Data updated Oct 16, 2026 from official FDA records · Editorial policy

16 open recalls name at least one SOLTIVE model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
867
Deaths reported
0
Injuries reported
97
Recalls
17

We found 16 models of SOLTIVE by Gyrus ACMI, LLC in the FDA's device registry (GUDID). Together they are named in 867 adverse event reports in the FDA's MAUDE database, received between Mar 22, 2021 and Sep 30, 2026 — classified as 97 injuries, 770 malfunctions.

The most frequently reported specific device problem is “Break” (201 reports).

17 FDA recall records name at least one of these models by its device identifier, 16 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 8 reportsDec 2023: 14 reportsJan 2024: 13 reportsFeb 2024: 20 reportsMar 2024: 6 reportsApr 2024: 5 reportsMay 2024: 18 reportsJun 2024: 11 reportsJul 2024: 14 reportsAug 2024: 20 reportsSep 2024: 20 reportsOct 2024: 28 reportsNov 2024: 8 reportsDec 2024: 8 reportsJan 2025: 7 reportsFeb 2025: 11 reportsMar 2025: 13 reportsApr 2025: 17 reportsMay 2025: 33 reportsJun 2025: 33 reportsJul 2025: 24 reportsAug 2025: 29 reportsSep 2025: 9 reportsOct 2025: 9 reportsNov 2025: 16 reportsDec 2025: 20 reportsJan 2026: 19 reportsFeb 2026: 17 reportsMar 2026: 12 reportsApr 2026: 31 reportsMay 2026: 14 reportsJun 2026: 14 reportsJul 2026: 0 reportsAug 2026: 4 reportsSep 2026: 19 reportsOct 2026: 0 reports20242025202633
Last 36 months. The most recent month may be incomplete.

SOLTIVE recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Feb 27, 2026Open, Classified

    Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.

    Z-1695-2026 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗

  2. Feb 27, 2026Open, Classified

    Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.

    Z-1694-2026 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗

  3. May 6, 2025Open, Classified

    Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

    Z-1923-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. May 6, 2025Open, Classified

    Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

    Z-1922-2025 · Root cause: Under Investigation by firm · names 7 of these models · Recall details · FDA record ↗

  5. May 6, 2025Open, Classified

    Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

    Z-1921-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  6. May 6, 2025Open, Classified

    Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

    Z-1920-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  7. Apr 4, 2025Open, Classified

    Preset treatment parameters are not consistently being used in accordance with the IFU.

    Z-1715-2025 · Root cause: Unknown/Undetermined by firm · names 4 of these models · Recall details · FDA record ↗

  8. Aug 13, 2024Open, Classified

    The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.

    Z-0135-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  9. Aug 13, 2024Open, Classified

    The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.

    Z-0134-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  10. May 31, 2024Open, Classified

    Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.

    Z-2506-2024 · Root cause: Device Design · Recall details · FDA record ↗

  11. May 31, 2024Open, Classified

    Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.

    Z-2505-2024 · Root cause: Device Design · Recall details · FDA record ↗

  12. May 31, 2024Open, Classified

    Difficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.

    Z-2504-2024 · Root cause: Device Design · Recall details · FDA record ↗

  13. Jan 29, 2024Open, Classified

    A non-compliant power cord may have been supplied in regions that do not utilize the Type B or Type E/F plug configurations.

    Z-1385-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  14. Jan 29, 2024Open, Classified

    A non-compliant power cord may have been supplied in regions that do not utilize the Type B or Type E/F plug configurations.

    Z-1384-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  15. Dec 4, 2023Open, Classified

    Affected products may potentially have a breached sterile pouch seal. A breached sterile pouch seal may lead to minor procedure delays (if noticed prior to use) or patient infections (if not noticed prior to use in patient).

    Z-0718-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  16. Dec 1, 2023Open, Classified

    Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.

    Z-0748-2024 · Root cause: Labeling Change Control · names 7 of these models · Recall details · FDA record ↗

  17. May 26, 2022Terminated

    Device model and size for affected lots of Laser Fibers may be mismatched between the product's carton, immediate pouch packaging, and/or the device.

    Z-1296-2022 · Root cause: Labeling mix-ups · names 4 of these models · Recall details · FDA record ↗

Most reported SOLTIVE problems

Device problems

  • Break201
  • Fire122
  • Communication or Transmission Problem88
  • Smoking67
  • Connection Problem57
  • Adverse Event Without Identified Device or Use Problem54
  • Sparking42
  • Wireless Communication Problem38

Patient problems

  • No Clinical Signs, Symptoms or Conditions776
  • Stenosis19
  • Unspecified Kidney or Urinary Problem18
  • Urethral Stenosis/Stricture14
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available12
  • Insufficient Information12
  • Renal Failure8
  • Burn(s)7

SOLTIVE models

16 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent SOLTIVE adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Gyrus ACMI, LLC products

All Gyrus ACMI, LLC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page