Optima Coil System OPTI0208HSS10 (REF N/A)Recalls & reported problems
Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant
All models of Optima Coil System (379)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053026706), received between Jul 20, 2023 and Jul 20, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Premature Separation” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
2 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- May 8, 2025Open, Classified
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
Z-1988-2025 · Root cause: Vendor change control · Recall details · FDA record ↗
- Jan 3, 2024Open, Classified
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Most reported problems
Device problems
- Premature Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 20, 2023Premature Separation; Premature Separation
IT WAS REPORTED THAT: "IT WAS A CASE OF ACOM ANEURYSM MEASURING 5.7MM*5MM WITH NECK OF 2 MM THE PHYSICAN PLANNED TO DO SIMPLE COILING. THE SUPPORT SYSTEM WAS VISTA BRITE 90CM AND FARGOMAX, THE VASCO10D WAS PLACED INSIDE THE ANEURYSM WITH SUPPORT OF HYBRID1214D. THE VASCO10D WAS CONNECTED WITH A FLUSH LINE. THE COIL WAS PREPARED AS PER IFU AND THEN ALIGNED TO THE HUB OF THE MICRO CATHETER, THE FIRST COIL SIZE OF 5MM*13MM OPTIMA COMPLEX10 WAS TAKEN ON INTRODUZING, THE COIL WENT SMOOTHLY THROUGH THE SLEEVE INTO THE HUB OF VASCO10 AND DETACHED LATER TOOK 4MM*13MM OPTIMA COMPLEX10 AND DETACHED NEXT PHYSICIAN TOOK 2MM*8MM OPTIMA HELICAL SUPERSOFT THE COIL WENT SMOOTHLY THROUGH THE SLEEVE INTO THE HUB. PHYSICIAN WAS PUSHING THE COIL INTO ANEURYSM BUT THE COIL WAS NOT MOVING LATER PHYSICIAN WAS RETERVING COIL BUT NO MOVEMENT WAS HAPPENING. COIL HAD PRE DETACHED BUT THE COIL WAS REMOVED WITH THE HELP OF CATCHMAXI40 COIL BEEN TAKEN OUT AND PHYSICIAN DECIDED TO LEAVE AND CASE WAS COMPLETED WITHOUT ISSUE." INCIDENT REPORT FORM SUBMITTED FOR THIS EVENT INDICATED NO PATIENT INJURY WAS SUSTAINED AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17364144 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.
- Z-0103-2023 · Sep 7, 2022 · Open, ClassifiedOptima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
- Z-1680-2022 · Jul 26, 2022 · Open, ClassifiedOptima Coil System Model OPTI0520CSF10. For endovascular embolization.
Other Device, Neurovascular Embolization devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page