Skip to content

Drive 10289BLRecalls & reported problems

Drive Devilbiss Healthcare · 3 Whl Alum Rollator,Blue,1/ea · Class 1

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00822383259345), received between Nov 2, 2017 and Mar 3, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Bent1
  • Collapse1

Patient problems

  • Bone Fracture(s)2
  • Damage to Ligament(s)1
  • Fall1
  • Laceration(s)1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 3, 2025Break; Break

    DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY AN ATTORNEY, WHO ALLEGED THAT THE CLIENT WAS INJURED WHEN HER ROLLATOR FAILED DURING USE RESULTING IN WOUNDS AND A BROKEN NOSE. AS PART OF ITS COMPLAINT REVIEW AND EVALUATION PROCESS, DRIVE DEVILBISS HEALTHCARE WILL ALSO NOTIFY THE MANUFACTURER OF THE PRODUCT ABOUT THIS COMPLAINT.

    Patient problems: Bone Fracture(s); Laceration(s)

    MAUDE report 21504893 ↗
  • InjuryJul 5, 2023Break; Break

    (B)(6) WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY AN END USER WHO STATED THAT "THE FRONT WHEEL BROKE OFF AND THE UNIT BROKE WHILE CROSSING THE STREET." THE END USER WAS TREATED AT AN URGENT CARE FACILITY, WHERE HER WRIST AND KNEE WERE BANDAGED. DRIVE REQUESTED THAT THE UNIT BE RETURNED FOR EVALUATION, AND WILL FILE AN UPDATE IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: Damage to Ligament(s)

    MAUDE report 17262488 ↗
  • InjuryNov 2, 2017Bent; Collapse

    DRIVE DEVILBISS HEALTHCARE IS THE INITIAL IMPORTER OF THE DEVICE. END-USER WAS USING DEVICE IN THE HOME IN AN AREA WHERE OUTDOOR CARPET AND TILES CONVERGED. AT THE BUMP IN THE FLOORING THE FRONT WHEEL COLLAPSED. SHE FELL AND FRACTURED HER FOOT. THE ITEM WAS RETURNED AND EVALUATED. ITEM RETURNED ON RMA (B)(4) HAD A COMPLAINT OF THE FRONT WHEEL COLLAPSING. THE UNIT SHOWS MODERATE USE WITH NO SIGNIFICANT DAMAGE TO THE FRAME OR WHEELS. THE BRAKES ON BOTH SIDES ARE CONFORMING. THE FRONT WHEEL HAS BENT FORWARD AND DOWNWARD BUT SHOWS NO DAMAGE TO THE WHEEL OR FRONT FORK. ROOT CAUSE MAY BE ATTRIBUTED TO MATERIAL DAMAGE DUE TO WELDING. THE NORMAL STRUCTURE MAY NOT HAVE BEEN STRONG ENOUGH DUE TO DAMAGED MATERIAL.

    Patient problems: Bone Fracture(s); Fall

    MAUDE report 6998165 ↗

Other WALKER, MECHANICAL devices

All WALKER, MECHANICAL devices →