Olympus BD-400P-2080Recalls & reported problems

Gyrus ACMI, LLC · EZDilate - Fixed Wire Esophageal 3-Stage Balloon Dilatation Catheter 18-20mm · Class 2

All models of Olympus (27)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EZDilate fixed-wire balloon dilatation catheter is used to widen narrowed parts of the esophagus during endoscopy. The 60 reports mostly describe the balloon not deflating properly, getting stuck in the scope channel and having to be cut out, or bursting, with 10 injuries and no deaths. One open recall, started in November 2023, covers balloon inflation, deflation and retrieval problems, including bursting, leaking and a case where a piece was left in the patient.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
59
Deaths reported
0
Injuries reported
9
Recalls naming it
1

Summary

The FDA's MAUDE database contains 59 adverse event reports that list this exact device (DI 00821925033214), received between Feb 1, 2021 and Nov 13, 2025 — classified as 9 injuries, 50 malfunctions.

The most frequently reported specific device problem is “Failure to Deflate” (18 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 10, 2023Class IIOpen, Classified

    Product balloon issues: inflating, deflating and/or retrieving the devices; reports of bursting and leaking and events involving foreign body in patient and prolonged procedure.

    Z-0938-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Deflate18
  • Deflation Problem14
  • Burst Container or Vessel10
  • Material Rupture6
  • Difficult to Remove5
  • Device Damaged Prior to Receipt by User3
  • Mechanical Problem2
  • Physical Resistance/Sticking2

Patient problems

  • No Clinical Signs, Symptoms or Conditions58
  • No Consequences Or Impact To Patient1
  • Pain1

Recent adverse event reports

The 25 most recent of 59 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2025Difficult to Remove; Failure to Deflate; Difficult to Remove; Failure to Deflate

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23550837 ↗
  • InjuryJul 14, 2025Detachment of Device or Device Component; Failure to Deflate; Detachment of Device or Device Component; Failure to Deflate

    IT WAS REPORTED UPON THE COMPLETION OF AN ESOPHAGOGASTRODUODENOSCOPY (ESG), THE USER WAS UNABLE TO DEFLATE THE SUBJECT DEVICE. THE USER HAD TO CUT THE BALLOON TO REMOVE IT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Pain

    MAUDE report 22493144 ↗
  • MalfunctionJul 7, 2025Failure to Deflate; Failure to Deflate

    IT WAS REPORTED DURING A THERAPEUTIC UPPER ENDOSCOPY PROCEDURE; THEY WERE NOT ABLE TO DEFLATE THE BALLOONS. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22420133 ↗
  • MalfunctionOct 14, 2024Failure to Deflate; Failure to Deflate

    IT WAS REPORTED THE BALLOON DIALATOR COULD NOT BE DEFLATED AND BECAME STUCK INSIDE THE SCOPE CHANNEL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20439378 ↗
  • MalfunctionJun 24, 2024Inflation Problem; Inflation Problem

    IT WAS REPORTED THE SINGLE-USE BALLOON INFLATOR EXPERIENCED AN OUT OF BOX FAILURE IN WHICH THE BALLOON WAS ALREADY INFLATED AND COULD NOT PASS THROUGH THE ENDOSCOPE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19593504 ↗
  • MalfunctionMar 12, 2024Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED, THAT THE BALLOON BURST WHEN INFLATED. THE ISSUE WAS FOUND DURING THE DIAGNOSTIC ENDOSCOPY PROCEDURE THAT WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18889315 ↗
  • MalfunctionJan 17, 2024

    INFLATION BALLOON WOULD NOT DEFLATE ENOUGH TO BE REMOVED FROM THE SCOPE. USED THE PUMP TO DEFLATE- IT WENT DOWN BUT NOT ENOUGH TO BE REMOVED FROM THE SCOPE. DISCONNECTED THE PUMP AND CONNECTED A SYRINGE AND TRIED TO REMOVE ANY WATER. TRIED TO REMOVE THE BALLOON AGAIN AND IT WOULD NOT BUDGE. HAD TO REMOVE SCOPE AND CUTTERS TO CUT IN 1/2 TO GET THE EQUIPMENT OUT OF THE SCOPE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18525982 ↗
  • MalfunctionDec 15, 2023Deflation Problem; Deflation Problem

