Skip to content

Optima Coil System OPTI0816CST10 (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053026355), received between Dec 23, 2021 and Dec 23, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 3, 2024Open, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · FDA record ↗

Most reported problems

Device problems

  • Migration1

Patient problems

  • Ischemia Stroke1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 23, 2021Migration; Migration

    IT WAS REPORTED THAT: "PREMATURE COIL DETACHMENT INSIDE MICROCATHETER". UPDATE 24NOV2021: "LEFT SYLVIAN ISCHEMIC STROKE HAS BEEN OBSERVED. DEFICIT OF RIGHT ARM WITH QUASI TOTAL RECOVERY". FOLLOW UP 25NOV2021: "HAS THE PATIENT RECOVERED FROM THE STROKE? FOR THE TIME BEING, THE STATUS OF THE RIGHT ARM DEFICIT IS: ALMOST COMPLETELY REGRESSIVE. IS IT KNOWN IF THE MALFUNCTION OF THE DEVICE (PREMATURE DETACHMENT) IS WHAT CAUSED THE STROKE? WERE THERE ANY PROCEDURAL COMPLICATIONS THAT COULD HAVE LED TO THE STROKE? AS THE COIL MIGRATED INTO SYLVIAN BRANCH AND WAS NOT ABLE TO BE NORMALLY RETRIEVE -BECAUSE DETACHED- IT HAS BEEN LONG AND DIFFICULT TO RETRIEVE IT WITH DEDICATED DEVICES, THERE IS A LINK BETWEEN PREMATURE DETACHMENT AND STROKE. THE FORM INDICATES THAT THE INCIDENT TOOK PLACE ON (B)(6) 2021. WHY WAS THIS ONLY JUST RECENTLY REPORTED TO BALT USA, DID THE HOSPITAL ONLY JUST NOW REPORT IT OR WAS NOTIFICATION RECEIVED FROM THE ANSM NOTIFICATION? THE COMPLICATION IN RELATION WITH THE PREMATURE DETACHMENT OF THE COIL HAS BEEN REPORTED TO US ONLY YESTERDAY. UPDATE 14DEC2021 - ADDITIONAL INFORMATION RECEIVED FROM ISSUER: "THE STATUS IS: SLIGHT RESIDUAL DEFICIT AT THE RIGHT HAND. PR PIEROT

    Patient problems: Ischemia Stroke

    MAUDE report 13078223 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.

  • Z-1988-2025 · May 8, 2025 · Open, Classified
    Optima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.