Mariner 41-1010Recalls & reported problems
Seaspine Orthopedics Corporation · Set Screw · Class 2
All models of Mariner (16)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Seaspine Mariner set screw is a small part of a pedicle screw system used to stabilize the spine during surgery. The reports mostly describe set screws loosening after implantation, sometimes causing rods to come loose and leading to additional surgery. There are 85 reports describing 1 death and 19 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 85
- Deaths reported
- 1
- Injuries reported
- 19
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 85 adverse event reports that list this exact device (DI 10889981155100), received between Dec 7, 2018 and May 30, 2024 — classified as 1 death, 19 injuries, 65 malfunctions.
The most frequently reported specific device problem is “Migration” (29 reports).
1 other recall cover devices cleared under the same FDA submission (K173882); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Migration29
- Detachment of Device or Device Component21
- Mechanical Problem20
- Premature Separation17
- Separation Problem15
- Separation Failure2
- Loose or Intermittent Connection1
- Loosening of Implant Not Related to Bone-Ingrowth1
Patient problems
- Failure of Implant71
- Patient Problem/Medical Problem18
- Insufficient Information7
- Pain6
- No Clinical Signs, Symptoms or Conditions2
- Death1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 85 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 30, 2024Migration; Migration
ON 02 MAY 2024, SEASPINE WAS MADE AWARE THAT A CUSTOMER UTILIZING THE MARINER MIS POSTERIOR FIXATION SYSTEM WAS EXPERIENCING AN "ISSUE WITH THE SET SCREWS LOWERING." MULTIPLE REQUESTS WERE MADE FOR CLARIFICATION, NO FURTHER INFORMATION WAS PROVIDED. IT IS PRESUMED THAT THE CUSTOMER MEANT "SET SCREW LOOSENING." CURRENTLY, NO REVISION SURGERIES ARE PLANNED, AND NO RETURNS ARE EXPECTED.
Patient problems: Insufficient Information
MAUDE report 19427282 ↗MalfunctionJan 5, 2024Migration; Migration
ON 13 DEC 2023, SEASPINE WAS MADE AWARE OF TWO SET SCREW LOOSENING EVENTS INVOLVING THE MARINER POSTERIOR FIXATION SYSTEM. THIS REPORT IS FOR THE SECOND PATIENT. IT WAS NOTED THAT A ROD HAD COME LOOSE, AND ADDITIONAL SURGERY IS REQUIRED. THE INITIAL SURGERY OCCURRED ON (B)(6) 2023; THE REVISION SURGERY HAS NOT YET BEEN SCHEDULED.
Patient problems: Insufficient Information
MAUDE report 18456899 ↗MalfunctionAug 22, 2023Separation Problem; Separation Problem
THIS EVENT WAS ORIGINALLY REPORTED ON 3012120772-2022-00019. THIS REPORT IS FOR 3 OF 5 LOT NUMBERS. A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S MARINER PEDICLE SCREW SYSTEM ON (B)(6) 2021. SEASPINE WAS MADE AWARE ON 11 APR 2022 THAT SET SCREWS HAD LOOSENED FROM THE CONSTRUCT IN THE POSTOPERATIVE PERIOD. THIS WAS NOTED AT THE 6-MONTH POST-OPERATIVE APPOINTMENT. ON 12 MAY 2022 SEASPINE WAS MADE AWARE THAT A REVISION SURGERY HAD OCCURRED ON (B)(6) 2022. EXPLANTS WERE RETURNED ON (B)(6) 2022.
Patient problems: Failure of Implant
MAUDE report 17599745 ↗MalfunctionAug 22, 2023Separation Problem; Separation Problem
THIS EVENT WAS ORIGINALLY REPORTED ON 3012120772-2022-00019. THIS REPORT IS FOR 2 OF 5 LOT NUMBERS. A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S MARINER PEDICLE SCREW SYSTEM ON (B)(6) 2021. SEASPINE WAS MADE AWARE ON(B)(6) 2022 THAT SET SCREWS HAD LOOSENED FROM THE CONSTRUCT IN THE POSTOPERATIVE PERIOD. THIS WAS NOTED AT THE 6-MONTH POST-OPERATIVE APPOINTMENT. ON (B)(6) 2022 SEASPINE WAS MADE AWARE THAT A REVISION SURGERY HAD OCCURRED ON (B)(6) 2022. EXPLANTS WERE RETURNED ON (B)(6) 2022.
