Skip to content

DRIVEMEDICAL GV974RD-TRecalls & reported problems

Drive Devilbiss Healthcare · Rollators4-Wheel Alum Product Description: F22 Rollator, Tall, Red · Class 1

All models of DRIVEMEDICAL (111)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00822383582238), received between Oct 8, 2018 and Mar 25, 2026 — classified as 4 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Material Fragmentation2
  • Break1

Patient problems

  • Bone Fracture(s)2
  • Fall2
  • Hernia2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 25, 2026Fracture

    PER DISTRIBUTOR MDR #243477-2026-00015: DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY THE END USER WHO REPORTED THE WELD AT THE CROSSBAR BROKE CAUSING THE END USER TO FALL. THE END USER CAUGHT HIMSELF AND NOTICED A HERNIA THAT EVENING IN HIS GROIN AREA. THE END USER REQUIRED EMERGENCY SURGERY TO REPAIR THE HERNIA. AS PART OF ITS COMPLAINT REVIEW AND EVALUATION PROCESS, DRIVE DEVILBISS HEALTHCARE WILL ALSO NOTIFY THE MANUFACTURER OF THE PRODUCT ABOUT THIS COMPLAINT.

    Patient problems: Hernia

    MAUDE report 24687747 ↗
  • InjuryMar 19, 2026Fracture

    DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY THE END USER WHO REPORTED THE WELD AT THE CROSSBAR BROKE CAUSING THE END USER TO FALL. THE END USER CAUGHT HIMSELF AND NOTICED A HERNIA THAT EVENING IN HIS GROIN AREA. THE END USER REQUIRED EMERGENCY SURGERY TO REPAIR THE HERNIA. AS PART OF ITS COMPLAINT REVIEW AND EVALUATION PROCESS, DRIVE DEVILBISS HEALTHCARE WILL ALSO NOTIFY THE MANUFACTURER OF THE PRODUCT ABOUT THIS COMPLAINT.

    Patient problems: Hernia

    MAUDE report 24640574 ↗
  • InjuryAug 20, 2019Break; Material Fragmentation; Break; Material Fragmentation

    DRIVE DEVILBISS HEALTHCARE IS THE INITIAL IMPORTER OF THE DEVICE WHIHC IS A ROLLATOR. OUR PMS TEAM WAS ADVISED OF THIS EVENT BY AN ATTORNEY'S LETTER. WE DO NOT HAVE POSSESSION OF THE DEVICE NOR DO WE EXPECT TO IN THE NEAR FUTURE. WE ARE FILING THIS REPORT IN AN OVERABUNDANCE OF CAUTION. THE ENUDER REPORTEDLY SUFFERED AN INJURY ON (B)(6) 2019 WHILE USING THE DEVICE. THE SUPPORTING LEG FOR A FRONT WHEEL SHEARED OFF THE PRODUCT CAUSING HER TO FALL FORWARD, FRACTURING HER WRIST.

    Patient problems: Bone Fracture(s); Fall

    MAUDE report 8912129 ↗
  • InjuryOct 8, 2018Material Fragmentation; Material Fragmentation

    DRIVE DEVILBISS HEALTHCARE IS THE INITIAL IMPORTER OF THE DEVICE WHICH IS A ROLLATOR. PMS IS COORDINATING THE RETURN OF THE DEVICE AND WHEEL FOR EVALUATION. END-USER WAS WALKING WITH THE DEVICE WHEN THE RIGHT WHEEL FELL OFF. HE FELL AND WAS INJURED. HE REPORTEDLY INJURED HIS SHOULDER AND NECK.

    Patient problems: Bone Fracture(s); Fall

    MAUDE report 7945234 ↗

Other WALKER, MECHANICAL devices

All WALKER, MECHANICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page