    THE OLYMPUS FIELD CORRECTIVE ACTION ADMINISTRATOR REPORTED (ON BEHALF OF THE CUSTOMER) THAT THE BALLOON DOES NOT DEFLATE WELL, THEREFORE WAS UNABLE TO PULL THE BALLOON OUT WHILE USING THE EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE. IT WAS NOTED THAT TO PULL OUT THE BALLOON, THE ENTIRE ENDOSCOPE WAS WITHDRAWN FROM THE PATIENT THEN THE BALLOON PART WAS CUT. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18338648 ↗
  • MalfunctionDec 14, 2023Deflation Problem; Difficult to Remove; Deflation Problem; Difficult to Remove

    THE CUSTOMER REPORTED TO OLYMPUS THAT WHILE USING THE EZDILATE BALLOON DILATOR FOR A THERAPEUTIC PROCEDURE, THE USER WAS UNABLE TO PULL THIS DEVICE FROM THE SCOPE AFTER IT WAS DILATED. THE ISSUE OCCURRED AT THE END OF THE CASE. THE PROCEDURE DID NOT NEED TO BE PROLONGED, AND PATIENT¿S ANESTHESIA OR SEDATION WAS NOT EXTENDED. PROCEDURE WAS COMPLETED BY CUTTING THE BALLOON FROM THE CATHETER TO REMOVE FROM THE SCOPE. THE ISSUE DID NOT AFFECT THE OUTCOME OF THE PROCEDURE AND NO MEDICAL INTERVENTIONS WERE REQUIRED. DEVICE CONNECTED TO SUBJECT DEVICE DURING MALFUNCTION WAS AN INFLATION DEVICE FOR ENDOSCOPIC BALLOON DILATATION MODEL# MAJ-1740. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT. RELATED PATIENT IDENTIFIERS: (B)(6), (B)(6) 2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18332436 ↗
  • MalfunctionDec 7, 2023Deflation Problem; Difficult to Remove; Deflation Problem; Difficult to Remove

    RELATED PATIENT IDENTIFIERS: (B)(6) BD-400P-2080 SN (B)(6 (B)(3) (B)(6) 2023, (B)(6) BD-400P-2080 SN (B)(6) (B)(3) (B)(6) 2023, AND (B)(6) BD-400P-2080 SN UNKNOWN (B)(3) (B)(6) 2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18284284 ↗
  • MalfunctionDec 7, 2023Deflation Problem; Difficult to Remove; Deflation Problem; Difficult to Remove

    THE CUSTOMER REPORTED TO OLYMPUS THAT WHILE USING THE EZDILATE BALLOON DILATOR FOR A THERAPEUTIC PROCEDURE, THE USER WAS UNABLE TO PULL THIS DEVICE FROM THE SCOPE AFTER IT WAS DILATED. THE ISSUE OCCURRED AT THE END OF THE CASE. THE PROCEDURE DID NOT NEED TO BE PROLONGED, AND PATIENT¿S ANESTHESIA OR SEDATION WAS NOT EXTENDED. PROCEDURE WAS COMPLETED BY CUTTING THE BALLOON FROM THE CATHETER TO REMOVE FROM THE SCOPE. THE ISSUE DID NOT AFFECT THE OUTCOME OF THE PROCEDURE AND NO MEDICAL INTERVENTIONS WERE REQUIRED. DEVICE CONNECTED TO SUBJECT DEVICE DURING MALFUNCTION WAS AN INFLATION DEVICE FOR ENDOSCOPIC BALLOON DILATATION MODEL# MAJ-1740. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT. RELATED PATIENT IDENTIFIER: (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18284228 ↗
  • MalfunctionNov 22, 2023Infusion or Flow Problem; Infusion or Flow Problem

    ON BEHALF OF THE CUSTOMER, AN OLYMPUS REPRESENTATIVE REPORTED THE FACILITY HAD ISSUES DURING DEFLATION OF THE BALLOON, THE BALLOON GOT CAUGHT ON THE CATHETER AND HAD TO BE CUT OFF. THE EVENT OCCURRED DURING A DIAGNOSTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT. INSPECTION AND TESTING OF THE RETURNED DEVICE REVEALED UNSUCCESSFUL BALLOON INFLATION. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING DEVICE EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18194523 ↗
  • MalfunctionOct 27, 2023Failure to Deflate; Failure to Deflate

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE EZDILATE BALLOON DILATOR (FW) 18-19-20 BALLOONS ARE NOT DEFLATING. THE EVENT OCCURRED DURING A THERAPEUTIC PROCEDURE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18019838 ↗
  • MalfunctionOct 17, 2023Deflation Problem; Deflation Problem