Patient problems: Failure of Implant
MAUDE report 17599683 ↗MalfunctionAug 22, 2023Separation Problem; Separation Problem
THIS EVENT WAS ORIGINALLY REPORTED ON 3012120772-2022-00019. THIS REPORT IS FOR 5 OF 5 LOT NUMBERS. A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S MARINER PEDICLE SCREW SYSTEM ON (B)(6) 2021. SEASPINE WAS MADE AWARE ON (B)(6) 2022 THAT SET SCREWS HAD LOOSENED FROM THE CONSTRUCT IN THE POSTOPERATIVE PERIOD. THIS WAS NOTED AT THE 6-MONTH POST-OPERATIVE APPOINTMENT. ON (B)(6) 2022 SEASPINE WAS MADE AWARE THAT A REVISION SURGERY HAD OCCURRED ON (B)(6) 2022. EXPLANTS WERE RETURNED ON (B)(6) 2022.
Patient problems: Failure of Implant
MAUDE report 17599266 ↗MalfunctionAug 22, 2023Premature Separation; Premature Separation
SEASPINE WAS MADE AWARE OF A POST-OP LOOSENING EVENT ON (B)(6) 2022 THAT OCCURRED 5 MONTHS ((B)(6) 2022) PRIOR TO THE EVENT DATE ((B)(6) 2022). THE CASE WAS FOR FRACTURE AT L4 ON A 46 YEAR OLD MALE. IT WAS COMPLETED IN A PERCUTANEOUS FASHION USING THE LARGEST SCREWS THE PEDICLE WOULD HOLD AS WELL AS A 6.0COCR ROD. A 105 IN/LB TORQUE WAS USED ON REGULAR SET CAPS TO FINAL LOCK EVERYTHING. THIS REPORT IS FOR 1 OF 2 LOT NUMBERS PROVIDED. THIS REPORT WAS ORGINALLY FILED UNDER 3012120772-2022-00031 BUT REPORTED TWO LOT NUMBERS. THIS REPORT SEPARATES OUT LOT AW135865D. LOT AW135871D IS REPORTED ON 3012120772-2023-00018.
Patient problems: Insufficient Information
MAUDE report 17599212 ↗MalfunctionAug 22, 2023Separation Problem; Separation Problem
THIS EVENT WAS ORIGINALLY REPORTED ON 3012120772-2022-00019. THIS REPORT IS FOR 4 OF 5 LOT NUMBERS. A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S MARINER PEDICLE SCREW SYSTEM ON (B)(6) 2021. SEASPINE WAS MADE AWARE ON (B)(6) 2022 THAT SET SCREWS HAD LOOSENED FROM THE CONSTRUCT IN THE POSTOPERATIVE PERIOD. THIS WAS NOTED AT THE 6-MONTH POST-OPERATIVE APPOINTMENT. ON 12 MAY 2022 SEASPINE WAS MADE AWARE THAT A REVISION SURGERY HAD OCCURRED ON (B)(6) 2022. EXPLANTS WERE RETURNED ON 16MAY2022.
Patient problems: Failure of Implant
MAUDE report 17599206 ↗MalfunctionAug 22, 2023Separation Problem; Separation Problem
THIS EVENT WAS ORIGINALLY REPORTED ON 3012120772-2022-00019. THIS REPORT IS FOR 1 OF 5 LOT NUMBERS. A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S MARINER PEDICLE SCREW SYSTEM ON (B)(6) 2021. SEASPINE WAS MADE AWARE ON 11 APR 2022 THAT SET SCREWS HAD LOOSENED FROM THE CONSTRUCT IN THE POSTOPERATIVE PERIOD. THIS WAS NOTED AT THE 6-MONTH POST-OPERATIVE APPOINTMENT. ON 12 MAY 2022 SEASPINE WAS MADE AWARE THAT A REVISION SURGERY HAD OCCURRED ON (B)(6) 2022. EXPLANTS WERE RETURNED ON 16MAY2022.
Patient problems: Failure of Implant
MAUDE report 17599175 ↗MalfunctionAug 22, 2023Premature Separation; Premature Separation
SEASPINE WAS MADE AWARE OF A POST-OP LOOSENING EVENT ON (B)(6) 2022 AT OCCURRED 5 MONTHS ((B)(6) 2022) PRIOR TO THE EVENT DATE ((B)(6) 2022). THE CASE WAS FOR FRACTURE AT L4 ON A 46 YEAR OLD MALE. IT WAS COMPLETED IN A PERCUTANEOUS FASHION USING THE LARGEST SCREWS THE PEDICLE WOULD HOLD AS WELL AS A 6.0COCR ROD. A 105 IN/LB TORQUE WAS USED ON REGULAR SET CAPS TO FINAL LOCK EVERYTHING. THIS REPORT IS FOR 2 OF 2 LOT NUMBERS PROVIDED. THIS REPORT WAS "ORIGINALLY" FILED UNDER 3012120772-2022-00031 BUT REPORTED TWO LOT NUMBERS. THIS REPORT SEPARATES OUT LOT AW135871D. LOT AW135865D WAS REPORTED ON 3012120772-2023-00024.
