SureGlide 35BXRecalls & reported problems
Gyrus ACMI, LLC · Guidewire 0.035" (0.89 mm) x 150 cm Regular Body, 3 cm Straight Tip (5/PK) · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00821925005730), received between Jun 16, 2021 and Nov 11, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Detachment of Device or Device Component1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 11, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, APPROXIMATELY 0.5CM PIECE OF THE LINING OF A GUIDE WIRE BROKE OFF FROM THE GUIDE WIRE AND WAS IN THE PATIENT'S BODY, UNTIL THE SURGEON REMOVED IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20658923 ↗MalfunctionOct 15, 2024
APPROXIMATELY 0.5CM PIECE OF THE LINING OF A GUIDE WIRE BROKE OFF FROM THE GUIDE WIRE AND WAS IN THE PATIENT'S BODY, UNTIL THE SURGEON REMOVED IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20450930 ↗InjuryJun 16, 2021Material Separation; Material Separation
IT IS REPORTED A PATIENT HAD A LEFT URETEROSCOPY WITH STENT FOR THE INDICATION OF LEFT URETERAL STONE (B)(6) 2020 USING A SURE GLIDE GUIDEWIRE AND AN HOLMIUM LASER (THE LASER SETTINGS WERE: 800 MJ AND 8 HZ. JOULES: 263.20, PULSES: 329, DURATION: 41 SECONDS). AFTER THE PROCEDURE, THE PATIENT EXPERIENCED URINARY URGENCY, FREQUENCY, AND URETHRAL PAIN. AN UNSPECIFIED TIME LATER, THE PATIENT HAD AN ULTRASOUND AND IT WAS FOUND THAT THE PATIENT HAD RETAINED A PORTION OF THE WIRE. THE PATIENT HAD ANOTHER URETEROSCOPY PROCEDURE (B)(6) 2021 TO TRY TO RETRIEVE THE BROKEN GUIDEWIRE WITH NO SUCCESS. THE PATIENT WAS THEN BOOKED FOR A LOWER URETERAL POLE STENT, UPPER POLE URETER URETEROSCOPY, REMOVAL OF FOREIGN BODY IN THE DISTAL URETER OF THE UPPER POLE URETER, URETERAL STENT IN UPPER POLE URETER AND REMOVAL OF STENT IN LOWER POLE AT THE LOCAL HOSPITAL TO HAVE THE DEVICE FRAGMENT REMOVED (B)(6) 2021 WHICH WAS SUCCESSFUL. THE PATIENT HAD A STENT PLACED AND CAME BACK SEVERAL WEEKS LATER TO HAVE THE STENT REMOVED AND IS NOW DOING FINE. NO FURTHER PROBLEMS. THE DEVICE FRAGMENT THAT WAS REMOVED FROM THE PATIENT WAS SENT TO PATHOLOGY. IT MEASURED 3.6CM X 0.1CM AND CAME BACK AS HYDROPHILIC NITINOL CORE
Patient problems: Foreign Body In Patient
MAUDE report 12014265 ↗
FDA clearance and classification
The FDA regulates product code EYA (stylet, ureteral) under 21 CFR 876.5130 as a Class I device. 21 CFR 876.5130 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SureGlide 35BX
Has the SureGlide 35BX been recalled?
No FDA recall record in our data names this exact device identifier (DI 00821925005730), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SureGlide 35BX?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jun 16, 2021 and Nov 11, 2024. In those reports, the most common specific device problems are “Break” (1), “Detachment of Device or Device Component” (1) and “Material Separation” (1); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SureGlide 35BX?
None of the reports are classified as deaths, 2 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SureGlide 35BX FDA approved?
Devices of this type (product code EYA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Gyrus ACMI, LLC.
Who makes the SureGlide 35BX, and is it still sold?
The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand SureGlide, model 35BX). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STYLET, URETERAL devices
- ZIPwire™ M0066802220Lake Region Medical · Stylet, Ureteral45 reports
- ZIPwire™ M006630222B0Lake Region Medical · Stylet, Ureteral21 reports
- OLYMPUS GWH3505RGyrus ACMI, LLC · STYLET, URETERAL16 reports
- ZIPwire™ M006630205B0Lake Region Medical · Stylet, Ureteral10 reports
- ZIPwire™ M0066802080Lake Region Medical · Stylet, Ureteral9 reports
- ZIPwire™ M0066802050Lake Region Medical · Stylet, Ureteral9 reports
- OLYMPUS GWH3505RAGyrus ACMI, LLC · STYLET, URETERAL5 reports
- ZIPwire™ M006630209B0Lake Region Medical · Stylet, Ureteral5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.