Socrates SC038-127-001Recalls & reported problems
Scientia Vascular LLC · Socrates Aspiration System: 127cm working length aspiration catheter · Class 2
All models of Socrates (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00818075011025), received between Oct 3, 2024 and Dec 27, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jan 2, 2025Class IIOpen, Classified
Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.
Z-1234-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 27, 2025Break; Break
DURING AN INTRACRANIAL COILING OF A PERICALLOSAL HEMORRHAGE ANEURYSM A SOCRATES 038 CATHETER FRACTURED AND BECAME DISPLACED/SEPARATED WHILE INSIDE OF A ZEBRA GUIDE CATHETER. THE PHYSICIAN NOTED THE ZEBRA CATHETER WAS KINKED BUT ALSO NOTED THE PROCEDURE WAS COMPLETED WITH A BENCHMARK CATHETER THROUGH THE SAME ZEBRA CATHETER WITHOUT INCIDENT AND THERE WAS NO INJURY TO THE PATIENT. THE PHYSICIAN NOTED TORTUOUS ANATOMY AND AN ACUTE TURN GOING FROM THE A1 TO THE A2 BRANCH. THE DISPLACED PIECE OF THE SOCRATES 038 CATHETER WAS REMOVED FROM THE ZEBRA CATHETER USING ASPIRATION BEFORE CONTINUING THE PROCEDURE USING THE BENCHMARK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23903168 ↗InjuryOct 3, 2024Break; Break
ON 03 SEP 2024, SCIENTIA VASCULAR WAS NOTIFIED THAT A SOCRATES 38 (SC038-127-001, LN: 029278) ASPIRATION CATHETER DISTAL TIP DETACHED DURING A STROKE PROCEDURE. THE BREAK OCCURRED WHEN THE DOCTOR WAS PULLING THE SOCRATES 038 ASPIRATION CATHETER INTO THE CEREBASE GUIDE CATHETER. THE BROKEN PIECE OF THE SOCRATES 038 REMAINED IN THE PATIENT SUPPORTED BY A STENT. THE EVENT WAS INITIALLY DESCRIBED AS FOLLOWS: "DURING A STROKE THE SOC 127 / HEADWAY DUO 167 / TIGER 13 WAS ADVANCED. TIGER 13 WAS DEPLOYED THEN CONSTRAINED 50% AND PULLED BACK. IT WAS THEN CONSTRAINED ANOTHER 25% TO FIT INTO THE GUIDE. WHILE PULLING EVERYTHING BACK INTO THE GUIDE THE TIP OF THE SOC127 DETACHED AND WENT DISTAL. THE DOCTOR TRIED TO RETRIEVE IT BUT COULD NOT. HE ENDED UP STETTING OVER IT." THE COMPLAINT UNIT (SOCRATES 038) WAS RETURNED TO SCIENTIA VASCULAR ON 10-SEP-2024.
Patient problems: Foreign Body In Patient
MAUDE report 20362704 ↗
Other Catheter, Thrombus Retriever devices
- Penumbra System 5MAXJET7KIT-BPenumbra, Inc. · Catheter, Thrombus Retriever1 recall14 deaths221 reports
- Penumbra System 3MAXCPenumbra, Inc. · CATHETER, THROMBUS RETRIEVER1 death241 reports
- EMBOTRAP ET309537Neuravi Limited · Catheter, Thrombus Retriever13 deaths177 reports
- ACE 5MAXACE068KITPenumbra, Inc. · CATHETER, THROMBUS RETRIEVER170 reports
- Embotrap ET009533Neuravi Limited · Catheter, Thrombus Retriever16 deaths153 reports
- Penumbra System 5MAXACE132Penumbra, Inc. · CATHETER, THROMBUS RETRIEVER3 deaths112 reports
- Penumbra System 5MAXJET7KIT-APenumbra, Inc. · Catheter, Thrombus Retriever1 death94 reports
- Trevo XP 90185Stryker Corporation · CATHETER, THROMBUS RETRIEVER1 recall1 death40 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page