Optima Coil System OPTI0306CSS10 (REF N/A)Recalls & reported problems
Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant
All models of Optima Coil System (379)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00818053025884), received between Aug 18, 2023 and Jan 24, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jan 3, 2024Class IIOpen, Classified
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Most reported problems
Device problems
- Contamination /Decontamination Problem1
- Insufficient Device Problem Information1
- Migration1
- Poor Visibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 24, 2025Migration; Migration
IT WAS REPORTED THAT: "THE PHYSICIAN WAS TREATING A PCOM ANEURYSM CASE. AFTER SUCCESSFULLY DEPLOYING FEW COILS FROM HEADWAY 17,. HE TOOK THIS 3*6 OPTIMA COIL AND PLACE IT INSIDE THE Y CONNECTOR AND WAITED FOR SALINE TO COME OUT OF THE SHEATH, ONCE THE DROPLETS COME OUT FROM THE SHEATH, THEN HE START DEPLOYING THE COIL INSIDE THE ANEURYSM AND AFTER SUCCEFUL DEPLOYMENT HE DETACHED THE COIL AD REMOVE THE PUSHER WIRE FROM THE MICRO CATHETER AND TOOK A FLOURO RUN THAT TIME HE FOUNDS THERE IS SOMETHING LYING NEAR THE NECK OF THE ANEURYSM AFTER TAKING FEW MORE SHOTS HE REALIZED THAT IT IS THE PART OF THE DETACHED PUSHER WIRE THEN HE HAD TO DEPLOY A STENT OVER IT AND COMPLETED THE CASE." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21230679 ↗MalfunctionNov 2, 2023Contamination /Decontamination Problem; Insufficient Device Problem Information; Contamination /Decontamination Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT: "<PATIENT INFORMATION> SEX: UNKNOWN. DISEASE NAME: AN. PROCEDURE: PLANED TAE. MICRO CATHETER: UNKNOWN. USED COILS: UNKNOWN. <DESCRIPTION> WHEN THE PHYSICIAN REMOVED PACKAGE OF OPTIMA AND TRIED TO ADVANCED IMPLANT COIL AS BEFORE USE CHECK, THE PHYSICIAN FOUND BROWN SUBSTANCE IN THE LUMEN OF INTRODUCER SHEATH AND THE COIL WOULD NOT ADVANCE. THEN THE PHYSICIAN OPENED NEW OPTIMA WITH SAME SIZE TO COMPLETE PROCEDURE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18056174 ↗MalfunctionAug 18, 2023Poor Visibility; Poor Visibility
IT WAS REPORTED THAT: PROCEDURE: EMERGENCY TAE FOR AN MICRO CATHETER: SL10/STRYKER OTHER COILS: TARGET/STRYKER, OPTIMA 3×8. <DESCRIPTION> AFTER PLACING OPTIMA 3×8, THE PHYSICIAN ADVANCED COIL INTO MICRO CATHETER. HE INSISTS THAT HE THINKS THE POSITION OF FLUOR MARKER WAS NOT CORRECT (IT WAS DISPLACED TO THE PROXIMAL SIDE.) SINCE WHEN HE STARTED FLUOROSCOPY OF TARGET LESION, HE FOUND THAT THE IMPLANT COIL HAD ALREADY ENTERED THE ANEURYSM DESPITE HE STOPPED ADVANCEMENT AT THE POINT OF THE FLUOR MARKER REACHED Y CONNECTOR. THE PROCEDURE WAS COMPLETED EXCHANGING TO ANOTHER OPTIMA OF SAME SIZE." INCIDENT REPORT FORM RECEIVED FOR THIS COMPLAINT INDICATED NO PATIENT INJURY WAS SUSTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17573028 ↗
FDA clearance and classification
Cleared through 510(k) K172390 on Feb 18, 2018 (decision: Substantially Equivalent). Submitted by Blockade Medical, LLC (D.B.A. Balt Usa) as “Optima Coil System”. The FDA review took 194 days. FDA record ↗
The FDA regulates product code HCG (device, neurovascular embolization) under 21 CFR 882.5950 as a Class II device. 21 CFR 882.5950 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.
- Z-1988-2025 · May 8, 2025 · Open, ClassifiedOptima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
- Z-0103-2023 · Sep 7, 2022 · Open, ClassifiedOptima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
- Z-1680-2022 · Jul 26, 2022 · Open, ClassifiedOptima Coil System Model OPTI0520CSF10. For endovascular embolization.
Common questions about the Optima Coil System OPTI0306CSS10
Has the Optima Coil System OPTI0306CSS10 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1041-2024 (Class II), began on Jan 3, 2024: Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Optima Coil System OPTI0306CSS10?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Aug 18, 2023 and Jan 24, 2025. In those reports, the most common specific device problems are “Contamination /Decontamination Problem” (1), “Migration” (1) and “Poor Visibility” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Optima Coil System OPTI0306CSS10?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Optima Coil System OPTI0306CSS10 FDA approved?
It was cleared through 510(k) K172390 on Feb 18, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Optima Coil System OPTI0306CSS10 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Optima Coil System OPTI0306CSS10, and is it still sold?
The FDA's device database lists Balt USA LLC as the company responsible for it (brand Optima Coil System, model OPTI0306CSS10). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Device, Neurovascular Embolization devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page