CRE Fixed Wire M00558380Recalls & reported problems

Boston Scientific Corporation · Esophageal Balloon Dilatation Catheter · Class 2

All models of CRE Fixed Wire (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific CRE Fixed Wire balloon, used to widen the esophagus, has 349 adverse event reports on file, with no deaths and 8 injuries. Reports mostly describe the balloon bursting, not deflating fully, or being hard to pull out of the scope, and a few mention tears or bleeding. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
344
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 344 adverse event reports that list this exact device (DI 08714729202011), received between May 24, 2017 and Sep 14, 2026 — classified as 8 injuries, 336 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (200 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 3 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 1 reportsAug 2024: 3 reportsSep 2024: 4 reportsOct 2024: 6 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 4 reportsJun 2025: 3 reportsJul 2025: 6 reportsAug 2025: 5 reportsSep 2025: 8 reportsOct 2025: 6 reportsNov 2025: 15 reportsDec 2025: 3 reportsJan 2026: 1 reportsFeb 2026: 5 reportsMar 2026: 4 reportsApr 2026: 10 reportsMay 2026: 5 reportsJun 2026: 6 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 6 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel200
  • Failure to Deflate45
  • Deflation Problem36
  • Material Puncture/Hole30
  • Adverse Event Without Identified Device or Use Problem25
  • Difficult to Remove15
  • Leak/Splash15
  • Use of Device Problem10

Patient problems

  • No Clinical Signs, Symptoms or Conditions253
  • No Consequences Or Impact To Patient76
  • No Known Impact Or Consequence To Patient8
  • Insufficient Information5
  • Hemorrhage/Blood Loss/Bleeding3
  • Laceration(s)2
  • Aspiration/Inhalation1
  • Chest Pain1

Recent adverse event reports

The 25 most recent of 344 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 14, 2026Burst Container or Vessel

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50764087 ↗
  • MalfunctionSep 14, 2026Failure to Deflate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755619 ↗
  • MalfunctionSep 14, 2026Failure to Deflate

    It was reported to Boston Scientific Corporation that a CRE Fixed Wire Dilatation Balloon was used in the esophagus during a Dilation procedure to treat stricture performed on (b)(6) 2026.\ \During the procedure, the dilatation balloon would not deflate after dilation. The gastroscope was withdrawn with the balloon remaining partially inflated. Once removed from the patient, the balloon was cut to facilitate separation of the dilator from the gastroscope. The procedure was prolonged by approximately 5 to 10 minutes due to the need for device exchange and reinsertion of the gastroscope. The procedure was completed using the original device.\ \There were no patient complications reported as a result of this event and the patient condition following procedure was reported to be fine.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755612 ↗
  • MalfunctionSep 14, 2026Failure to Deflate

    It was reported to Boston Scientific Corporation that a CRE Fixed Wire Dilatation Balloon was used in the esophagus during a Dilation procedure to treat stricture performed on (b)(6) 2026.\ \During the procedure, the dilatation balloon would not deflate after dilation. The gastroscope was withdrawn with the balloon remaining partially inflated. Once removed from the patient, the balloon was cut to facilitate separation of the dilator from the gastroscope. The procedure was prolonged by approximately 5 to 10 minutes due to the need for device exchange and reinsertion of the gastroscope. The procedure was completed using the original device.\ \There were no patient complications reported as a result of this event and the patient condition following procedure was reported to be fine.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755611 ↗
  • MalfunctionSep 14, 2026Failure to Deflate

    It was reported to Boston Scientific Corporation that a CRE Fixed Wire Dilatation Balloon was used in the esophagus during a Dilation procedure to treat stricture performed on (b)(6)2026.\ \During the procedure, the dilatation balloon would not deflate after dilation. The gastroscope was withdrawn with the balloon remaining partially inflated. Once removed from the patient, the balloon was cut to facilitate separation of the dilator from the gastroscope. The procedure was prolonged by approximately 5 to 10 minutes due to the need for device exchange and reinsertion of the gastroscope. The procedure was completed using the original device.\ \There were no patient complications reported as a result of this event and the patient condition following procedure was reported to be fine.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755610 ↗
  • MalfunctionSep 14, 2026Failure to Deflate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755609 ↗
  • MalfunctionJun 30, 2026Use of Device Problem; Failure to Deflate

