Skip to content

Optima Coil System OPTI0923CST10 (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053021893), received between Aug 15, 2019 and Aug 15, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

2 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. May 8, 2025Open, Classified

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    Z-1988-2025 · Root cause: Vendor change control · FDA record ↗

  2. Jan 3, 2024Open, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Fistula1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 15, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT: "PHYSICIAN WAS TREATING A CCF FISTULA. HE WAS TRYING TO REMOVE A 9X23 COMPLEX STANDARD OPTIMA AND THE COIL BROKE OFF THE DETACHER LEAVING PART OF THE COIL IN THE FISTULA AND PART IN THE CATHETER. HE HAD TO USE A STENT RETRIEVER TO RETRIEVE THE COIL."

    Patient problems: Fistula

    MAUDE report 8899552 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.

  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.