Optima Coil System OPTI0517CSF10 (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053021756), received between May 13, 2019 and May 13, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 3, 2024Open, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Aneurysm1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT: "THE DOCTOR WAS TREATING A 5.18MM X 5.78MM RUPTURED MCA BIFURCATION ANEURYSM. THE PHYSICIAN HAD TWO MICROCATHETERS (SL10) IN THE ANEURYSM, ONE WITH THE OPTIMA COIL AND THE OTHER WITH A STRYKER 6X20 3D. AS BOTH COILS WERE ADVANCED TO DETACHMENT ZONES, DR. (B)(6) ATTEMPTED TO DETACH THE OPTIMA COIL, THE DETACHER BLINKED RED AND GREEN AS SOON AS THE COIL PUSHER WAS BOTTOMED OUT. DR. (B)(6) PRESSED THE DETACHMENT BUTTON AT LEAST 10 TIMES, AFTER MULTIPLE FAILED ATTEMPTS, A SECOND DETACHER WAS OPENED WITH THE SAME RESULT. A THIRD DETACHER WAS OPENED AND THE SAME EVENTS OCCURRED. DR. (B)(6) THEN ATTEMPTED TO WITHDRAW THE OPTIMA, APPROXIMATELY 4 CM OF COIL WAS PULLED BACK WHEN THE COIL AND PUSHER DETACHED. DR. (B)(6) BEGAN WITHDRAWING THE MICROCATHETER AND COIL WHICH CAUSED THE STRYKER COIL AND OPTIMA COIL TO RELOCATE TO THE MCA ARTERY. DR. (B)(6) THEN WENT UP WITH A GOOSENECK SNARE TO RETRIEVE THE COIL MASS. HE WAS ABLE TO SNARE THE OPTIMA COIL AND MICROCATHETER AND REMOVED THEM FROM THE PATIENT. DR. (B)(6) WAS THEN ABLE TO USE A WIRE AND MICROCATHER TO RE- POSITION THE STRYKER COIL BACK INTO THE ANEURYSM. THE CASE ENDED WITH NO PATIENT INJURY AND IS COMPLICATION FRE

    Patient problems: Aneurysm; Patient Problem/Medical Problem

    MAUDE report 8605820 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.

  • Z-1988-2025 · May 8, 2025 · Open, Classified
    Optima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →