Skip to content

Optima Coil System OPTI0310CSS10 (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

All models of Optima Coil System (379)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00818053025907), received between Apr 8, 2021 and Aug 11, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

2 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. May 8, 2025Class IIOpen, Classified

    Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.

    Z-1988-2025 · Root cause: Vendor change control · Recall details · FDA record ↗

  2. Jan 3, 2024Class IIOpen, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Migration1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Stroke/CVA1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2023Stretched; Stretched

    IT WAS REPORTED THAT: "PATIENT INFORMATION: SEX: UNKNOWN. DISEASE NAME: DAVF. PROCEDURE: PLANNED PROCEDURE FOR DAVF. MICRO CATHETER: UNKNOWN. GUIDE WIRE: UNKNOWN. USED COILS:10 THE OTHER COILS INCLUDING COMPETITIVE PRODUCT. DESCRIPTION: AFTER DETACHMENT OF THE OPTIMA, THE PHYSICIAN RETRIEVED DELIVERY PUSHER AND FOUND THAT THE FLUOROSCOPY SUGGESTED THE POSSIBILITY OF UNRAVELING THE IMPLANT COIL IN TARGET LESION. THE PHYSICIAN MANAGED TO RETRIEVE UNRAVELED COIL WITH SNARE. THE PROXIMAL END OF RETRIEVED COIL WAS CERTAINLY UNRAVELED. ALTHOUGH 10 OTHER COILS INCLUDING COMPETITIVE DEVICE HAD BEEN PLACED THE TARGET LESION AT THAT POINT (THE COMPLAINT OPTIMA WAS 11TH COIL IN THE PROCEDURE), THE PHYSICIAN DETERMINED THAT THE RETRIEVED COIL LOOKS OPTIMA. THE PROCEDURE WAS FINISHED WITHOUT ANY FURTHER TREATMENT. THE PHYSICIAN ALSO COMMENTED THAT IT MIGHT BE CAUSED BY TOO EARLY WITHDRAWAL OF DELIVERY PUSHER WHICH MIGHT HAVE HOOKED AND DRAGGED IMPLANT COIL. THUS, THE PHYSICIAN DOES NOT ASSUME IT PRODUCT DEFECT AND THE FACILITY DISCARDED COIL SYSTEM." INCIDENT REPORT FORM SUBMITTED FOR THIS EVENT INDICATED NO PATIENT INJURY WAS SUSTAINED AS A RESULT OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17515365 ↗
  • MalfunctionApr 8, 2021Detachment of Device or Device Component; Migration; Detachment of Device or Device Component; Migration

    IT WAS REPORTED THAT: "I WAS JUST INFORMED OF THE ISSUE IN THE CASE TODAY. THEPHYSICIAN WAS REPAIRING A RUPTURED LEFT PCOMM ANEURYSM AND WAS PUTTING IN HIS FOURTH OPTIMA COIL A 3X10 CSF. HE WAS TAKING THE COIL IN AND OUT TO REPOSITION WHEN HE DECIDED COIL WASN'T GOING WHERE HE WANTED AND BEGAN PULLING COIL OUT AND IT PREMATURE DETACHED IN THE CATHETER. THE COIL INADVERTENTLY DETACHED PARTIALLY WITHIN THE CATHETER AND THE ANUERYSM, BASICALLY THE DETACHMENT ZONE BROKE. HE LOST THE COIL OUT THE FRONT OF THE CATHETER WHEN HE WENT TO REMOVE AND THE COIL FLEW UP THE MCA AND CAUSED A BLOCKAGE IN THE ARTERY. HE WAS ABLE TO GET PAST THE COIL AND STENT IT ALONG THE VESSEL IN PLACE. PATIENT HAD SOME STROKE LIKE SYMPTOMS BUT HAS RECOVERED."

    Patient problems: No Clinical Signs, Symptoms or Conditions; Stroke/CVA

    MAUDE report 11638875 ↗

FDA clearance and classification

Cleared through 510(k) K172390 on Feb 18, 2018 (decision: Substantially Equivalent). Submitted by Blockade Medical, LLC (D.B.A. Balt Usa) as “Optima Coil System”. The FDA review took 194 days. FDA record ↗

The FDA regulates product code HCG (device, neurovascular embolization) under 21 CFR 882.5950 as a Class II device. 21 CFR 882.5950 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.

  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

Common questions about the Optima Coil System OPTI0310CSS10

Has the Optima Coil System OPTI0310CSS10 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1988-2025 (Class II), began on May 8, 2025: Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Optima Coil System OPTI0310CSS10?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Apr 8, 2021 and Aug 11, 2023. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (1), “Migration” (1) and “Stretched” (1); the most common patient problem is “Stroke/CVA” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Optima Coil System OPTI0310CSS10?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Optima Coil System OPTI0310CSS10 FDA approved?

It was cleared through 510(k) K172390 on Feb 18, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Optima Coil System OPTI0310CSS10 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Optima Coil System OPTI0310CSS10, and is it still sold?

The FDA's device database lists Balt USA LLC as the company responsible for it (brand Optima Coil System, model OPTI0310CSS10). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page