Patient problems: Insufficient Information
MAUDE report 17599098 ↗InjuryJan 31, 2023Migration; Migration
SEASPINE WAS MADE AWARE ON (B)(6) 2023 THAT A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF A SEASPINE MARINER MIS SYSTEM. THE INDEX SURGERY WAS ON (B)(6) 2022. THE DISTRIBUTOR REPORTED THAT A CONSTRUCT FAILURE OCCURRED POST-OPERATIVELY. THE X-RAYS PROVIDED APPEAR TO SHOW THAT THE P/N: 41-1010 SET SCREW LOOSENED AT L4.
Patient problems: Insufficient Information
MAUDE report 16273161 ↗InjuryDec 22, 2022Migration; Migration
PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S MARINER PEDICLE SCREW SYSTEM ON (B)(6) 2022. ON (B)(6) 2022 IT WAS IDENTIFIED THAT THE S1 SET SCREW LOOSENED, CAUSING THE ROD TO DISENGAGE FROM THE SUPERIOR TULIP HEAD AND SLID INFERIORLY. REMOVAL AND RE-INSTRUMENTATION OF THE FIXATION OCCURRED ON (B)(6) 2022. SEASPINE WAS MADE AWARE OF THIS EVENT ON (B)(6) 2022. THERE WERE 2 SCREWS AND 1 ROD P/N MM1-155040) INVOLVED IN THIS SINGLE EVENT. LOT NUMBERS AW154138D/AW127247D.
Patient problems: Insufficient Information
MAUDE report 16039948 ↗InjuryDec 22, 2022Migration; Migration
SEASPINE WAS MADE AWARE OF AN EVENT INVOLVING MULTIPLE SET SCREWS LOOSENING POST-OPERATIVELY. THE INDEX SURGERY WAS ON (B)(6) 2022 AND THE EVENT DATE IS ON (B)(6) 2022 THERE WERE 10 SCREWS OVER 6 LOTS INVOLVED AND ARE AS FOLLOWS: PRODUCT LOT NUMBER PRODUCT QUANTITY AW155418D 4, AW154141D 2, AW132005D 1, AW147643D 1, AW154681D 1, AW154138D 1. THIS PATIENT HAD A PRIOR FUSION FROM L2-L4. THEY HAD A PREVIOUS FRACTURE AT T12 WHICH WAS CEMENTED. THEY KYPHOSED ABOVE THEIR HARDWARE AT L1. L1 AND L2 SUBSEQUENTLY FUSED TOGETHER. IF I REMEMBER CORRECTLY[?]THEY ALSO HAD SOME CERVICAL WORK DOWN THAT'S NOT SHOWN.
Patient problems: Insufficient Information
MAUDE report 16039921 ↗MalfunctionNov 23, 2022Premature Separation; Premature Separation
ON (B)(6) 2022 THE SEASPINE CLINICAL TEAM PROVIDED AN UPDATED SAE FORM FOR PATIENT (B)(6). THE EVENT DESCRIPTION IN THE UPDATED FORM NOTED THAT ON (B)(6) 2021 THE PATIENT HAS PLATEAUED IN IMPROVEMENT WITH PAIN ON THE LEFT SIDE LOW BACK. ALSO, NOTED BILATERAL SET NUT LOOSENING AT L5. REVISION TO BE CONSIDERED IF THE PATIENT DOES NOT CONTINUE TO IMPROVE. AT THAT TIME THE PATIENT CONTINUED PHYSICAL THERAPY AND THE HOPE WAS THAT BRACING WITH THE INTERBODY FUSION WOULD LEAD TO A SOLID FUSION AND REDUCTION IN PAIN. NOTE THAT THE INITIAL LEFT L5 SET SCREW LOOSENING WAS RECORDED UNDER PR (B)(4). TREATMENT DESCRIPTION: REVISION POSTERIOR SPINAL FUSION L5-S1 WITH HARDWARE REMOVAL AND REINSERTION WITH ANOTHER SYSTEM.
Patient problems: Failure of Implant; Pain
MAUDE report 15854374 ↗MalfunctionSep 13, 2022Separation Failure; Premature Separation; Separation Failure; Premature Separation
SEASPINE WAS NOTIFIED ON (B)(6) 2022 OF A SCREW BACK OUT EVENT. ON THE FILMS PROVIDED, THERE WAS ALSO A ROD FRACTURE OBSERVED.
