Skip to content

Optima Coil System OPTI1650CSFMXRecalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

All models of Optima Coil System (379)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053025341), received between Apr 16, 2026 and Apr 16, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Premature Separation” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

3 other recalls cover devices cleared under the same FDA submission (K200030); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 8, 2025Class IIOpen, Classified

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    Z-1988-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

  2. Jan 3, 2024Class IIOpen, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 16, 2026Premature Separation

    IT WAS REPORTED THAT: "DURING A COIL EMBOLIZATION OF A SPINAL AVM, THE PHYSICIAN ATTEMPTED TO DEPLOY AN OPTIMA COMPLEX-18 (20 × 65) AS THE SEVENTH COIL THROUGH A HEADWAY 27 MICROCATHETER. DURING DEPLOYMENT, WHILE ATTEMPTING TO REPOSITION, HE REPORTED FEELING A "POP" AND SUBSEQUENTLY LOST ONE-TO-ONE CONTROL OF THE COIL. MULTIPLE ATTEMPTS WERE MADE USING DIFFERENT GUIDEWIRES AND COIL PUSHERS TO ADVANCE THE REMAINING PORTION OF THE COIL OUT OF THE MICROCATHETER; HOWEVER, THESE ATTEMPTS WERE UNSUCCESSFUL. THE COIL WAS ULTIMATELY UNINTENTIONALLY SNARED BY A COIL PUSHER WITHIN THE MICROCATHETER, AND THE PHYSICIAN WAS ABLE TO SUCCESSFULLY REMOVE BOTH THE COIL AND MICROCATHETER. AFTER RE-ACCESSING WITH A HEADWAY DUO 156 CM MICROCATHETER, THE PHYSICIAN PROCEEDED WITH ADDITIONAL COIL DEPLOYMENT. DURING PLACEMENT OF AN OPTIMAX SOFT (16 × 50), HE AGAIN EXPERIENCED LOSS OF ONE-TO-ONE CONTROL, AND THE COIL WAS PREMATURELY DETACHED WITH GREATER THAN 50% OF THE COIL REMAINING WITHIN THE MICROCATHETER. ATTEMPTS TO ADVANCE THE COIL USING GUIDEWIRES WERE UNSUCCESSFUL. A SYRINGE WAS THEN ATTACHED TO THE HUB, AND NEGATIVE PRESSURE WAS APPLIED WHILE WITHDRAWING THE MICROCATHETER, RESULTING IN SUCCESSFUL

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24899848 ↗

FDA clearance and classification

Cleared through Special 510(k) K200030 on Feb 1, 2020 (decision: Substantially Equivalent). Submitted by Balt USA, LLC as “Optima Coil System”. The FDA review took 25 days. FDA record ↗

The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K200030. One submission often covers many models, so check the product and lot details in each record.

  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.
  • Z-0660-2022 · Jan 4, 2022 · Open, Classified
    The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil del

Common questions about the Optima Coil System OPTI1650CSFMX

Has the Optima Coil System OPTI1650CSFMX been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1988-2025 (Class II), began on May 8, 2025: Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Optima Coil System OPTI1650CSFMX?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 16, 2026. That report lists device problem “Premature Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Optima Coil System OPTI1650CSFMX?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Optima Coil System OPTI1650CSFMX FDA approved?

It was cleared through Special 510(k) K200030 on Feb 1, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Optima Coil System OPTI1650CSFMX MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Optima Coil System OPTI1650CSFMX, and is it still sold?

The FDA's device database lists Balt USA LLC as the company responsible for it (brand Optima Coil System, model OPTI1650CSFMX). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page