Optima Coil System OPTI0256CSS10 (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

All models of Optima Coil System (379)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00818053025853), received between Mar 10, 2022 and Jan 31, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

2 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 8, 2025Class IIOpen, Classified

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    Z-1988-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

  2. Jan 3, 2024Class IIOpen, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Premature Separation2
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 31, 2025Nonstandard Device; Nonstandard Device

    ON DECEMBER 15, 2024, BALT USA RECEIVED COMPLAINT: (B)(4) (LOT: F230800685) FROM (B)(6). IT WAS REPORTED, "DURING DEPLOYMENT OF THE COIL IN THE ANEURYSM, THE RADIOPAQUE MARKER OF THE DETACH ZONE COULD NOT BE SEEN USING ANGIOGRAPHY. AFTER MOVING SEVERAL TIMES, IT COULD NOT BE SEEN AND WAS REMOVED." THE RETURNED UNIT WAS ANALYZED IN-HOUSE AND THEN SENT TO A SENT FOR MATERIAL ANALYSIS TO AN EXTERNAL MATERIALS CHARACTERIZATION LAB FOR FURTHER TESTING. ON JANUARY 3, 2025, THE ANALYSIS FROM THE MATERIALS CHARACTERIZATION LAB WAS RECEIVED WHICH CONCLUDED THE MATERIALS OF THE RO MARKER ALONG THE DELIVERY PUSHER ARE CONSISTENT WITH A 300 SERIES STAINLESS STEEL. THIS DATA SUPPORTS THE MATERIAL BEING THE CAUSE FOR THE VISIBILITY ISSUE DESCRIBED IN COMPLAINT#: (B)(4) CONFIRMING THE RO MARKER WAS NOT MADE WITH THE INTENDED WIRE MATERIAL TO BE ABLE TO EASILY VISUALIZE THE RADIOPAQUE SECTION OF THE DELIVERY PUSHER. FOR THIS REASON, BALT USA INITIATED HHE-031 ON 03JAN2025 FOR THIS ISSUE TO FULLY ASSESS THE IMPACT OF THE SUPPLIER'S HYPOCOIL SUBASSEMBLY BEING MANUFACTURED USING THE UNINTENDED COIL WIRE MATERIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21289088 ↗
  • MalfunctionAug 18, 2023Premature Separation; Premature Separation

    IT WAS REPORTED THAT: "WHEN THE 0256CSS COIL WAS INSERTED INTO THE ANEURYSM, IT SUDDENLY DETACHED AND AROUND HALF WAS FLOATING AND BLOCKING THE BLOOD VESSEL. THE PHYSICIAN PUSHED IN A 0154CSS AND TOOK EMERGENCY MEASURES, BUT THE BLOOD VESSEL FROM THE OUTSIDE OF THE ANEURYSM TO THE PCOM WAS BLOCKED. THEREFORE, AFTER REMOVING THE 0154CSS, IT WAS RE-ADJUSTED. FORTUNATELY, IT WAS JUDGED THAT THERE WAS NO SIGNIFICANT HARM TO THE PATIENT AND THE PROCEDURE WAS CLOSED." 09NOV2022, ADDITIONAL INFORMATION PROVIDED BY ISSUER - "·REGARDING THE UNIT UNDER REFERENCE ((B)(6)): WAS THE 0154CSS COIL A PART OF THE INITIAL PLANNED PROCEDURE OR WAS THIS A PHYSICIAN INTERVENTION? (B)(6)'S ANSWER: BOTH ARE CORRECT. THE DOCTOR PLANNED ON USING IT AFTER 0256CSS, BUT SINCE THE 0256CSS WAS PRE-DETACHED 0154CSS WAS USED. WAS A DETACHMENT CYCLE ATTEMPTED FOR THE 0256CSS COIL? TAEWOONG'S ANSWER: THE PROBLEM OCCURRED BECAUSE IT WAS DETACHED WITHOUT DETACHMENT ATTEMPT. CAN YOU CONFIRM THAT THE 0154CSS COIL WAS NOT DETACHED IN THE ANEURYSM? TAEWOONG'S ANSWER: THERE WERE NO PROBLEMS WITH 0154CSS COIL. 0256CSS WAS NOT DETACHED, AND COULD NOT BE COMPLETELY PUSHED WITH 0154CSS, SO IT WAS SLIGHTLY PROTRUDED INTO THE B

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17573384 ↗
  • MalfunctionMar 10, 2022Premature Separation; Premature Separation

    IT WAS REPORTED THAT: WHEN THE PHYSICIAN TRIED TO INSERT OPTIMA AS SIXTH COIL IN STENT ASSISTED PROCEDURE, THE COIL PROTRUSION INTO THE STENT LUMEN SO THE PHYSICIAN TRIED TO REPOSITION IT. THEN HE FELT RESISTANCE. HE FOUND THE COIL WAS DETACHED. COIL WAS RETRIEVED USING SNARE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13721338 ↗

FDA clearance and classification

Cleared through 510(k) K172390 on Feb 18, 2018 (decision: Substantially Equivalent). Submitted by Blockade Medical, LLC (D.B.A. Balt Usa) as “Optima Coil System”. The FDA review took 194 days. FDA record ↗

The FDA regulates product code HCG (device, neurovascular embolization) under 21 CFR 882.5950 as a Class II device. 21 CFR 882.5950 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.

  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

Common questions about the Optima Coil System OPTI0256CSS10

Has the Optima Coil System OPTI0256CSS10 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1988-2025 (Class II), began on May 8, 2025: Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Optima Coil System OPTI0256CSS10?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Mar 10, 2022 and Jan 31, 2025. In those reports, the most common specific device problems are “Premature Separation” (2) and “Nonstandard Device” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Optima Coil System OPTI0256CSS10?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Optima Coil System OPTI0256CSS10 FDA approved?

It was cleared through 510(k) K172390 on Feb 18, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Optima Coil System OPTI0256CSS10 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Optima Coil System OPTI0256CSS10, and is it still sold?

The FDA's device database lists Balt USA LLC as the company responsible for it (brand Optima Coil System, model OPTI0256CSS10). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Neurovascular Embolization devices