Patient problems: Failure of Implant
MAUDE report 15415031 ↗MalfunctionMay 9, 2022Premature Separation; Premature Separation
A PATIENT UNDERWENT SPINAL SURGERY UTILIZING SEASPINE'S MARINER PEDICLE SCREW SYSTEM ON (B)(6) 2021 . SEASPINE WAS MADE AWARE ON (B)(6) 2022 THAT SET SCREWS HAD LOOSENED FROM THE CONSTRUCT IN THE POSTOPERATIVE PERIOD.
Patient problems: Failure of Implant
MAUDE report 14342789 ↗InjuryApr 7, 2022Premature Separation; Premature Separation
A PATIENT UNDERWENT A 5 LEVEL SPINAL SURGERY FROM LEVEL L1-S1 ON (B)(6) 2021 CONSISTING OF SEASPINE'S MARINER PEDICLE SCREW SYSTEM. SEASPINE WAS MADE AWARE OF A REVISION SURGERY THAT TOOK PLACE ON (B)(6) 2022 DUE TO IMPLANTS LOOSENING FROM THE CONSTRUCT IN THE POSTOPERATIVE PERIOD.
Patient problems: Failure of Implant
MAUDE report 14043451 ↗MalfunctionMar 23, 2022Premature Separation; Premature Separation
A PATIENT UNDERWENT A 3 LEVEL SPINAL SURGERY FROM LEVEL L3-L5 ON (B)(6) 2021 CONSISTING OF SEASPINE'S MARINER PEDICLE SCREW SYSTEM. SEASPINE WAS MADE AWARE ON 25 FEB 2022 THAT THE TWO SET SCREWS AT THE L3 LEVEL LOOSENED FROM THE CONSTRUCT IN THE POSTOPERATIVE PERIOD.
Patient problems: Failure of Implant
MAUDE report 13857946 ↗InjuryFeb 3, 2022Premature Separation; Premature Separation
A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S MARINER PEDICLE SCREW SYSTEM AND SEASPINE'S REGATTA AND REEF TO INTERVERTEBRAL FUSION DEVICES ON (B)(6) 2020. ON (B)(6) 2022, IT WAS REPORTED THAT A REVISION SURGERY OCCURRED ON (B)(6) 2020 TO REPLACE THREE MARINER SET SCREWS THAT LOOSENED IN THE POSTOPERATIVE PERIOD. TWO OF THE THREE SET SCREWS REPORTEDLY LOOSENED A SECOND TIME, PROMPTING A SECOND REVISION SURGERY, WHICH OCCURRED ON (B)(6) 2022, PROMPTING REPORT 3012120772-2022-00001.
Patient problems: Failure of Implant
MAUDE report 13444909 ↗InjuryFeb 3, 2022Premature Separation; Premature Separation
A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S MARINER PEDICLE SCREW SYSTEM AND SEASPINE'S REGATTA AND REEF TO INTERVERTEBRAL FUSION DEVICES ON (B)(6) 2020. ON (B)(6) 2022, IT WAS REPORTED THAT A REVISION SURGERY OCCURRED ON (B)(6) 2020 TO REPLACE THREE MARINER SET SCREWS THAT LOOSENED IN THE POSTOPERATIVE PERIOD (REF: 3012120772-2022-00002 FOR THIS EVENT). TWO OF THE THREE SET SCREWS REPORTEDLY LOOSENED AGAIN, PROMPTING A SECOND REVISION SURGERY, WHICH OCCURRED ON (B)(6) 2022, PROMPTING THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 13444889 ↗MalfunctionNov 9, 2021Mechanical Problem; Mechanical Problem
PATIENT UNDERWENT INITIAL PROCEDURE ON (B)(6) 2020. DURING FOLLOW-UP CT SCAN, IT WAS NOTICED BY THE SURGEON, THAT THE SET SCREW NO LONGER LOCKED INTO THE POLYAXIAL HEAD. THE SURGEON DECIDED TO REOPERATE ON THIS PATIENT, TO REMOVE THE SET SCREW AND POLYAXIAL HEAD AND REPLACE WITH THE NEW ONES. THE REASON FOR THE LOOSENING IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12782870 ↗MalfunctionOct 21, 2021Separation Problem; Separation Problem
DOCTOR CONDUCTED A REMOVAL OF IMPLANTS ON A PATIENT THAT HAD AN L3-S1 ALIF + POST SCREW FIXATION WITH MIS MARINER ON (B)(6) 2021. UPON REMOVING THE S1 AND L5 SCREWS, HE NOTICED THAT THE SET SCREWS HAD LOOSENED SIGNIFICANTLY. ALTHOUGH THE SET SCREWS DID NOT FULLY REMOVE, THE THERE WAS NO TORQUE RETAINED ON THE IMPLANTS. OF NOTE, THE INDEX SURGERY WAS PERFORMED WITH 85LB TORQUE LIMITERS ON (B)(6) 2020. IMAGES WERE PROVIDED THAT SHOW THE BOTTOMS OF THE SET SCREWS THAT WERE LOOSE.
