Optima Coil System OPTI1345CSFMXRecalls & reported problems
Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (K200030); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jan 3, 2024Open, Classified
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
Z-1041-2024 · Root cause: Material/Component Contamination · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K200030. One submission often covers many models, so check the product and lot details in each record.
- Z-1988-2025 · May 8, 2025 · Open, ClassifiedOptima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
- Z-0103-2023 · Sep 7, 2022 · Open, ClassifiedOptima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
- Z-1680-2022 · Jul 26, 2022 · Open, ClassifiedOptima Coil System Model OPTI0520CSF10. For endovascular embolization.
- Z-0660-2022 · Jan 4, 2022 · Open, ClassifiedThe Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil del
Other Device, Vascular, For Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.