Optima Coil System OPTI0308CSS10 (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00818053021541), received between Sep 22, 2020 and Mar 7, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 3, 2024Open, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Expulsion1
  • Unintended Movement1

Patient problems

  • Aneurysm1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 7, 2023Expulsion; Expulsion

    IT WAS REPORTED THAT: "DURING THE ANEURYSM EMBOLIZATION PROCEDURE LOCATED IN THE ANTERIOR COMMUNICATING ARTERY, IN WHICH THERE WAS A COAXIAL SYSTEM WITH BENCHMARK AND A HEADWAY 17 MICROCATHETER, ANEURYSM CATHETERIZATION WAS PERFORMED. THE COIL OPTIMA 3MM X 8CM COMPLEX DEVICE IS ADVANCED, WHICH ADVANCES NORMALLY THROUGH THE MICROCATHETER HUB, THE HYPOTUBE CONTAINING THE CONTAINER IS REMOVED, THE COIL CONTINUES TO BE ADVANCED, THEN THE TWO COIL LOOPS ARE RELEASED WITHOUT INCONVENIENCE. HOWEVER, IN THE THIRD LOOP A "CLICK" IS FELT WHEN THE COIL IS PUSHED AFTER WHICH IT IS FULLY RELEASED, WITH THE PROXIMAL PORTION OF THE COIL IN THE LEFT ANTERIOR CEREBRAL ARTERY AND LEFT INTERNAL CAROTID ARTERY. AFTER THIS, THERE IS NO POSSIBILITY OF ANTEROGRADE OR RETROGRADE COIL MOVEMENT, SO IT IS DECIDED TO PLACE AN "LVIS" STENT TRYING TO EXCLUDE THE COIL DEVICE FROM CIRCULATION." INCIDENT REPORT FORM SUBMITTED TO BALT INDICATED THAT NO PATIENT INJURY WAS SUSTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16499646 ↗
  • MalfunctionSep 22, 2020Detachment of Device or Device Component; Unintended Movement; Detachment of Device or Device Component; Unintended Movement

    IT WAS REPORTED: DURING COILING PROCEDURE OF A PCOM ANX, THE PHYSICIAN BEGAN DELIVERING A 3X8 CSS10 COIL. THIS WAS THE FOURTH COIL OF THE CASE STARTING WITH A 4X7CSFMX, 3X10CSS10, AND 3X8CSS10. THE PHYSICIAN WAS ABLE TO DELIVER APPROXIMATELY 75% OF THE COIL, HOWEVER, THE LAST COUPLE CMS KEPT PUSHING THE MICROCATHETER OUT OF THE ANX. THE PHYSICIAN ATTEMPTED TO REPOSITION THE COIL MULTIPLE TIMES OVER APPROXIMATELY 15 MINUTES WITH NO SUCCESS. WHILE REMOVING THE COIL, THE PHYSICIAN FELT A SLIGHT HITCH PROMPTING TO BELIEVE THE COIL DETACHED. ONLY A COUPLE CM REMAINED IN THE MC SO SHEN THE PHYSICIAN TRIED REMOVING BOTH THE COIL AND THE MC, THE COIL RELEASED FROM THE TIP OF THE MC AND FLOWED DISTAL TO THE ANX BUT WAS STILL ANCHORED WITHIN THE COIL MASS. THE PHYSICIAN WAS ABLE TO LASSO THE COIL AND REMOVE THE COIL COMPLETELY WITH NO ADVERSE EVENTS. WE THEN CONTINUED COILING THE ANX SUCCESSFULLY AND WAS PLEASED WITH THE RESULTS.

    Patient problems: Aneurysm

    MAUDE report 10570826 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.

  • Z-1988-2025 · May 8, 2025 · Open, Classified
    Optima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →