Optima Coil System OPTI0615CSF10Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

All models of Optima Coil System (379)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00818053026188), received between Mar 7, 2025 and Jun 18, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Migration” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K200030); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 3, 2024Open, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Migration2
  • Difficult or Delayed Separation1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 18, 2025Migration; Difficult or Delayed Separation; Migration; Difficult or Delayed Separation

    IT WAS REPORTED THAT: "THE PHYSICIAN WAS PERFORMING AN EMBOLIZATION TO TREAT A RECANALIZED ANEURYSM IN THE TERMINUS OF THE ICA. PHYSICIAN PREVIOUSLY USED STRYKER COILS TO FILL ANEURYSMAL SPACE ALONG WITH Y STENTS. DUE TO THE NATURE OF THE FLOW, THE COILS MOVED TO THE DOME OF THE ANEURYSM AND THE PHYSICIAN NEEDED TO TREAT THE NECK OF THE ANEURYSM. HE BEGAN WITH TWO 9X30 TARGET XL COILS. BOTH COILS FORMED INSIDE THE ANEURYSM BUT IT WAS DIFFICULT TO VIEW COILS FORMING DUE TO THE COIL MASS THAT WAS BLOCKING IT. THE PHYSICIAN THEN USED A TWO BALT COILS - FIRST AND OPTIMAX AND THEN AN OPTIMA. BOTH COILS PACKED NICELY, AND HE ASKED FOR A TETRA COIL BUT THE REP ADVISED HIM AGAINST IT. THEY DISCUSSED FINISHING WITH A STRYKER COIL, BUT THE PHYSICIAN CHOSE TO USE AN OPTIMA 6X15 AS THE FINAL COIL AT THE NECK. PHYSICIAN WAS INSTRUCTED MULTIPLE TIMES TO TEST COIL BEFORE INSERTING INTO PATIENT. HE ADVANCED THE COIL THROUGH THE 8 FR SHEATH/TRACSTAR/SOFIA/SL 10 AND BEGAN TO DEPLOY COIL INSIDE THE COIL MASS. THE COIL KEPT LOOPING INTO ANOTHER BRANCH AND DR. PULLED BACK SLIGHTLY ON MICROCATHETER TO RETRACT COIL SLIGHTLY AND READVANCED. THE COIL THEN CONTINUED GOING INTO ANOTHER ARTERIAL BRANCH, SO HE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22286185 ↗
  • MalfunctionMar 7, 2025Stretched; Migration; Stretched; Migration

    IT WAS REPORTED THAT: "ACOM ANEURYSM TREATMENT WAS PLANNED USING SL-10 MICROCATHETER AND TRAXCESS-14 WIRE. SL-10 WAS PARKED INTO THE ANEURYSM OVER TRAXCESS-14 WIRE. FIRST OPTIMA COMPLEX-10 SOFT 6/15 WAS PREPARED AS PER IFU. DURING DEPLOYMENT OF COIL, ONE LOOP OF COIL CAME OUT OF THE ANEURYSM, SO OPERATOR TRIED TO REPOSITION IT, BUT DURING REPOSITION IT WAS OBSERVED THAT THE COIL WAS NOT GOING FORWARD, SO OPERATOR DECIDED TO TAKE IT BACK FROM SYSTEM AND IT WAS OBSERVED THAT THE COIL GOT STRETCHED. THUS, THE COIL WAS REMOVED FROM THE SYSTEM. THE CASE WAS COMPLETED USING 5 MORE OPTIMA COILS WITHOUT ANY DIFFICULTY." INCIDENT REPORT FORM SUBMITTED FOR THIS COMPLAINT INDICATES THAT NO PATIENT INJURY WAS SUSTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21548731 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K200030. One submission often covers many models, so check the product and lot details in each record.

  • Z-1988-2025 · May 8, 2025 · Open, Classified
    Optima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.
  • Z-0660-2022 · Jan 4, 2022 · Open, Classified
    The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil del

Other Device, Vascular, For Promoting Embolization devices

All Device, Vascular, For Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page