Optima Coil System OPTI0206CSS10 (REF N/A)Recalls & reported problems
Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00818053025808), received between Apr 26, 2023 and Apr 18, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
2 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 8, 2025Open, Classified
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
Z-1988-2025 · Root cause: Vendor change control · FDA record ↗
- Jan 3, 2024Open, Classified
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
Z-1041-2024 · Root cause: Material/Component Contamination · FDA record ↗
Most reported problems
Device problems
- Fracture1
- Migration1
- Premature Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 18, 2025Migration; Premature Separation; Migration; Premature Separation
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
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21869279 ↗MalfunctionApr 26, 2023Fracture; Fracture
IT WAS REPORTED THAT: "THE XCEL BLINKED GREEN AND REG LIGHT IN THE BEGINNING, BUT TRYING AGAIN IT WORKED AND THE PROCEDURE WAS STARTED. AFTER INSERTION INTO THE ANEURYSM DETACH WAS TRIED BUT NO SOUND WAS DETECTED AND DID NOT DETACH. OTHER METHODS SUCH AS CHANGING THE XCEL WAS TRIED BUT DID NOT WORK, THEREFORE THE IMPLANT WAS TAKEN OUT AND CUT BY FORCE." 30MAR2023, ADDITIONAL INFORMATION RECEIVED BY THE ISSUER: "WE WERE TOLD THE IMPLANT FAILED TO DETACH, AND WAS REMOVED FROM THE ANEURYSM AND THE STRETCHED PART OF THE COIL WAS CUT WITH A SNARE. IT IS ASSUMED A PART OF THE COIL WAS REMOVED AND A PART REMAINED IN THE BODY." 04APR2023, ADDITIONAL INFORMATION RECEIVED BY THE ISSUER: "THE COIL WAS CAUGHT DURING REMOVAL, AND SNARE WAS USED TO REMOVE BY HOOKING BUT IT WAS STRETCHED AND CUT AND THROWN AWAY, THEREFORE IT IS DIFFICULT TO CHECK. WHEN REMOVING THE COIL FROM THE CATHETER FROM PULLING THE PUSHER, IT WAS CAUGHT AND COULD NOT BE REMOVED, THEREFORE A SNARE WAS USED. THE USER WAS ABLE TO CONFIRM THE TOTALITY OF THE IMPLANT COIL WAS CAPTURED WITHIN THE MICROCATHETER THROUGH THE RADIOPAQUE MARKER IN THE VIDEO (THE USER IS EXPERIENCED WITH THE USE OF THE DEVICE). THE SNARE STRETCHED THE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16818254 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.
- Z-0103-2023 · Sep 7, 2022 · Open, ClassifiedOptima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
- Z-1680-2022 · Jul 26, 2022 · Open, ClassifiedOptima Coil System Model OPTI0520CSF10. For endovascular embolization.
Other Device, Neurovascular Embolization devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.