    THE CUSTOMER REPORTED THAT THE BALLOON WOULD NOT DEFLATE ALL THE WAY AFTER ALL ATTEMPTS WERE MADE. THE ISSUE OCCURRED DURING A THERAPEUTIC ESOPHAGOGASTRODUODENOSCOPY PROCEDURE. THE PROCEDURE WAS DELAYED BY 2 MINUTES AND WAS COMPLETED WITHOUT HARM OR INJURY TO THE PATIENT. THERE WERE NO REPORTS OF HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17949741 ↗
  • MalfunctionOct 4, 2023Material Rupture; Material Rupture

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE EZDILATE BALLOON DILATOR (FW) 18-19-20 DURING A DIAGNOSTIC ENDOSCOPY PROCEDURE THE BALLOON BURST/RUPTURED WHILE INSIDE A PATIENT. THERE WAS NO DELAY, AND THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THE DEVICE WAS DISCARDED AFTER USE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17867624 ↗
  • MalfunctionAug 29, 2023Deflation Problem; Deflation Problem

    THE CUSTOMER REPORTED THAT THE BALLOON DID NOT DEFLATE COMPLETELY AND GOT STUCK IN THE SCOPE WHILE WITHDRAWING IT FROM THE SCOPE. THE REPORTED ISSUE OCCURRED DURING A THERAPEUTIC ESOPHAGEAL DILATION PROCEDURE. REPORTEDLY, THE BALLOON WAS CUT FROM THE SCOPE, BUT DILATION WAS COMPLETED FIRST. THE REPORTED ISSUE DID NOT AFFECT THE OUTCOME OF THE PROCEDURE. THERE WAS NO REPORT OF PATIENT OR USER INJURY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17646949 ↗
  • MalfunctionAug 24, 2023Material Rupture; Material Rupture

    THE CUSTOMER REPORTED THAT THE BALLOON POPPED DURING A THERAPEUTIC ESOPHAGEAL DILATION PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORT OF PATIENT OR USER INJURY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17619815 ↗
  • MalfunctionAug 24, 2023Deflation Problem; Deflation Problem

    THE CUSTOMER REPORTED THAT THE BALLOON DID NOT DEFLATE COMPLETELY AND GOT STUCK IN THE SCOPE WHILE WITHDRAWING IT FROM THE SCOPE. THE REPORTED ISSUE OCCURRED DURING A THERAPEUTIC ESOPHAGEAL DILATION PROCEDURE. REPORTEDLY, THE BALLOON WAS CUT FROM THE SCOPE, BUT DILATION WAS COMPLETED FIRST. THE REPORTED ISSUE DID NOT AFFECT THE OUTCOME OF THE PROCEDURE. THERE WAS NO REPORT OF PATIENT OR USER INJURY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17619512 ↗
  • MalfunctionAug 24, 2023Deflation Problem; Deflation Problem

    THE CUSTOMER REPORTED THAT THE BALLOON DID NOT DEFLATE COMPLETELY AND GOT STUCK IN THE SCOPE WHILE WITHDRAWING IT FROM THE SCOPE. THE REPORTED ISSUE OCCURRED DURING A THERAPEUTIC ESOPHAGEAL DILATION PROCEDURE. REPORTEDLY, THE BALLOON WAS CUT FROM THE SCOPE, BUT DILATION WAS COMPLETED FIRST. THE REPORTED ISSUE DID NOT AFFECT THE OUTCOME OF THE PROCEDURE. THERE WAS NO REPORT OF PATIENT OR USER INJURY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17619361 ↗
  • MalfunctionAug 22, 2023Mechanical Problem; Physical Resistance/Sticking; Mechanical Problem; Physical Resistance/Sticking

    THE CUSTOMER REPORTED THAT WHEN USING AN EZ DILATE BALLOON DILATOR (FW) 18-19-20, THE BALLOON GOT STUCK IN THE ENDOSCOPE AND NEEDED TO BE CUT TO REMOVE IT. THE EVENT OCCURRED DURING THE THERAPEUTIC PROCEDURE (UPPER GI ENDOSCOPY) AND WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THIS REPORT IS LINKED TO PATIENT IDENTIFIER: (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17597752 ↗
  • MalfunctionAug 21, 2023Physical Resistance/Sticking; Physical Resistance/Sticking

