Skip to content

Optima Coil System OPTI2265COM18 (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053022067), received between Aug 26, 2020 and Aug 26, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Premature Separation” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 3, 2024Open, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · FDA record ↗

Most reported problems

Device problems

  • Premature Separation1

Patient problems

  • Aneurysm1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2020Premature Separation; Premature Separation

    IT WAS REPORTED THROUGH AN AUTOMATED REPORTING SYSTEM FROM ACTIVE CLINICAL STUDY: "THE CASE WAS ACOM NARROW NECK ANEURYSM MEASURING 25MMX18MM AND NECK MEASURED 6MM.THE PHYSICIAN PLANNED TO DO SIMPLE COILING OF THE ANEURYSM.THE SUPPORT HARDWARE WAS 8F NEURONMAX ,6F FARGOMAX ,VASCO 10D & TRAXCESS GUIDE WIRE .THE COILS USED WERE COMPLEX18 22X65-2NO,20X65-2NO,14X47-1NO,COMPLEX10STANDARD10X34-1NO. THE LAST COIL WAS OPTIMA HELICAL STANDARD10X30-1NO TO COMPLETE THE PROCEDURE. OUT OF THE COMPLEX18- 22X65 GOT PRE MATURELY DETACHED AND THE SAME WAS RETRIEVED BY USING VASCO 21 AND SOLITAIRE STENT RETRIEVER. WE WERE NOT ABLE TO COLLECT THE DEVICE PUSHWIRE, COILS AND MICRO CATH AS THE PATIENT WAS SUSPECTED FOR COVID 19 AND ALL THE THEM WERE DISCARDED AS PER HOSPITAL POLICY. BUT WE WERE ABLE TO GET THE RECORDED IMAGES."

    Patient problems: Aneurysm

    MAUDE report 10456781 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.

  • Z-1988-2025 · May 8, 2025 · Open, Classified
    Optima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.