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED DURING AN UPPER ENDOSCOPY PROCEDURE FOR DILATION PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE DEVICE DID NOT FULLY DEFLATE AS INTENDED. TO FACILITATE REMOVAL, THE WIRE WAS CUT, WHICH ALLOWED COMPLETE DEFLATION OF THE BALLOON. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENTS CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25673500 ↗
  • MalfunctionJun 17, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE TO TREAT DISPHASIA PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE BALLOON BURST UPON INFLATION. IT WAS REPORTED THAT NO PIECE OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER CRE FIXED WIRE DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION FOLLOWING PROCEDURE WAS REPORTED TO HAVE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25540719 ↗
  • MalfunctionJun 17, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE TO TREAT ESOPHAGITIS PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE BALLOON BURST WHILE INFLATING AT 20 MM. IT WAS REPORTED THAT NO PART OF THE BALLOON WAS DETACHED INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25534448 ↗
  • MalfunctionJun 16, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE FOR DYSPHASIA PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE BALLOON BURST AFTER REACHING A DIAMETER OF 20MM. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PIECE OF BALLOON DETACHED INSIDE THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENTS CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25524941 ↗
  • MalfunctionJun 16, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE FOR DYSPHASIA PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE BALLOON BURST AFTER REACHING A DIAMETER OF 20MM. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PIECE OF BALLOON DETACHED INSIDE THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENTS CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25524914 ↗
  • MalfunctionJun 9, 2026Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGEAL DILATION PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE BALLON RUPTURED BEFORE USED. IT WAS REPORTED THAT NO PIECE OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE FIXED WIRE DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25448718 ↗
  • MalfunctionMay 31, 2026Burst Container or Vessel

    NOTE: THIS REPORT PERTAINS TO ONE OF TWO CRE FIXED WIRE DILATATION BALLOONS USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO CRE FIXED WIRE DILATATION BALLOON WERE ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE TO TREAT DYSPHASIA PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, BALLOON BURST WHILE INFLATING TO 20MM. ADDITIONALLY, THE SAME PROBLEM OCCURRED WITH THE OTHER CRE FIXED WIRE DILATATION BALLOON. IT WAS REPORTED THAT NO PIECE OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER CRE FIXED WIRE DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION FOLLOWING PROCEDURE WAS REPORTED TO HAVE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25342775 ↗
  • MalfunctionMay 28, 2026Entrapment of Device; Use of Device Problem; Deformation Due to Compressive Stress; Failure to Deflate

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) DILATION PROCEDURE TO TREAT TROUBLE SWALLOWING PERFORMED ON 29 APR 2026. DURING WITHDRAWAL, THE BALLOON HAD BEEN INSERTED THROUGH THE ENDOSCOPE AND USED ON THE PATIENT. FOLLOWING COMPLETION OF DILATION, THE TECHNICIAN ATTEMPTED TO DEFLATE THE BALLOON; HOWEVER, THE BALLOON DID NOT FULLY DEFLATE AT THE DISTAL END AND COULD NOT BE WITHDRAWN FROM THE SCOPE. AS A RESULT, THE PHYSICIAN REMOVED THE ENDOSCOPE FROM THE PATIENT, AND THE TECHNICIAN USED CUTTING WIRE TO REMOVE THE BALLOON CATHETER. UPON INSPECTION, A KINK IN THE CATHETER WAS OBSERVED, AND THE BALLOON APPEARED TWISTED, POSSIBLY DUE TO WITHDRAWAL ATTEMPTS. THE PROCEDURE WAS PROLONGED AS A RESULT OF THIS EVENT. PHOTOS PROVIDED OF THE DEVICE OUTSIDE THE PATIENT SHOW THE CATHETER IS KINKED, STRETCHED AND MECHANICALLY CUT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION FOLLOWING PROCEDURE WAS REPORTED TO HAVE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25318076 ↗
  • MalfunctionMay 26, 2026Difficult to Remove; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS USED IN THE ESOPHAGUS DURING A DILATION PROCEDURE TO TREAT DYSPHAGIA PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, ONCE THE BALLON WAS DEFLATED, IT COULD NOT BE WITHDRAWN FROM THE SCOPE. THERE WAS A SMALL BUMP, RIDGE OR RAISED PORTION AT THE TIP OF THE BALLOON (WHERE THE TIP OF THE CATHETER AND THE BALLOON MEET) MAKING IT DIFFICULT TO BE REMOVED FROM THE SCOPE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25284799 ↗
  • MalfunctionMay 13, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGEAL DILATION PROCEDURE FOR SCHATZKIS RING PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE BALLOON BURST AFTER REACHING A DIAMETER OF 19MM. NO PIECE OF BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25161402 ↗
  • MalfunctionMay 5, 2026Failure to Deflate