Patient problems: Failure of Implant
MAUDE report 12677646 ↗MalfunctionSep 16, 2021Separation Problem; Separation Problem
SURGEON REPORTED THAT A SET SCREW BACKOUT OCCURRED APPROXIMATELY 6 MONTHS POST-OPERATION. THE REPORT NOTED THE LOOSENING OCCURRED AT THE BOTTOM OF THE CONSTRUCT. WHILE SURGICAL REVISION IS PLANNED, THERE IS NO DEFINITIVE DATE FOR THE SURGERY BETWEEN THE DATE THAT THE ISSUE WAS REPORTED AND THE TIME OF THIS REPORT. THERE WERE NO PRODUCTS RETURNED FOR EVALUATION AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 12481904 ↗MalfunctionAug 27, 2021Separation Problem; Separation Problem
ON (B)(6) 2021, IT WAS REPORTED THAT BACKED OUT SCREWS WERE OBSERVED DURING A PROCEDURE. THERE WERE NO RESPONSES FROM THE REPORTER OR THE FACILITY TO FOLLOW UP ATTEMPTS REQUESTING FURTHER INFORMATION.
Patient problems: Failure of Implant
MAUDE report 12380783 ↗MalfunctionJul 6, 2021Loose or Intermittent Connection; Separation Problem; Loose or Intermittent Connection; Separation Problem
PATIENT UNDERWENT REVISION SURGERY ON (B)(6) 2021, FOR UNSPECIFIED REASONS AT WHICH TIME IT WAS NOTED THAT 1 OUT OF 6 LOCK SCREWS HAD BACKED OUT. THE PRODUCT IS NOT AVAILABLE FOR RETURN. THE REASON FOR THE REVISION SURGERY IS UNKNOWN AND NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Failure of Implant
MAUDE report 12124751 ↗MalfunctionJun 9, 2021Separation Problem; Separation Problem
A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S MARINER PEDICLE SCREW SYSTEM FROM LEVELS L3-L5. A LATERAL X-RAY REVEALED THAT A SET SCREW MIGRATED IN THE POSTOPERATIVE PERIOD.
Patient problems: Failure of Implant
MAUDE report 11970405 ↗
FDA clearance and classification
Cleared through 510(k) K173882 on Feb 6, 2018 (decision: Substantially Equivalent). Submitted by SeaSpine Orthopedics Corporation as “SeaSpine Mariner Pedicle Screw System”. The FDA review took 47 days. FDA record ↗
The FDA regulates product code NKB (thoracolumbosacral pedicle screw system) under 21 CFR 888.3070 as a Class II device. 21 CFR 888.3070 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Intended to provide immobilization and stabilization of spinal segments in thoracic, lumbar, and sacral spine as an adjunct to fusion.
Recalls under the same FDA submission
These recalls involve devices cleared under K173882. One submission often covers many models, so check the product and lot details in each record.
- Z-2762-2026 · May 28, 2026 · Open, ClassifiedSeaSpine Polyaxial Head REF Number: 41-3010
Common questions about the Mariner 41-1010
Has the Mariner 41-1010 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10889981155100), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the Mariner 41-1010?
The FDA's MAUDE database lists 85 adverse event reports for this device, received between Dec 7, 2018 and May 30, 2024. In those reports, the most common specific device problems are “Migration” (29), “Detachment of Device or Device Component” (21) and “Mechanical Problem” (20); the most common patient problems are “Failure of Implant” (71), “Patient Problem/Medical Problem” (18) and “Pain” (6). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Mariner 41-1010?
1 report is classified as a death, 19 as injuries and 65 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Mariner 41-1010 FDA approved?
It was cleared through 510(k) K173882 on Feb 6, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Mariner 41-1010, and is it still sold?
The FDA's device database lists Seaspine Orthopedics Corporation as the company responsible for it (brand Mariner, model 41-1010). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Thoracolumbosacral pedicle screw system devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system122 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page