Aristotle 24-200-002Recalls & reported problems

Scientia Vascular LLC · Guidewire-Standard · Class 2

All models of Aristotle (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
2
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00818075010134), received between Jul 30, 2024 and Feb 11, 2026 — classified as 2 deaths, 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Break” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 6 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break9

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Foreign Body In Patient3
  • Obstruction/Occlusion2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2026Break

    IF WAS REPORTED THAT DURING A NEUROVASCULAR PROCEDURE AN ARISTOTLE 24 GUIDEWIRE WAS DEFECTIVE. THERE WAS NO PATIENT INJURY AS RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24319495 ↗
  • MalfunctionFeb 11, 2026Break

    IT WAS REPORTED THAT DURING A PROCEDURE AN ARISTOTLE 24 GUIDEWIRE WAS DEFECTIVE AND THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24319417 ↗
  • MalfunctionFeb 11, 2026Break

    DURING A PROCEDURE, THE PHYSICIAN WAS USING THE GUIDEWIRE WHEN THEY NOTICED THERE WAS A FRACTURE OF THE DISTAL TIP UNDER FLUOROSCOPY AND REMOVED THE WIRE WITH NO PATIENT INJURY. THE PROCEDURE WAS COMPLETED WITH NO INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24319398 ↗
  • MalfunctionFeb 10, 2026Break

    DURING A NEUROVASCULAR PROCEDURE THE PHYSICIAN WAS USING THE GUIDEWIRE AND WHEN THEY REMOVED THE WIRE FROM THE PATIENT AFTER BEING INTRACRANIAL THE WIRE WAS DAMAGED. THE GUIDEWIRE MALFUNCTION OCCURRED INSIDE OF THE PATIENT AND NO INTERVENTION WAS REQUIRED TO RETRIEVE ANY PORTION OF THE DAMAGED DEVICE. THERE WAS NO INJURY TO THE PATIENT AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24309623 ↗
  • MalfunctionFeb 10, 2026Break

    DURING A NEUROVASCULAR PROCEDURE THE PHYSICIAN WAS USING THE GUIDEWIRE AND IT BROKE. ALL OF THE WIRE WAS ABLE TO BE REMOVED AND ANOTHER DEVICE WAS PULLED TO COMPLETE THE PROCEDURE. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24309621 ↗
  • MalfunctionFeb 10, 2026Break

    DURING A NEUROVASCULAR PROCEDURE THE PHYSICIAN WAS PUTTING THE WIRE IN AND FELT FRICTION. THE PHYSICIAN NOTICED THE DISTAL END OF THE WIRE WAS STARTING TO UNRAVEL AND SUBSEQUENTLY REMOVED THE GUIDEWIRE AND REPLACED IT WITH ANOTHER DEVICE. THE PROCEDURE WAS COMPLETED WITH NO ISSUE AND NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24309563 ↗
  • MalfunctionDec 2, 2025Break; Break

    DURING A ANEURYSM CASE USING AN ARISTOTLE 24 (A24-200-002) GUIDEWIRE BY RADIAL ACCESS THE PHYSICIAN NOTED THE CATHETER KINKED AND THE GUIDEWIRE GOT CAUGHT AND STRETCHED. THE PHYSICIAN MENTIONED THE GUIDEWIRE PERFORMING WELL GOING INTO THE ICA AND THE GUIDEWIRE TIP WAS RESPONDING AS EXPECTED UP UNTIL THE GUIDEWIRE BROKE. THE PHYSICIAN ALSO MENTIONED THEY DECIDED TO PERFORM THE PROCEDURE THROUGH THE ARM / SHOULDER AND THAT IT WAS A TORTUOUS PATHWAY WITH A SHARP TURN AND WAS HARD TO NAVIGATE. THE PHYSICIAN LOST TIP CONTROL WHEN THEY WITNESSED THE CATHETER KINKING AND THE GUIDEWIRE STRETCHING. THE PHYSICIAN PULLED EVERYTHING OUT AND DECIDED TO SWITCH ACCESS TO THE GROIN INSTEAD USING A NEW ARISTOTLE 24. THE PROCEDURE WAS COMPLETED WITHOUT ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23699343 ↗
  • InjuryFeb 20, 2025Break; Break

