Skip to content

Optima Coil System OPTI0206CSS10 (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

All models of Optima Coil System (379)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00818053025808), received between Apr 26, 2023 and Apr 18, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

2 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 8, 2025Class IIOpen, Classified

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    Z-1988-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

  2. Jan 3, 2024Class IIOpen, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture1
  • Migration1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 18, 2025Migration; Premature Separation; Migration; Premature Separation

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

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21869279 ↗
  • MalfunctionApr 26, 2023Fracture; Fracture

    IT WAS REPORTED THAT: "THE XCEL BLINKED GREEN AND REG LIGHT IN THE BEGINNING, BUT TRYING AGAIN IT WORKED AND THE PROCEDURE WAS STARTED. AFTER INSERTION INTO THE ANEURYSM DETACH WAS TRIED BUT NO SOUND WAS DETECTED AND DID NOT DETACH. OTHER METHODS SUCH AS CHANGING THE XCEL WAS TRIED BUT DID NOT WORK, THEREFORE THE IMPLANT WAS TAKEN OUT AND CUT BY FORCE." 30MAR2023, ADDITIONAL INFORMATION RECEIVED BY THE ISSUER: "WE WERE TOLD THE IMPLANT FAILED TO DETACH, AND WAS REMOVED FROM THE ANEURYSM AND THE STRETCHED PART OF THE COIL WAS CUT WITH A SNARE. IT IS ASSUMED A PART OF THE COIL WAS REMOVED AND A PART REMAINED IN THE BODY." 04APR2023, ADDITIONAL INFORMATION RECEIVED BY THE ISSUER: "THE COIL WAS CAUGHT DURING REMOVAL, AND SNARE WAS USED TO REMOVE BY HOOKING BUT IT WAS STRETCHED AND CUT AND THROWN AWAY, THEREFORE IT IS DIFFICULT TO CHECK. WHEN REMOVING THE COIL FROM THE CATHETER FROM PULLING THE PUSHER, IT WAS CAUGHT AND COULD NOT BE REMOVED, THEREFORE A SNARE WAS USED. THE USER WAS ABLE TO CONFIRM THE TOTALITY OF THE IMPLANT COIL WAS CAPTURED WITHIN THE MICROCATHETER THROUGH THE RADIOPAQUE MARKER IN THE VIDEO (THE USER IS EXPERIENCED WITH THE USE OF THE DEVICE). THE SNARE STRETCHED THE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16818254 ↗

FDA clearance and classification

Cleared through 510(k) K172390 on Feb 18, 2018 (decision: Substantially Equivalent). Submitted by Blockade Medical, LLC (D.B.A. Balt Usa) as “Optima Coil System”. The FDA review took 194 days. FDA record ↗

The FDA regulates product code HCG (device, neurovascular embolization) under 21 CFR 882.5950 as a Class II device. 21 CFR 882.5950 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.

  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

Common questions about the Optima Coil System OPTI0206CSS10

Has the Optima Coil System OPTI0206CSS10 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1988-2025 (Class II), began on May 8, 2025: Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Optima Coil System OPTI0206CSS10?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Apr 26, 2023 and Apr 18, 2025. In those reports, the most common specific device problems are “Fracture” (1), “Migration” (1) and “Premature Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Optima Coil System OPTI0206CSS10?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Optima Coil System OPTI0206CSS10 FDA approved?

It was cleared through 510(k) K172390 on Feb 18, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Optima Coil System OPTI0206CSS10 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Optima Coil System OPTI0206CSS10, and is it still sold?

The FDA's device database lists Balt USA LLC as the company responsible for it (brand Optima Coil System, model OPTI0206CSS10). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page