Skip to content

Artisan Power MB-8713030-NRecalls & reported problems

Artisan Power LLC · Replacement Battery for B. Braun Perfusor Space Syringe Infusion Pump - 8713030 · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00819595015418), received between Mar 31, 2026 and Mar 31, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Difficult to Setup or Prepare1
  • Insufficient Flow or Under Infusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2026Device Difficult to Setup or Prepare; Insufficient Flow or Under Infusion

    AS REPORTED BY THE USER FACILITY: WE HAVE SEEN A COUPLE SAFEDAY REPORTS RECENTLY ABOUT THERE BEING MORE VERSED LEFT IN THE SYRINGE THAN THERE SHOULD HAVE BEEN BASED ON CHARTING AND THE RATE OF INFUSION. VERSED COMES IN A MANUFACTURER SYRINGE THAT IS A BD 30ML SYRINGE. WHEN LOOKING AT THE PUMP LOGS, IT WOULD SHOW THAT THE NURSE CHOSE THE 50ML SYRINGE SIZE WHEN PROGRAMMING THE PUMP, CAUSING THE DISCREPANCY AND ULTIMATELY ALTERING DRUG DELIVERY. BUT WHEN YOU PLACE A 30ML SYRINGE IN THE PUMP, IT ONLY PROMPTS FOR 20ML OR 30ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742980 ↗

Other Battery, Replacement, Rechargeable devices

All Battery, Replacement, Rechargeable devices →

Sources and updates