Optima Coil System OPTI0307CSSMXRecalls & reported problems
Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant
All models of Optima Coil System (379)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053025044), received between Apr 7, 2023 and Apr 7, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Migration” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
3 other recalls cover devices cleared under the same FDA submission (K200030); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- May 8, 2025Class IIOpen, Classified
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
Z-1988-2025 · Root cause: Vendor change control · Recall details · FDA record ↗
- Jan 3, 2024Class IIOpen, Classified
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Most reported problems
Device problems
- Migration1
- Premature Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 7, 2023Migration; Premature Separation; Migration; Premature Separation
IT WAS REPORTED THAT: "NEURON MAX 90CM, SL10 45 DEGREE MC PARALLEL WITH A SCEPTER XC 4X11. THE PLAN WAS TO DO A BALLOON ASSISTED COILING OF A WIDE NECK SPHERICAL ANEURSYM ON THE ICA. INITIALLY FRAMED WITH A 4X7 OPTIMAX SOFT COMPLEX ALONG WITH BALLOON ASSISTANCE. THE FRAME SET AND DEPLOYED WITHOUT ANY ISSUES. THE FOLLOWING COIL WAS A 3X7 OPTIMAX SUPERSOFT COMPLEX AND AS THE INITIAL PHASES OF DEPLOYMENT WAS INITIATED, A LITTLE RESISTANCE WAS FELT BY THE FELLOW WHEN THE BALLOON WAS UP. THE ATTENDING PHYSICIAN TOOK OVER AND TRIED TO ADVANCE AND PULL BACK AS THE COIL WAS NOT SITTING RIGHT. AS SOON AS HE ATTEMPTED THAT MANEUVER, HE LOST 1:1 CONTROL AND THE COIL PREMATURELY DETACHED WITH IT PARTIALLY IN THE SL10 45 DEGREE MICROCATHETER. CAREFUL REMOVAL OF THE SL10 45 DEGREE WAS ATTEMPTED, HOWEVER THE 3X7 OPTIMAX COIL FLIPPED OUT AND MIGRATED INTO THE MCA. AS A RESULT, THE PHYSICIAN HAD TO SNARE THE THE COIL OUT OF THE MCA AND WAS ABLE TO SAFELY REMOVE THE 3X7 OPTIMAX WHILE KEEPING THE FRAMED COIL (4X7 OPTIMAX SOFT). THE CASE WAS CONCLUDED AFTER REMOVAL OF THE 3X7 OPTIMAX SUPERSOFT AND THE PATIENT IS DOING WELL."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16702315 ↗
FDA clearance and classification
Cleared through Special 510(k) K200030 on Feb 1, 2020 (decision: Substantially Equivalent). Submitted by Balt USA, LLC as “Optima Coil System”. The FDA review took 25 days. FDA record ↗
The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K200030. One submission often covers many models, so check the product and lot details in each record.
- Z-0103-2023 · Sep 7, 2022 · Open, ClassifiedOptima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
- Z-1680-2022 · Jul 26, 2022 · Open, ClassifiedOptima Coil System Model OPTI0520CSF10. For endovascular embolization.
- Z-0660-2022 · Jan 4, 2022 · Open, ClassifiedThe Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil del
Common questions about the Optima Coil System OPTI0307CSSMX
Has the Optima Coil System OPTI0307CSSMX been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1988-2025 (Class II), began on May 8, 2025: Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Optima Coil System OPTI0307CSSMX?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 7, 2023. That report lists device problems “Migration” (1) and “Premature Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Optima Coil System OPTI0307CSSMX?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Optima Coil System OPTI0307CSSMX FDA approved?
It was cleared through Special 510(k) K200030 on Feb 1, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Optima Coil System OPTI0307CSSMX MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Optima Coil System OPTI0307CSSMX, and is it still sold?
The FDA's device database lists Balt USA LLC as the company responsible for it (brand Optima Coil System, model OPTI0307CSSMX). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Device, Vascular, For Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization117 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page