Skip to content

AtriCure cryoICE cryoSPHERE Cryoablation System (REF A001009)Recalls & reported problems

Atricure, Inc. · CRYOS, Packaged Assembly · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818354016321), received between Jul 7, 2022 and Jul 7, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 7, 2022Crack; Crack

    ON (B)(6) 2021 WHEN PREPARING FOR A MODIFIED PECTUS EXCAVATUM PROCEDURE, THE CRYOS PROBE WAS CONNECTED TO THE ACM GENERATOR, LAID ON A TABLETOP AND A TEST FREEZE WAS INITIATED. WHEN THE PROBE REACHED -40 DEGREES, THE ACM COUNTED DOWN FOR ABOUT 10 SECONDS AND THEN SCRUB TECHNICIAN VENTED IT BY PRESSING THE N2O MANUAL EXHAUST KNOB. WHEN DOING THIS A LOUD "POP" SOUND OCCURRED AND THE DEVICE HANDLE SEPARATED. THERE WAS NO PATIENT HARM. DURING DEVICE EVALUATION, IT WAS OBSERVED THAT THE EXHAUST CRIMP SLEEVE WAS NOT PRESENT IN THE DEVICE. THIS DISCOVERY ESCALATED THIS PRODUCT PROBLEM TO A REPORTABLE EVENT PER PART 803 ON (B)(6) 2022.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14950614 ↗

Other Device, Surgical, Cryogenic devices