HYBRID Guidewire HYBRID1214DARecalls & reported problems
Balt USA LLC · HYBRID Guidewire is a coiled wire that is designed to fit inside a percutaneous microcatheter for the purpose of directing the catheter through a blood vessel. It is used in association with catheters to enable insertion and easier navigation into the intracranial vascular branches for diagnositc or therapeutic use. Guidewires constitute an important element that conditions the success of the operation. They can be pushed and torqued to move forward through patient's tortuous anatomy. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00818053022791), received between Nov 22, 2023 and Oct 1, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Inaccurate Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 1, 2024Break; Break
IT WAS REPORTED THAT: "12/14 HYBRID WIRE WAS ATTEMPTED FOR DIFFICULT ANATOMY AT THE A2 TAKEOFF. THE DISTAL END OF WIRE SNAPPED OFF IN THE CATHETER SO THEY HAD TO REMOVE THE WIRE AND CATHETER. THE PORTION OF THE CATHETER IN WHICH THE WIRE SNAPPED OFF DID NOT SEEM TORTUOUS, THE PHYSICIAN WAS CONFUSED WHY IT WOULD BREAK AT THAT POINT. I HAVE THE WIRE TO RETURN, NOT THE CATHETER" INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20345062 ↗MalfunctionAug 14, 2024Break; Break
IT WAS REPORTED THAT: "TIP OF WIRE BROKE OFF AND WAS NOT RETRIEVABLE". INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19983162 ↗MalfunctionMar 12, 2024Break; Break
IT WAS REPORTED THAT: "ON (B)(6) DURING A PROSTATE ARTERY EMBOLIZATION A HYBRIB 1214DA GUIDE WIRE WAS INTRODUCED INTO A MICRO CATHETER. AS SOON AS THE PHYSICIAN PULLED THE WIRE BACK, HE SAYS THAT THE WIRE WAS STUCK AND BROKE. THEY HAD TO PULL THE WHOLE SYSTEM OUT AND START OVER ON THIS SIDE." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18890763 ↗MalfunctionFeb 16, 2024Break; Break
IT WAS REPORTED THAT: "ON (B)(6) DURING A PROSTATE ARTERY EMBOLIZATION A HYBRIB 1214DA GUIDE WIRE WAS INTRODUCED INTO A PROWLER SELECT PLUS-J SHAPE MICRO CATHETER. AS SOON AS THE PHYSICIAN PULLED THE WIRE BACK, HE SAYS THAT THE WIRE WAS STUCK AND BROKE. THEY HAD TO PULL THE WHOLE SYSTEM OUT AND START OVER ON THIS SIDE. WHEN THEY MOVED ON TO TREAT THE OTHER SIDE, THEY USED A SWIFT NINJA STEERABLE MICRO CATHETER AND INTRODUCED A HYBRID 1214DA FOR THIS SIDE AS WELL, AND THE SAME THING HAPPENED. WHEN THEY PULLED THE WIRE BACK, THE WIRE WAS BROKEN." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18721120 ↗MalfunctionFeb 16, 2024Break; Break
IT WAS REPORTED THAT: "ON JAN. 17TH DURING A PROSTATE ARTERY EMBOLIZATION A HYBRIB 1214DA GUIDE WIRE WAS INTRODUCED INTO A PROWLER SELECT PLUS-J SHAPE MICRO CATHETER. AS SOON AS THE PHYSICIAN PULLED THE WIRE BACK, HE SAYS THAT THE WIRE WAS STUCK AND BROKE. THEY HAD TO PULL THE WHOLE SYSTEM OUT AND START OVER ON THIS SIDE. WHEN THEY MOVED ON TO TREAT THE OTHER SIDE, THEY USED A SWIFT NINJA STEERABLE MICRO CATHETER AND INTRODUCED A HYBRID 1214DA FOR THIS SIDE AS WELL, AND THE SAME THING HAPPENED. WHEN THEY PULLED THE WIRE BACK, THE WIRE WAS BROKEN." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18721106 ↗MalfunctionNov 22, 2023Inaccurate Information; Inaccurate Information
IT WAS IDENTIFIED DURING THE INVESTIGATION OF SUPPLIER BALT EXTRUSION'S HEALTH HAZARD EVALUATION PROCESS (BALT EXTRUSION HHE-030) THAT THE PRODUCT LOT NUMBER 00521955 MAY HAVE BEEN IMPACTED BY A POTENTIAL MIX-UP OF PRODUCTS. FOR THIS LOT NUMBER (00521955), THE INDICATED HYBRID GUIDEWIRE MODEL IS HYBRID1214DA ON THE POUCH AND CARTON LABEL. FOLLOWING BALT EXTRUSION'S INVESTIGATION, IT WAS NOTED THAT A POTENTIAL DEVICE MIX-UP MAY HAVE OCCURRED WITH ANOTHER MANUFACTURING LOT NUMBER (00520501) WHICH CORRESPONDS TO A DIFFERENT MODEL/GUIDEWIRE SIZE (HYBRID008J)
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18194814 ↗
Other Wire, Guide, Catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page