Soltive TFL-CPLURecalls & reported problems
Gyrus ACMI, LLC · TFL PREMIUM LASER UNIT - COMPONENT PART · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00821925044586), received between Dec 13, 2022 and Dec 10, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Fire” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
5 other recalls cover devices cleared under the same FDA submission (K242191); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Feb 27, 2026Open, Classified
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
Z-1694-2026 · Root cause: Process control · FDA record ↗
- Apr 4, 2025Open, Classified
Preset treatment parameters are not consistently being used in accordance with the IFU.
Z-1715-2025 · Root cause: Unknown/Undetermined by firm · FDA record ↗
Most reported problems
Device problems
- Fire1
- Output Problem1
Patient problems
- Laceration(s)1
- No Clinical Signs, Symptoms or Conditions1
- Unspecified Kidney or Urinary Problem1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 10, 2025Fire; Fire
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23774857 ↗InjuryDec 13, 2022Output Problem
PATIENT UNDERWENT SURGICAL LASER TREATMENT OF LEFT URETERAL STONES ON (B)(6) 2022. THE PROVIDER USED THE OLYMPUS SOLTIVE THULIUM LASER WITH 0.4 -1J ENERGY, FREQUENCY OF 16-40JZ, POWER OF 12.8-40W OVER 9MINS 52 SECS FOR TOTAL ENERGY OF 32.57KJ. THE PATIENT UNDERWENT A SECOND LASER LITHOTRIPSY OF LEFT URETERAL STONE ON (B)(6) 2022 WITH THE SAME LASER, 1J ENERGY WITH 20HZ FREQUENCY, 20W POWER OVER 39 MINUTES FOR A TOTAL OF 59.26KJ WAS USED. THE PATIENT RETURNED A THIRD TIME FOR SIMILAR PROBLEM, AND IT WAS DETERMINED THAT HE HAD A SEGMENT OF AVASCULARIZED NONVIABLE URETER THAT REQUIRED CORRECTIVE SURGERY. PATIENT UNDERWENT THE CORRECTIVE SURGERY AND HAS DONE WELL SINCE.
Patient problems: Laceration(s); Unspecified Kidney or Urinary Problem
MAUDE report 15977674 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K242191. One submission often covers many models, so check the product and lot details in each record.
- Z-1695-2026 · Feb 27, 2026 · Open, ClassifiedBrand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An el
- Z-1923-2025 · May 6, 2025 · Open, ClassifiedBrand Name: Powered Laser Surgical Instrument Product Name: Soltive Reusable Laser Fibers Model/Catalog Number: TFL-FBX150R, TFL-FBX200R, TFL-FBX365R, TFL-FBX550R, TFL-FBX940R, TFL
- Z-1921-2025 · May 6, 2025 · Open, ClassifiedBrand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Premium SuperPulsed Laser System Model/Catalog Number: TFL-PLS, TFL-PLS-JA (Japan only); Software Version: N/A P
- Z-1922-2025 · May 6, 2025 · Open, ClassifiedBrand Name: Powered Laser Surgical Instrument Product Name: Soltive SuperPulsed Laser Fibers - Single Use Model/Catalog Numbers: TFL-FBX150BS, TFL-FBX200BS, TFL-FBX150S, TFL-FBX200
- Z-1920-2025 · May 6, 2025 · Open, ClassifiedBrand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Number: TFL-SLS, TFL-SLS-JA (Japan only); Product Description: SOLTIV
Other Powered laser surgical instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports