RxSight Light Adjustable Lens (RxLAL) 60005-2160Recalls & reported problems
Rxsight, Inc. · LAL, Sterile, Packaged, 16.0 Diopter · Class 3 · Implant
All models of RxSight Light Adjustable Lens (RxLAL) (62)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00818806020098), received between Aug 12, 2020 and Mar 18, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Optical Distortion” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Optical Distortion1
- Optical Problem1
Patient problems
- Visual Disturbances1
- Visual Impairment1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 18, 2021Optical Problem; Optical Problem
SITE REPORTED THAT PATIENT WITH LIGHT ADJUSTABLE LENS IMPLANTED IN THE SULCUS HAD THE LENS EXPLANTED ON (B)(6) 2021 FROM THE RIGHT EYE AS THE DESIRED REFRACTIVE OUTCOME WAS NOT ACHIEVED AFTER LIGHT ADJUSTMENT TREATMENTS.
Patient problems: Visual Impairment
MAUDE report 11518222 ↗InjuryAug 12, 2020Optical Distortion; Optical Distortion
PATIENT IMPLANTED WITH LIGHT ADJUSTABLE LENS ((B)(6) 2020) IN THE RIGHT EYE ADMITTED TO POOR COMPLIANCE WITH UV SPECTACLES IMMEDIATELY AFTER SURGERY. PATIENT DEVELOPED A LOSS OF VISUAL ACUITY. THE LIGHT ADJUSTABLE LENS WAS EXPLANTED (SERIAL NUMBER OF EXPLANTED LENS: (B)(6)) AND EXCHANGED WITH ANOTHER LIGHT ADJUSTABLE LENS (SERIAL NUMBER OF IMPLANTED LENS: (B)(6)) OF THE SAME DIOPTER POWER ON (B)(6) 2020. PATIENT AGAIN DID NOT COMPLY WITH UV SPECTACLE WEAR AND DEVELOPED A LOSS OF VISUAL ACUITY. THE LENS WAS EXPLANTED ON (B)(6) 2020 AND EXCHANGED WITH A 3-PIECE MONOFOCAL IOL. RXSIGHT'S FIRST AWARENESS OF THE EXPLANT ON (B)(6) 2020 INVOLVING SERIAL NUMBER (B)(6) WAS (B)(6) 2020. RXSIGHT'S FIRST AWARENESS OF THE EXPLANT ON (B)(6) 2020 INVOLVING SERIAL NUMBER (B)(6) WAS (B)(6) 2020.
Patient problems: Visual Disturbances
MAUDE report 10397103 ↗
Other Light Adjustable Lens (Lal) And Light Delivery Device (Ldd) devices
- RxSight Light Adjustable Lens (RxLAL) 60005-3205Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)50 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3200Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)47 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3210Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)39 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3190Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)32 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3195Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)31 reports
- Light Adjustable Lens (LAL) 60007-0200Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)31 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3215Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)30 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3185Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)29 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page