RxSight Light Adjustable Lens (RxLAL) 60005-2185Recalls & reported problems
Rxsight, Inc. · LAL, Sterile, Packaged, 18.5 Diopter · Class 3 · Implant
All models of RxSight Light Adjustable Lens (RxLAL) (62)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00818806020142), received between Sep 11, 2020 and Mar 11, 2022 — classified as 6 injuries.
The most frequently reported specific device problem is “Optical Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Optical Problem4
- Optical Distortion2
Patient problems
- Visual Impairment4
- Visual Disturbances2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMar 11, 2022Optical Problem; Optical Problem
THE SITE REPORTED THAT A LIGHT ADJUSTABLE LENS HAD BEEN EXPLANTED FROM THE PATIENT'S LEFT EYE AS THE DESIRED REFRACTIVE OUTCOME WAS NOT ACHIEVED AFTER LIGHT ADJUSTMENT TREATMENTS. ANOTHER LAL WAS IMPLANTED DURING THE EXCHANGE ON (B)(6) 2022. RXSIGHT'S FIRST AWARENESS OF THE EVENT OCCURRED ON (B)(6) 2022.
Patient problems: Visual Impairment
MAUDE report 13733983 ↗InjuryAug 31, 2021Optical Problem; Optical Problem
SITE REPORTED THAT THE LIGHT ADJUSTABLE LENS WAS EXPLANTED DUE TO VISUAL DISTURBANCES. RXSIGHT'S FIRST AWARENESS OF THE EVENT WAS (B)(6) 2021.
Patient problems: Visual Impairment
MAUDE report 12394391 ↗InjuryAug 10, 2021Optical Problem; Optical Problem
SITE REPORTED THAT THE LIGHT ADJUSTABLE LENS WAS EXPLANTED AS THE DESIRED REFRACTIVE OUTCOME WAS NOT ACHIEVED AFTER LIGHT ADJUSTMENT TREATMENTS. RXSIGHT'S FIRST AWARENESS OF THE EVENT WAS 7/15/2021.
Patient problems: Visual Impairment
MAUDE report 12302294 ↗InjuryFeb 9, 2021Optical Problem; Optical Problem
SITE REPORTED THAT THE LIGHT ADJUSTABLE LENS WAS EXPLANTED FROM THE PATIENT'S RIGHT EYE AS THE DESIRED REFRACTIVE OUTCOME WAS NOT ACHIEVED AFTER LIGHT ADJUSTMENT TREATMENTS. RXSIGHT'S FIRST AWARENESS OF THE EVENT OCCURRED ON (B)(6) 2021.
Patient problems: Visual Impairment
MAUDE report 11305419 ↗InjuryOct 9, 2020Optical Distortion; Optical Distortion
PATIENT REPORTED TO HAVE SPENT TIME INDOORS BY WINDOWS WITHOUT UV PROTECTIVE EYEWEAR RESULTING IN DECREASED VISUAL ACUITY IN THE LEFT EYE. PREMATURE PHOTOPOLYMERIZATION CONFIRMED, RESULTING IN EXPLANT OF THE LENS.
Patient problems: Visual Disturbances
MAUDE report 10659106 ↗InjurySep 11, 2020Optical Distortion; Optical Distortion
TWO DAYS FOLLOWING LIGHT TREATMENT, PATIENT NOTED A DECREASE IN ACUITY AND QUALITY OF VISION. EXAMINATION CONFIRMED A DECREASE IN BCVA. RETROILLUMINATION NOTED A DEFECT NEAR THE CENTER OF THE LENS CONSISTENT WITH PREMATURE PHOTOPOLYMERIZATION OF THE LENS.
Patient problems: Visual Disturbances
MAUDE report 10522405 ↗
Other Light Adjustable Lens (Lal) And Light Delivery Device (Ldd) devices
- RxSight Light Adjustable Lens (RxLAL) 60005-3205Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)50 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3200Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)47 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3210Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)39 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3190Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)32 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3195Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)31 reports
- Light Adjustable Lens (LAL) 60007-0200Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)31 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3215Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)30 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3185Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)29 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page