Skip to content

Drive PAN6300HBFB-P55658Recalls & reported problems

Drive Devilbiss Healthcare · Beds LTC Accessories Product Description Bariatric Foot Board with Staff control · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00822383008127), received between Jan 4, 2023 and Jan 4, 2023 — classified as 1 death.

The most frequently reported specific device problem is “Compatibility Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Compatibility Problem1
  • Device Handling Problem1

Patient problems

  • Physical Entrapment1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathJan 4, 2023Compatibility Problem; Device Handling Problem; Compatibility Problem; Device Handling Problem

    DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A BARIATRIC BED BY A DISTRIBUTOR, WHO STATED THAT "THE END USER HAD BEEN FOUND DECEASED WITH THEIR HEAD CAUGHT IN THE FRAME OF THE BED." DRIVE INVESTIGATED THE INCIDENT AND DETERMINED THAT THERE WERE NO BED RAILS OR HANDLES ON THE BED; THAT THE END USER'S HEAD WAS STUCK BETWEEN THE DECK AND THE BASE OF THE BED AT THE FOOT SECTION; THAT THE FACILITY WHERE THE END USER LIVED HAD ATTACHED A BED CONTROL PENDANT FROM A DIFFERENT MODEL BED; AND THAT SAID BED CONTROL PENDANT DID NOT OPERATE PROPERLY WITH THE BED IN QUESTION, WITH SEVERAL BUTTONS NOT OPERATING AT ALL. DRIVE ALSO LEARNED THAT THE PATIENT WAS NOT IN BED AT THE TIME OF THE INCIDENT, BUT WAS RATHER ON THE FLOOR, POTENTIALLY TRYING TO RETRIEVE AN ITEM THAT HAD FALLEN BEHIND THE BED, AND POTENTIALLY ATTEMPTED TO USE THE INCOMPATIBLE BED PENDANT TO MAKE IT EASIER FOR HIM TO REACH THE ITEM.

    Patient problems: Physical Entrapment

    MAUDE report 16096589 ↗

Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices

All BED, AC-POWERED ADJUSTABLE HOSPITAL devices →