Optima Coil System OPTI1040CSFMXRecalls & reported problems
Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053025280), received between Jul 28, 2025 and Jul 28, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Migration” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
3 other recalls cover devices cleared under the same FDA submission (K200030); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 8, 2025Open, Classified
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
Z-1988-2025 · Root cause: Vendor change control · FDA record ↗
- Jan 3, 2024Open, Classified
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
Z-1041-2024 · Root cause: Material/Component Contamination · FDA record ↗
Most reported problems
Device problems
- Migration1
- Premature Separation1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 28, 2025Stretched; Migration; Premature Separation; Stretched; Migration; Premature Separation
IT WAS REPORTED THAT: "<PATIENT INFORMATION> SEX: UNKNOWN DISEASE NAME: AS PER BELOW. PROCEDURE: UNKNOWN MICRO CATHETER: HEADWAY 17/TERUMO ASSIST STENT: ENTERPRISE/JOHNSON&JOHNSON Y CONNECTOR: UNKNOWN USED COILS: UNKNOWN. <DESCRIPTION> IN THIS CASE OF RECANALIZATION OF A LARGE ANEURYSM OVER 15 MM NEAR THE P1 SEGMENT, AN ASSISTING ENTERPRISE STENT WAS DEPLOYED IN THE PARENT VESSEL AND THE MICROCATHETER WAS JAILED TO INITIATE THE PROCEDURE. WHILE INSERTING THE OPTIMA COIL SYSTEM AS THE FIRST COIL AND ADVANCING IT APPROXIMATELY 30 CM, THE OPERATOR INADVERTENTLY CAUGHT THEIR HAND, CAUSING THE MICROCATHETER TO RETRACT SIGNIFICANTLY, DISLODGING IT FROM THE STENT AND RESULTING IN THE TIP MOVING FAR DISTALLY INTO THE BASILAR ARTERY, AWAY FROM THE ANEURYSM. SUBSEQUENT ATTEMPTS WERE MADE TO RE-NAVIGATE THE MICROCATHETER BACK INTO THE ANEURYSM, BUT THE COIL HAD STRETCHED, MAKING IT IMPOSSIBLE TO RETRACT THE COIL INTO THE MICROCATHETER. FURTHER ATTEMPTS TO RETRIEVE THE COIL USING A SNARE AND OTHER DEVICES WERE UNSUCCESSFUL, AS THE COIL BROKE DURING THE RETRIEVAL PROCESS AND ULTIMATELY COULD NOT BE RECOVERED. FINALLY, THE MICROCATHETER WAS REINSERTED THROUGH THE STENT STRUTS, WITH THE TAIL OF T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22631509 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K200030. One submission often covers many models, so check the product and lot details in each record.
- Z-0103-2023 · Sep 7, 2022 · Open, ClassifiedOptima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
- Z-1680-2022 · Jul 26, 2022 · Open, ClassifiedOptima Coil System Model OPTI0520CSF10. For endovascular embolization.
- Z-0660-2022 · Jan 4, 2022 · Open, ClassifiedThe Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil del
Other Device, Vascular, For Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.