    WHILE TRYING TO POSITION A PHENOM 27 IN A FUSIFORM ANEURYSM THE PHYSICIAN DECIDED TO REPLACE THE SYNCHRO GUIDEWIRE WITH AN ARISTOTLE 24 GUIDEWIRE. WITHIN TWO MINUTES OF USING THE ARISTOTLE 24 GUIDEWIRE THE PHYSICIAN REMOVED THE WIRE TO RESHAPE THE TIP. WHEN HE REMOVED THE WIRE APPROXIMATELY 1.5 INCHES TO 2 INCHES OF THE DISTAL PORTION OF THE GUIDEWIRE BROKE OFF AND REMAINED INSIDE OF THE ANEURYSM. THE BROKEN PORTION OF THE GUIDEWIRE WAS THEN ENCAPSULATED INSIDE OF THE ANEURYSM WITH THE FLOW DIVERTER.

    Patient problems: Foreign Body In Patient

    MAUDE report 21421574 ↗
  • DeathAug 20, 2024Break; Break

    ON 22 JUL 2024, SCIENTIA VASCULAR WAS NOTIFIED THAT AN A24-200-002 GUIDEWIRE FRACTURED DURING A NEUROVASCULAR PROCEDURE, ATTEMPTS TO GATHER MORE INFORMATION BETWEEN THE SCIENTIA VASCULAR SALES REPRESENTATIVE AND THE HOSPITAL OCCUR, AND NO ADDITIONAL INFORMATION WAS PROVIDED. ON 26 JUL 2024, DETAILS OF THE EVENT WERE PROVIDED TO SCIENTIA VASCULAR. THE EVENT WAS DESCRIBED AS FOLLOWS: "DURING MECHANICAL THROMBECTOMY FOR ACUTE STROKE, THE DISTAL TIP OF THE REPERFUSION CATHETER TIP SEPARATED. AFTER SUCCESSFUL RECANALIZATION, IT APPEARED THAT THE CATHETER TIP WAS RETAINED IN THE THROMBOSIS AND ULTIMATELY MIGRATED; RESULTED IN M1 OCCLUSION. AN ARISTOTLE 24 GUIDEWIRE WAS USED TO TRY TO OBTAIN THE BROKEN CATHETER TIP, THE GUIDEWIRE ALSO FRACTURED AND CONTRIBUTED TO THE OCCLUSION. OCCLUSION COULD NOT BE CLEARED. MANY ATTEMPTS TO RETRIEVE CATHETER TIP AND GUIDEWIRE WITH A VARIETY OF STRATEGIES WERE UNSUCCESSFUL." THE COMPLAINT GUIDEWIRE WAS NOT RETURNED TO SCIENTIA VASCULAR FOR EVALUATION, DESPITE SALES REPRESENTATIVE ATTEMPTS TO COLLECT THE GUIDEWIRE.

    Patient problems: Foreign Body In Patient; Obstruction/Occlusion

    MAUDE report 20005961 ↗
  • DeathJul 30, 2024

    DURING MECHANICAL THROMBECTOMY FOR ACUTE STROKE, THE DISTAL TIP OF THE REPERFUSION CATHETER TIP SEPARATED. AFTER SUCCESSFUL RECANALIZATION, IT APPEARED THAT THE CATHETER TIP WAS RETAINED IN THE THROMBOSIS AND ULTIMATELY MIGRATED; RESULTED IN M1 OCCLUSION. AN ARISTOTLE 24 GUIDEWIRE WAS USED TO TRY TO OBTAIN THE BROKEN CATHETER TIP, THE GUIDEWIRE ALSO FRACTURED AND CONTRIBUTED TO THE OCCLUSION. OCCLUSION COULD NOT BE CLEARED. MANY ATTEMPTS TO RETRIEVE CATHETER TIP AND GUIDEWIRE WITH A VARIETY OF STRATEGIES WERE UNSUCCESSFUL.

    Patient problems: Foreign Body In Patient; Obstruction/Occlusion

    MAUDE report 19876094 ↗

Other Wire, Guide, Catheter devices

All Wire, Guide, Catheter devices →

Sources and updates