    THE CUSTOMER REPORTED THAT WHEN USING AN EZ DILATE BALLOON DILATOR (FW) ON (B)(6) 2020, THE BALLOON GOT STUCK IN THE ENDOSCOPE AND NEEDED TO BE CUT TO REMOVE IT. THE EVENT OCCURRED DURING THE THERAPEUTIC PROCEDURE (UPPER GI ENDOSCOPY) AND WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS ABOUT A FIVE-MINUTE DELAY AND NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE REPORT IS LINKED TO PATIENT IDENTIFIERS: (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17589067 ↗
  • MalfunctionAug 17, 2023Deflation Problem; Deflation Problem

    THE CUSTOMER REPORTED THAT THE BALLOON FAILED DURING A DIAGNOSTIC ESOPHAGEAL DILATION PROCEDURE. REPORTEDLY, THE BALLOON DID NOT FULLY DEFLATE. THE DOCTOR WAS ABLE TO RETRIEVE THE WHOLE BALLOON, AFTER REMOVING THE SCOPE FROM THE PATIENT. THE INTENDED PROCEDURE WAS COMPLETED USING THE SAME DEVICE WITHOUT A DELAY. THERE WAS NO REPORT OF PATIENT OR USER INJURY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17568519 ↗
  • MalfunctionAug 16, 2023Burst Container or Vessel; Burst Container or Vessel

    THE PATIENT WAS UNDER MONITORED ANESTHESIA CARE (MAC) DURING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17558797 ↗
  • MalfunctionAug 14, 2023Deflation Problem; Deflation Problem

    THE CUSTOMER CONFIRMED THE PATIENT DID NOT REQUIRE ADDITIONAL ANESTHESIA. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PROCEDURE WAS PRACTICALLY ENDED WHEN THE DEVICE MALFUNCTIONED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS FROM THE MALFUNCTION. THIS DEVICE MALFUNCTION DID NOT AFFECT PATIENT OUTCOMES. NO OTHER DEVICES USED DURING THE EVENT THAT COULD HAVE CONTRIBUTED TO THE MALFUNCTION. OTHER RELATED PATIENT IDENTIFIERS: (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17541311 ↗
  • MalfunctionAug 14, 2023Deflation Problem; Deflation Problem

    ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER ON (B)(6) 2023. THE CUSTOMER CONFIRMED THAT THE ISSUE OCCURRED AT THE END OF THE PROCEDURE ONCE THEY WERE COMPLETED AND REMOVED THE BALLOON. THE CONDITION OF THE PATIENT WAS NOT IMPACTED BY THE FAILURE. THE TYPE OF PROCEDURE BEING PERFORMED WAS AN EGD WITH DIL. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE SAME DEVICE. THE PROCEDURE WAS NOT PROLONGED. THE PATIENT¿S ANESTHESIA OR SEDATION WAS NOT EXTENDED. THERE WAS NO MEDICAL INTERVENTION REQUIRED. IT IS UNKNOWN AS TO WHETHER ANY OTHER DEVICES WERE CONNECTED TO THE DEVICE WHEN THE FAILURE OCCURRED. THE DEVICE HAS BEEN RETURNED PER THE CUSTOMER. THE PATIENT DEMOGRAPHICS AND PRE-EXISTING CONDITIONS ARE NOT AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17540673 ↗

FDA clearance and classification

The FDA regulates product code KNQ (dilator, esophageal) under 21 CFR 876.5365 as a Class II device. 21 CFR 876.5365 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Olympus BD-400P-2080

Has the Olympus BD-400P-2080 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0938-2024 (Class II), began on Nov 10, 2023: Product balloon issues: inflating, deflating and/or retrieving the devices; reports of bursting and leaking and events involving foreign body in patient and prolonged procedure. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Olympus BD-400P-2080?

The FDA's MAUDE database lists 59 adverse event reports for this device, received between Feb 1, 2021 and Nov 13, 2025. In those reports, the most common specific device problems are “Failure to Deflate” (18), “Deflation Problem” (14) and “Burst Container or Vessel” (10); the most common patient problem is “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Olympus BD-400P-2080?

None of the reports are classified as deaths, 9 as injuries and 50 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Olympus BD-400P-2080 FDA approved?

Devices of this type (product code KNQ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Gyrus ACMI, LLC.

Who makes the Olympus BD-400P-2080, and is it still sold?

The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand Olympus, model BD-400P-2080). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Dilator, esophageal devices

All Dilator, esophageal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page