Skip to content

LineSider LS-N6SG7545Recalls & reported problems

Integrity Implants Inc. · 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated - 7.5mm x 45mm · Class 2 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818613022735), received between Jan 7, 2022 and Jan 7, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier.

3 other recalls cover devices cleared under the same FDA submission (K203367); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 27, 2022Terminated

    Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

    Z-0419-2023 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 7, 2022Fracture; Fracture

    ON 08-DEC-2021 ACCELUS WAS INFORMED OF A POST-OPERATIVE FAILURE ON A NON-MODULAR STANDARD SCREW. ACCELUS WAS ADVISED THAT THE SCREW HEAD APPEARS IN IMAGING TO HAVE BROKEN AT THE POINT WHERE IT ATTACHES TO THE SHAFT (OR TULIP WAS DISSOCIATED FROM THE SCREW) AND THIS MAY BE CONSISTENT WITH A POTENTIAL WELD FAILURE THOUGH INVESTIGATION IS STILL UNDERWAY. THE ORIGINAL PROCEDURE WAS A 2 LEVEL TLIF AND FIXATION PROCEDURE AND TOOK PLACE ON (B)(6) 2021. CURRENTLY THERE ARE NO PLANS TO PERFORM A REVISION SURGERY AS THE PATIENT IS DOING WELL AND HEALING AS EXPECTED. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13196810 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K203367. One submission often covers many models, so check the product and lot details in each record.

  • Z-0421-2023 · Aug 3, 2022 · Terminated
    LineSider Spinal System pedicle screws, 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Model Number LS-N6SG6545. Thoracolumbosacral pedicle screw system; Spinal i
  • Z-0418-2023 · Jan 27, 2022 · Terminated
    LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sol
  • Z-0420-2023 · Jan 27, 2022 · Terminated
    LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.