Light Adjustable Lens (LAL) 60007-0210Recalls & reported problems
Rxsight, Inc. · LAL+, Light Adjustable Lens, Sterile, Packaged, 21.0 Diopter · Class 3 · Implant
All models of Light Adjustable Lens (LAL) (33)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This RxSight Light Adjustable Lens is an implantable eye lens whose power is fine-tuned with light after surgery. The 28 reports mostly describe blurred vision and other visual disturbances after the light treatments, with several lenses removed and replaced. No deaths were reported, and 28 injuries were. There are no recalls naming this device. Reports of this kind rose from 7 in the earlier 12-month period to 18 in the most recent one.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 24
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00818806021606), received between Mar 12, 2024 and Jun 2, 2026 — classified as 24 injuries.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem24
Patient problems
- Blurred Vision16
- Visual Disturbances9
- Unspecified Eye / Vision Problem3
- Capsular Bag Tear1
- Diplopia/Double Vision1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
InjuryJun 2, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+) WAS EXPLANTED FROM THE LEFT EYE DUE TO THE PATIENT'S DISSATISFACTION WITH THEIR VISION AFTER ONE LIGHT ADJUSTMENT. A LENS FROM A DIFFERENT MANUFACTURER WAS IMPLANTED AS A REPLACEMENT AND THE PATIENT IS REPORTED TO BE DOING WELL POST-OPERATIVELY. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Unspecified Eye / Vision Problem
MAUDE report 25362277 ↗InjuryMay 6, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+) WAS EXPLANTED FROM THE RIGHT EYE DUE TO BLURRY VISION. A NEW LAL WAS IMPLANTED AS A REPLACEMENT AND THE PATIENT IS REPORTED TO BE DOING WELL POST-OPERATIVELY.
Patient problems: Blurred Vision
MAUDE report 25095048 ↗InjuryMay 1, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ WAS EXPLANTED FROM THE LEFT EYE DUE TO VISUAL DISTURBANCES AND BLURRY VISION.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 25060327 ↗InjuryApr 14, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED DUE TO BLURRY VISION AND VISUAL DISTURBANCES AND A NEW LIGHT ADJUSTABLE LENS (LAL, SN (B)(6), +20.5D) IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 24879073 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED DUE TO BLURRY VISION AND VISUAL DISTURBANCES AND A NEW LIGHT ADJUSTABLE LENS (LAL, SN (B)(6), +21.0D) IMPLANTED AS A REPLACEMENT. DURING THE EXPLANT PROCEDURE, A POSTERIOR CAPSULE TEAR OCCURRED AND AN ANTERIOR VITRECTOMY WAS PERFORMED.
Patient problems: Blurred Vision; Capsular Bag Tear; Visual Disturbances
MAUDE report 24720037 ↗InjuryMar 20, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0) WAS EXPLANTED DUE TO BLURRY VISION AND A NEW LAL+ (SN (B)(6), +20.5) IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision
MAUDE report 24654895 ↗InjuryMar 20, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED DUE TO VISUAL DISTURBANCES AND BLURRY VISION.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 24654452 ↗InjuryMar 13, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN: (B)(6), +21.0D) WAS EXPLANTED DUE TO BLURRY VISION AND THE PATIENT'S DISSATISFACTION WITH THE CHOSEN REFRACTIVE TARGET.
Patient problems: Blurred Vision
MAUDE report 24592290 ↗InjuryMar 12, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED DUE TO BLURRY VISION AND A NEW LIGHT ADJUSTABLE LENS (LAL, SN (B)(6), +21.0D) IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision
MAUDE report 24585983 ↗InjuryMar 4, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED FROM THE LEFT EYE DUE TO BLURRY VISION AND VISUAL DISTURBANCES AND A NEW LIGHT ADJUSTABLE LENS (LAL, SN (B)(6), +21.0D) IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 24508905 ↗InjuryFeb 5, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A BILATERALLY IMPLANTED PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6)) WAS EXPLANTED FROM THE RIGHT EYE DUE TO BLURRY DISTANCE VISION. A NEW LAL (SN (B)(6)) WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision
MAUDE report 24270021 ↗InjuryJan 15, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN: (B)(6), +21.0D) WAS EXPLANTED DUE TO THE PATIENT'S DISSATISFACTION WITH THE CHOSEN REFRACTIVE TARGET AND A NEW LAL+ (SN: (B)(6), +21.0D) IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision
MAUDE report 24087854 ↗InjuryJan 14, 2026Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED DUE TO THE PATIENT'S DISSATISFACTION WITH THE CHOSEN REFRACTIVE TARGET AND VISUAL DISTURBANCES.
