Optima Coil System OPTI0204CSSMXRecalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053024993), received between May 29, 2026 and May 29, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Migration” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K200030); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 3, 2024Open, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · FDA record ↗

Most reported problems

Device problems

  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 29, 2026Migration

    IT WAS REPORTED THAT: "THE PHYSICIAN WAS TREATING A RUPTURED RIGHT PCOM ANEURYSM. SHE WAS ABLE TO SUCCESSFULLY DEPLOY A 4X7 OPTIMAX. SHE FOLLOWED WITH A 2X4 OPTIMAX. SHE WAS ABLE TO POSITION THE COIL AFTER ATTEMPTING MULTIPLE ADJUSTMENTS PULLING/PUSHING TO FIND THE RIGHT PLACEMENT. SHE RECEIVED A SUCCESSFUL GREEN LIGHT SEQUENCE ON THE DETACHER. SHE SLOWLY PULLED BACK THE PUSHER WIRE FROM THE COIL MASS UNTIL THE WIRE WAS NO LONGER VISIBLE ON THE SCREEN WHEN THE TAIL OF THE COIL GOT SUCKED INTO THE MICROCATHETER. SHE USED THE PUSHER WIRE TO REPOSITION THE COIL AND TUCK IT BACK INTO THE MASS THEN STARTED PULLING BACK ONCE AGAIN. AGAIN, THE PUSHER WIRE WAS OFF THE SCREEN WHEN THE COIL SUCKED BACK INTO THE MICRO. SHE THEN ADVANCED THE MICRO AND GUIDEWIRE AT THE SAME TIME WHICH FREED THE COIL; HOWEVER, IT DISPLACED THE LAST LOOP AND IT MOVED INTO THE ICA. SHE TRIED ONCE AGAIN TO PUSH THE TAIL BACK INTO THE COIL MASS USING THE GUIDEWIRE AND MICRO WHICH ENDED UP FURTHER KICKING THE TAIL INTO THE ICA. SHE WAS WORRIED ABOUT THE COIL FLYING AWAY AND DECIDED TO STENT DOWN THE COIL TAIL. NO FURTHER COILS WERE DEPLOYED. SHE USED A 45 DEGREE SL10 AS THE MICRO." UPDATE 01MAY2026 - ADDITIONAL INFOR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25333034 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K200030. One submission often covers many models, so check the product and lot details in each record.

  • Z-1988-2025 · May 8, 2025 · Open, Classified
    Optima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.
  • Z-0660-2022 · Jan 4, 2022 · Open, Classified
    The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil del

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.