Optima Coil System OPTI0203CSS10 (REF N/A)Recalls & reported problems
Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant
All models of Optima Coil System (379)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00818053025785), received between Mar 25, 2021 and Aug 15, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Stretched” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
2 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 8, 2025Class IIOpen, Classified
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
Z-1988-2025 · Root cause: Vendor change control · Recall details · FDA record ↗
- Jan 3, 2024Class IIOpen, Classified
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Most reported problems
Device problems
- Stretched2
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 15, 2024Stretched; Stretched
IT WAS REPORTED THAT: "DURING THE PROCEDURE THE COIL WAS STRETCHED AND WAS TAKEN OUT OF THE BODY WITH SOLITAIRE STENT. TOTAL OF 3 OPTIMA COIL STRETCH DEFECTS OCCURRED IN THE PROCEDURE, SO 1 COIL AND 2 DELIVERY SYSTEM WILL BE RETURNED. 2 OF THE COILS WERE LOST DURING THE RETRIEVAL WITH THE SOLITAIRE STENT. IT WAS DIFFICULT TO VISUALLY DISTINGUISH THE RETURNED COIL IMPLANT FROM THE 3 STRETCH FAILURE, SO THE LABEL WILL BE ATTACHED." UPDATE 18JUL2024 - ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: " 1. WHEN EXACTLY DID THE STRETCHING OCCUR FOR EACH OF THE THREE COILS? - WHEN POSITIONING THE COIL BY INSERTING AND REMOVING. 2. WAS THE STENT DEVICE USED ONLY TO RETRACT ONE COIL? OR WAS A STENT DEVICE USED FOR EACH OF THE THREE STRETCHED COILS? - ONE STENT WAS USED TO REMOVE 3 COILS. 3. DID THE PATIENT SUSTAIN ANY INJURY? - NO 4. WHEN IT IS MENTIONED THAT "2 OF THE COILS WERE LOST DURING THE RETRIEVAL WITH THE SOLITAIRE STENT", WHAT WAS THE OUTCOME OF THIS PROCESS? WHAT HAPPENED TO THESE TWO COILS? - IT WAS DISCARDED WITH THE SOLITAIRE STENT." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19994840 ↗MalfunctionAug 15, 2024Stretched; Stretched
IT WAS REPORTED THAT: "DURING THE PROCEDURE THE COIL WAS STRETCHED AND WAS TAKEN OUT OF THE BODY WITH SOLITAIRE STENT. TOTAL OF 3 OPTIMA COIL STRETCH DEFECTS OCCURRED IN THE PROCEDURE, SO 1 COIL AND 2 DELIVERY SYSTEM WILL BE RETURNED. 2 OF THE COILS WERE LOST DURING THE RETRIEVAL WITH THE SOLITAIRE STENT. IT WAS DIFFICULT TO VISUALLY DISTINGUISH THE RETURNED COIL IMPLANT FROM THE 3 STRETCH FAILURE, SO THE LABEL WILL BE ATTACHED." UPDATE 18JUL2024: ADDITIONAL INFORMATION RECEIVED FROM THE ISSUER: "WHEN EXACTLY DID THE STRETCHING OCCUR FOR EACH OF THE THREE COILS? WHEN POSITIONING THE COIL BY INSERTING AND REMOVING. WAS THE STENT DEVICE USED ONLY TO RETRACT ONE COIL? OR WAS A STENT DEVICE USED FOR EACH OF THE THREE STRETCHED COILS? ONE STENT WAS USED TO REMOVE 3 COILS. DID THE PATIENT SUSTAIN ANY INJURY? NO WHEN IT IS MENTIONED THAT "2 OF THE COILS WERE LOST DURING THE RETRIEVAL WITH THE SOLITAIRE STENT", WHAT WAS THE OUTCOME OF THIS PROCESS? WHAT HAPPENED TO THESE TWO COILS? IT WAS DISCARDED WITH THE SOLITAIRE STENT." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19994452 ↗MalfunctionMar 25, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT: "THE PHYSICIAN WAS BALLOON ASSIST COILING A RUPTURED MCA BIRFURCATION ANEURYSM. THE PHYSICIAN HAD PREVIOUSLY USED A 2.5X6 AXIUM, THEN OPTIMA 2X6 CSS, OPTIMA 2X4 CSS. THE COIL MASS WASN'T SITTING WELL DUE TO WIDE NECK. THE PHYSICIAN USED THE BALLOON TO PUSH THE COILS BACK INTO THE ANUERYSM. THE PHYSICIAN ADDED AN ADDITIONAL 2X3 OPTIMA CSS AND IT PACKED AND HELD THE COILS IN PLACE. THE PHYSICIAN WAS USING AN ADDITIONAL 2X3 OPTIMA CSS. THE PHYSICIAN DIDN'T LIKE HOW THE COIL WAS SITTING AND DECIDED TO PULL OUT THE COIL. THE PHYSICIAN FELT SOME FRICTION WHEN REMOVING AND CONTINUED TO PULL ON PUSHER TO THE REMOVE THE COIL AND THE COIL BROKE OFF THE PUSHER AT THE DETACHMENT ZONE LEAVING THE COIL HALF IN THE MICRO CATHETER. THE PHYSICIAN USED A SYRINGE FILLED WITH SALINE TO PUSH THE COIL OUT OF MICROCATHETER. THE PHYSICIAN HAD TO USE AN ATLAS STENT TO SECURE THE COILS DOWN. THE PHYSICIAN BELIEVES HE HAD TRAPPED THE COIL ON THE SIDE WALL BETWEEN THE CATHETER AND BALLOON. THE PHYSICAN DOESN'T BELIEVE THE COIL WAS TANGLED. PUSHER AVAILABLE FOR RETURN, NOT COIL."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11571181 ↗
FDA clearance and classification
Cleared through 510(k) K172390 on Feb 18, 2018 (decision: Substantially Equivalent). Submitted by Blockade Medical, LLC (D.B.A. Balt Usa) as “Optima Coil System”. The FDA review took 194 days. FDA record ↗
The FDA regulates product code HCG (device, neurovascular embolization) under 21 CFR 882.5950 as a Class II device. 21 CFR 882.5950 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.
- Z-0103-2023 · Sep 7, 2022 · Open, ClassifiedOptima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
- Z-1680-2022 · Jul 26, 2022 · Open, ClassifiedOptima Coil System Model OPTI0520CSF10. For endovascular embolization.
Common questions about the Optima Coil System OPTI0203CSS10
Has the Optima Coil System OPTI0203CSS10 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1988-2025 (Class II), began on May 8, 2025: Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Optima Coil System OPTI0203CSS10?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Mar 25, 2021 and Aug 15, 2024. In those reports, the most common specific device problems are “Stretched” (2) and “Detachment of Device or Device Component” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Optima Coil System OPTI0203CSS10?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Optima Coil System OPTI0203CSS10 FDA approved?
It was cleared through 510(k) K172390 on Feb 18, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Optima Coil System OPTI0203CSS10 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Optima Coil System OPTI0203CSS10, and is it still sold?
The FDA's device database lists Balt USA LLC as the company responsible for it (brand Optima Coil System, model OPTI0203CSS10). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Device, Neurovascular Embolization devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page