Prestige Coil System PRES0420HPKPLRecalls & reported problems
Balt USA LLC · The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature. It is designed for use with the Ultra Detachment Controller. · Class 2 · Implant
All models of Prestige Coil System (97)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
5 other recalls cover devices cleared under the same FDA submission (K200030); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 8, 2025Class IIOpen, Classified
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
Z-1989-2025 · Root cause: Vendor change control · Recall details · FDA record ↗
- Jan 3, 2024Class IIOpen, Classified
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
Z-1042-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K200030. One submission often covers many models, so check the product and lot details in each record.
- Z-1988-2025 · May 8, 2025 · Open, ClassifiedOptima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI020
- Z-1041-2024 · Jan 3, 2024 · Open, ClassifiedNeurovascular embolization device. The Optima Coil System is a series specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracra
- Z-0103-2023 · Sep 7, 2022 · Open, ClassifiedOptima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.
- Z-1680-2022 · Jul 26, 2022 · Open, ClassifiedOptima Coil System Model OPTI0520CSF10. For endovascular embolization.
- Z-0660-2022 · Jan 4, 2022 · Open, ClassifiedThe Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil del
Other Device, Vascular, For Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page