Patient problems: Visual Disturbances
MAUDE report 24067260 ↗InjuryDec 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN: (B)(6), +21.0D) WAS EXPLANTED FROM THE RIGHT EYE DUE TO BLURRY VISION. A NEW LAL (SN: (B)(6), +21.0D) WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision
MAUDE report 23706531 ↗InjurySep 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED FROM THEIR RIGHT EYE DUE TO BLURRY VISION. A STANDARD LAL (SN (B)(6), +21.5D) WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision
MAUDE report 23068761 ↗InjuryAug 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED. A NEW LAL+ (SN (B)(6), +21.0D) WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Unspecified Eye / Vision Problem
MAUDE report 22723047 ↗InjuryJun 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED DUE TO BLURRY VISION AND DIPLOPIA AND A NEW LAL+ (SN (B)(6), +21.5D) IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision; Diplopia/Double Vision
MAUDE report 22204131 ↗InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED DUE TO THE PATIENT'S DISSATISFACTION WITH VISION AND A NEW LIGHT ADJUSTABLE LENS (SN (B)(6), +20.5D) WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision
MAUDE report 21644928 ↗InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT IMPLANTED WITH A LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) ON (B)(6) 2023 HAD THE LENS EXPLANTED ON (B)(6) 2024 DUE TO UNHAPPINESS WITH THEIR UNCORRECTED VISION AND A NEW LIGHT ADJUSTABLE LENS (SN (B)(6), +20.5D) IMPLANTED AS A REPLACEMENT.
Patient problems: Blurred Vision
MAUDE report 21044401 ↗InjuryNov 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS IMPLANTED BACKWARDS DURING PRIMARY CATARACT SURGERY. THE TREATING PHYSICIAN OBSERVED THE ISSUE DURING THE POST-OPERATIVE 1-DAY EXAM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20664359 ↗InjuryNov 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT THE PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED DUE TO PATIENT DISSATISFACTION WITH THEIR VISION. THE LAL+ WAS EXPLANTED AND A LIGHT ADJUSTABLE LENS (LAL, SN (B)(6), +21.0D) WAS IMPLANTED AS A REPLACEMENT.
Patient problems: Unspecified Eye / Vision Problem
MAUDE report 20626216 ↗InjuryAug 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A BILATERALLY IMPLANTED PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6) WAS EXPLANTED FROM THEIR LEFT EYE DUE TO CHANGE IN REFRACTIVE TARGETS DURING ADJUSTMENT PERIOD FROM DISTANCE TO NEAR. RXSIGHT'S FIRST AWARENESS OF THE EVENT WAS ON 07/29/2024.
Patient problems: Visual Disturbances
MAUDE report 20082204 ↗InjuryJul 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A SITE REPORTED TO RXSIGHT THAT A PATIENT'S LIGHT ADJUSTABLE LENS+ (LAL+, SN (B)(6), +21.0D) WAS EXPLANTED DUE TO GLARE SYMPTOMS AND A NEW LAL (SN (B)(6), +20.5) IMPLANTED AS A REPLACEMENT. RXSIGHT'S FIRST AWARENESS OF THIS EVENT WAS ON 06/05/2024.
Patient problems: Visual Disturbances
MAUDE report 19680530 ↗InjuryMar 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REFERENCE REPORT #(B)(4) FOR THE REPORT ON THE RIGHT EYE (OD).
Patient problems: Visual Disturbances
MAUDE report 18887245 ↗
Other Light Adjustable Lens (Lal) And Light Delivery Device (Ldd) devices
- RxSight Light Adjustable Lens (RxLAL) 60005-3205Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)50 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3200Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)47 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3210Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)39 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3190Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)32 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3195Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)31 reports
- Light Adjustable Lens (LAL) 60007-0200Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)31 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3215Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)30 reports
- RxSight Light Adjustable Lens (RxLAL) 60005-3185Rxsight, Inc. · Light Adjustable Lens (Lal) And Light Delivery Device (Ldd)29 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page