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGEAL DILATATION (EGD) PROCEDURE FOR STRICTURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE BALLOON WAS NOT DEFLATING PROPERLY. THE BALLOON WAS PULLED OUT FROM THE PATIENT BY WAITING FOR FULL DEFLATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25088200 ↗
  • MalfunctionApr 30, 2026Use of Device Problem; Failure to Deflate

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE FOR DILATION PERFORMED ON (B)(6) 2026. DURING WITHDRAWAL, THE BALLOON WAS NOT ABLE TO DEFLATE, IT HAD TO BE CUT TO BE REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25050525 ↗
  • MalfunctionApr 29, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2026. DURING PROCEDURE, THE BALLOON POPPED. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25035193 ↗
  • MalfunctionApr 29, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2026. DURING PROCEDURE, THE BALLOON POPPED. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25035192 ↗
  • MalfunctionApr 23, 2026Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE TO TREAT STRICTURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, WHILE DILATING THE BALLOON AT 19MM THE BALLOON BURST. IT WAS REPORTED THAT NO PIECE OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24974420 ↗
  • MalfunctionApr 20, 2026Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN EGD (ESOPHAGOGASTRODUODENOSCOPY) WITH DILATION PROCEDURE FOR DYSPHAGIA PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE BALLOON BURST AFTER REACHING A DIAMETER OF 20MM. NO PIECE OF BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24927594 ↗
  • MalfunctionApr 13, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE MID ESOPHAGUS DURING A DILATION PROCEDURE FOR STRICTURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE BALLOON BURST AFTER REACHING 19MM OF DILATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME MODEL. NO PIECE OF BALLOON DETACHED INSIDE THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24867494 ↗
  • MalfunctionApr 8, 2026Difficult to Remove; Adverse Event Without Identified Device or Use Problem; Failure to Deflate

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS USED IN THE ESOPHAGUS DURING A CRE ESOPHAGEAL BALLOON DILATION PROCEDURE FOR ESOPHAGEAL STRICTURE PERFORMED ON (B)(6) 2026. DURING WITHDRAWAL, THE BALLOON DIDN'T DEFLATE COMPLETELY AND COULDN'T BE TAKEN OUT OF THE SCOPE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24812366 ↗
  • MalfunctionApr 7, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE FIXED WIRE DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE TO TREAT DYSPHAGIA PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, WHEN THE BALLOON REACHED 20 MM, IT BURST AND BEGAN TO DEFLATE. NO DEBRIS WAS OBSERVED IN THE ESOPHAGUS, AND ONLY MINIMAL BLEEDING WAS NOTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24805327 ↗

FDA clearance and classification

The FDA regulates product code KNQ (dilator, esophageal) under 21 CFR 876.5365 as a Class II device. 21 CFR 876.5365 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CRE Fixed Wire M00558380

Has the CRE Fixed Wire M00558380 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08714729202011), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the CRE Fixed Wire M00558380?

The FDA's MAUDE database lists 344 adverse event reports for this device, received between May 24, 2017 and Sep 14, 2026. In those reports, the most common specific device problems are “Burst Container or Vessel” (200), “Failure to Deflate” (45) and “Deflation Problem” (36); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (3), “Laceration(s)” (2) and “Aspiration/Inhalation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CRE Fixed Wire M00558380?

None of the reports are classified as deaths, 8 as injuries and 336 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CRE Fixed Wire M00558380 FDA approved?

Devices of this type (product code KNQ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Boston Scientific Corporation.

Who makes the CRE Fixed Wire M00558380, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand CRE Fixed Wire, model M00558380). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.