHED,THE MICRO WAS OUT OF THE ANUERYSM.THE PHYSCIAN WANTED TO WITHDRAW THE COIL IN ORDER TO REPOSITION THE MICRO WITH WIRE AND WHEN HE WAS ATTEMPTING TO WITHDRAW THE COIL IT HAD PREMATURELY DETACHED (NOTE:- NO MANUPILATION WAS DONE TO PUSH THE MICRO WITH THE COIL SUPPORT).NO MOVEMENT OF COIL WAS OBSERVED ON WITHDRAWL OF THE PUSH WIRE THE COIL HAD COME OUT HALF STRECHED ,THE PHYSCIAN USED THE HYBRID 12/14 THROUGH THE VASCO 10 TO ENSURE NO COIL WAS STILL PRESENT IN THE MICRO WITH THE ECLIPSE INFLATED BUT DURING THIS PROCESS SOME PART OF THE COIL CAME INTO THE ACA.THE MICRO AND BALLOON WAS REMOVED. THE PHYSCIAN USED XT 27 AND SOLITAIRE 4X20,CAPTURED THE COIL & REMOVED IT. POST REMOVAL OF THE STRECHED COIL ,THE PHYSC
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11334011 ↗MalfunctionSep 10, 2018Stretched; Unraveled Material; Stretched; Unraveled Material
IT WAS REPORTED THAT: "THE SURGEON TRIED TO PLACE 2MM X 3CM COMPLEX FINISH COIL INTO THE AN BUT THE COIL HAS UNRAVELED DUE TO THE TIP OF SL-10 DROPPED OFF FROM THE AN. THEN UNRAVELED COIL(PRIMARY COIL) MOVED TO SIPHON REGION BY BLOOD FLOW. THEREFORE, A STENT (4.5 X 3 NEUROFORM ATLAS / STRYKER) WAS PLACED SIPHON REGION TO LOCK UNRAVELED COIL AND THE PUSHER WIRE WAS REMOVED, HOWEVER A PART OF COIL TAIL REMAINS WITHIN THE PATIENT. SURGEON'S COMMENTS: HE COMMENTED THAT THIS CASE MIGHT HAD BEEN OCCURRED DUE TO THE COIL CATCHING ON STRUT OF THE STENT. "
Patient problems: Aneurysm; Patient Problem/Medical Problem
MAUDE report 7862107 ↗MalfunctionSep 10, 2018Stretched; Unraveled Material; Stretched; Unraveled Material
IT WAS REPORTED THAT: "4TH COIL (2MM X 3CM COMPLEX FINISH COIL) WAS STRETCHED (UNRAVELED). THE MICROCATHETER WAS DROPPED OFF FROM THE AN UNINTENTIONALLY DURING INSERTION OF THE COIL AND REMAINING 5MM COIL COULD NOT BE PLACED. THE SURGEON TRIED TO RE-INSERT THE MICROCATHETER INTO THE AN BUT IT COULD NOT BE SUCCESSFUL. SO, WHOLE COIL SYSTEM WAS REMOVED FROM THE PATIENT. SURGEON'S COMMENTS: THE SURGEON COMMENTED THAT THE SELECTION OF LONG COIL (3CM) WAS NOT PROPERLY FOR THIS CASE. UNRAVEL OF COIL MIGHT HAVE DUE TO OPERATION OF MICROCATHETER DURING RE-INSERTION."
Patient problems: Aneurysm; Patient Problem/Medical Problem
MAUDE report 7862104 ↗
Other Device, Neurovascular Embolization devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page