Skip to content

NITREX N183002Recalls & reported problems

Covidien LP · Guidewire · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821684058213), received between Dec 6, 2018 and Dec 6, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 6, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    PHYSICIAN USED A NITREX GUIDEWIRE WITH A TRAILBLAZER DEVICE DURING TREATMENT OF A 230MM PLAQUE CTO (CHRONIC TOTAL OCCLUSION-100%) IN THE PATIENT¿S MID SUPERFICIAL FEMORAL ARTERY (SFA). MODERATE CALCIFICATION AND SLIGHT TORTUOSITY ARE REPORTED. IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. THE VESSEL WAS NOT PRE-DILATED. IT IS REPORTED THAT THE TIP OF THE NITREX DEVICE WAS PROLAPSED OUTSIDE .035 X 135 TRAILBLAZER IN ATTEMPT TO CROSS SFA CTO. WHEN WITHDRAWING THE WIRE FROM THE TRAILBLAZER, THE NITREX WIRE BROKE. THE PHYSICIAN THEN REMOVED THE TRAILBLAZER WITH THE REMAINING WIRE (WITH TIP). NO PART OF THE WIRE REMAINED IN THE PATIENT. THE PROCEDURE WAS THEN COMPLETED USING A .035 X 260 (GLIDEWIRE ADVANTAGE) NON-MEDTRONIC WIRE. NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8137213 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →