Skip to content

Optima Coil System OPTI0930CST10 (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

All models of Optima Coil System (379)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053026386), received between Nov 3, 2020 and Nov 3, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jan 3, 2024Class IIOpen, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Physical Resistance/Sticking1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2020Break; Separation Failure; Physical Resistance/Sticking

    IT WAS REPORTED: "THE PHYSICIAN WAS DOING A VENOUS EMBO OF A DAVF WHERE HE WAS COILING OFF THE TRANSVERSE SINUS. HIS 37TH COIL WAS A 9X30 COMPLEX STANDARD. THE CATHETER HAD LOTS OF FORWARD TENSION AND HE WAS PACKING IN A TIGHT SPOT. HE ATTEMPTED TO DETACH AND THE COIL DIDN'T RELEASE AFTER FIRST CYCLE. HE TRIED A SECOND CYCLE AND XCEL WAS GIVING CHRISTMAS TREE LIGHTS AND THE COIL DIDN'T DETACH. HE TRIED A THIRD TIME AND PUSHED THE COIL AND PUSHER WELL PAST THE PROXIMAL LINE AND IT STILL DIDN'T POP OFF THE PUSHER. HE TRIED DETACHING A FOURTH TIME. HE THEN TRIED TO REMOVE THE COIL AND IT WAS LODGED IN THE COIL MASS. HE PULLED UNTIL THE COIL BROKE AND SEPARTED. THE COIL AND PUSHER STRETCHED AS HE REMOVED. HE THEN HAD TO REMOVE THE CATHETER AND THERE WAS COIL AND PUSHER STRETCHED OUT OF THE SHEATH. HE PULLED AND BROKE OFF MORE OF THE COIL AND PUSHER. HE HAD TO REMOVE THE SHEATH BECAUSE THERE WAS COIL AND PUSHER DOWN THE SHEATH. HE PULLED ANOTHER SEGMENT OF COIL OUT UNTIL IT BROKE. HE PLACED ANOTHER SHEATH AND CATHETER AND WAS ABLE TO FINISH THE CASE WITHOUT HARM TO THE PATIENT. HE USED 10 MORE OPTIMAS WITHOUT ANY INCIDENT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10778881 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.

  • Z-1988-2025 · May 8, 2025 · Open, Classified
    Optima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page