Optima Coil System OPTI0510HSS10 (REF N/A)Recalls & reported problems

Balt USA LLC · The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller. · Class 2 · Implant

All models of Optima Coil System (379)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818053026836), received between Oct 7, 2022 and Oct 7, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K172390); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Jan 3, 2024Open, Classified

    Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

    Z-1041-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 7, 2022Fracture; Stretched; Fracture; Stretched

    IT WAS REPORTED THAT: "AFTER NON DETACHMENT OF HSS 5X10, DURING INSERTION OF THE SAME SIZED 5X10, IT WAS PUSHED AND PULLED 2~3 TIMES AS IT COULD NOT BE INSERTED PROPERLY. WHEN CHECKED IT WAS FOUND THAT IT WAS STRETCHED, AND SOLITAIRE STENT RETRIEVER WAS USED FOR PARTIAL RETRIEVAL OF THE STRETCHED COIL. THE REST WAS ATTACHED TO THE VESSEL WALL USING A DOMESTIC INTRACRANIAL STENT.". 29SEP2022, ADDITIONAL INFORMATION PROVIDED BY ISSUER: "O CAN YOU CONFIRM THAT THERE WAS NOT HARM/ INJURY SUSTAINED BY THE PATIENT DURING/ POST-PROCEDURE? WHAT IS THE CURRENT STATUS OF THE PATIENT? - THERE WAS NO HARM / INJURY SUSTAINED BY THE PATIENT. O WAS THE ANATOMY TORTUOUS? - WE WERE INFORMED THE ANATOMY WAS TORTUOUS.".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15565009 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172390. One submission often covers many models, so check the product and lot details in each record.

  • Z-1988-2025 · May 8, 2025 · Open, Classified
    Optima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
  • Z-0103-2023 · Sep 7, 2022 · Open, Classified
    Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
  • Z-1680-2022 · Jul 26, 2022 · Open, Classified
    Optima Coil System Model OPTI0520CSF10. For endovascular embolization.

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →