Skip to content

Socrates recallZ-1234-2025 · Scientia Vascular, Inc. · initiated Jan 2, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Scientia Vascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.

FDA root cause category: Under Investigation by firm

What the company is doing

On 01/02/2025, the firm initially notified customer via email an "URGENT: MEDICAL DEVICE RECALL" Letter informing them that through internal evaluation of finished goods of the Socrates 38 Aspiration Catheter, one lot of pouches containing the 127 cm Socrates 38 Aspiration Catheter was identified with pouches that had a gap in the sterile barrier seal. This was a result of the pouch not being centered with the sealing bar during the sealing process. On 01/10/2025, The firm sent an updated letter informing customers of additional product lots were added to the Recall. Customers are instructed to: Segregate Recalled Product. Immediately remove all impacted (see list above) Socrates 38 Catheter(s) from their inventory (regardless of its location) and segregate this product in a secure location for return to Scientia Vascular, Inc. In situations where a device has already been used in a patient procedure, closely monitor patients for signs and symptoms of post-surgical infection. Customer Service will be available to provide additional information and provide replacement orders at [email protected] or (888)385-9016; Monday through Friday, 8:00 AM to 4:30 PM, Mountain Time.

Recall details

Recall number
Z-1234-2025
Status
Open, Classified
Initiated by the company
Jan 2, 2025
Posted by the FDA
Feb 26, 2025
Recalling firm
Scientia Vascular, Inc.
Quantity
120 devices
Product code
NRY
FDA submissions
K223913
Distribution
U..S. Nationwide distribution in the states of AL, AZ, FL, GA, HI, IL, IN, KS, MN, NC, NJ, NY, PA, SD, TX, VA, WI and WV.

Product description

Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute ischemic stroke

Code information (models, lots, serial numbers)

Catalog Number: SC038-127-001 UDI-DI code: 00818075011025 Lot Numbers: 030471 030317 030345 030376 030427 030451

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K223913 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.