Skip to content

DRIVEMEDICAL RTL10266DTRecalls & reported problems

Drive Devilbiss Healthcare · Rollators4-Wheel Alum Product Description: Nitro Duet, Red, 1 Ea. · Class 1

All models of DRIVEMEDICAL (111)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00822383972497), received between Sep 20, 2021 and Oct 21, 2023 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Material Integrity Problem1
  • Sharp Edges1

Patient problems

  • Fall3
  • Bone Fracture(s)2
  • Head Injury1
  • Insufficient Information1
  • Laceration(s)1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 21, 2023Break; Break

    DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY AN END USER'S DAUGHTER, WHO STATED THAT THE "RIGHT HANDLE SUDDENLY BENT DOWN WHERE THE PLASTIC HANDLE MEETS THE METAL FRAME" AND REPORTED THAT HER MOTHER LOST HER BALANCE AND BROKE HER LEG. THE END USER WAS TAKEN TO THE HOSPITAL AND UNDERWENT A WEEK OF TREATMENT FOR HER INJURIES. DRIVE REQUESTED THE UNIT BE RETURNED FOR INVESTIGATION AND WILL FILE AN UPDATE IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: Bone Fracture(s); Fall

    MAUDE report 17981882 ↗
  • InjuryAug 9, 2023Break; Break

    DRIVE (B)(4) WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY AN END USER WHO STATED THAT SHE WAS SITTING ON THE ROLLATOR AND BEING PUSHED THROUGH A THRESHOLD WHEN THE "[RIGHT] FRONT WHEEL BUCKLED AND THREW HER OUT OF THE CHAIR" AND SHE REPORTEDLY BROKE HER RIGHT ANKLE. THE END USER WAS TAKEN TO THE HOSPITAL TO TREAT HER BROKEN ANKLE. DRIVE REQUESTED THE UNIT BE RETURNED FOR INVESTIGATION AND WILL FILE AN UPDATE IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: Bone Fracture(s); Fall

    MAUDE report 17490546 ↗
  • MalfunctionJun 28, 2023Break; Break

    DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY A PROVIDER, WHO STATED THAT "THE FRAME SNAPPED WHILE USING THE ROLLATOR," ALTHOUGH NO SPECIFIC INFORMATION REGARDING THE MECHANISM OF FAILURE OR THE EXACT LOCATION OR NATURE OF THE BROKEN FRAME WAS PROVIDED. THE END USER WAS REPORTEDLY INJURED, BUT THE DETAILS OF THE INJURY WERE NOT PROVIDED. DRIVE IS CURRENTLY INVESTIGATING THE INCIDENT, INCLUDING ATTEMPTING TO RETRIEVE THE PRODUCT AND INSPECT IT, AND WILL FILE AN UPDATE IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 17216633 ↗
  • InjuryOct 4, 2022Material Integrity Problem; Material Integrity Problem

    DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY THE END USER'S DAUGHTER, WHO REPORTED THAT "ONE OF THE BOLTS ATTACHING THE BACK REST FELL OUT AND MY MOTHER FELL OFF THE BACK, LANDED ON HER BACK AND HIT THE BACK OF HER HEAD ON THE CONCRETE SIDEWALK. LATER WE ADMINISTERED PAIN RELIEF MEDICINE AND PAIN RUB ON HER SKIN OF HER LOWER BACK. SHE WILL BE GETTING AN X-RAY ON HER BACK." THE END USER DECLINED TO RETURN THE PRODUCT FOR EVALUATION.

    Patient problems: Fall; Head Injury

    MAUDE report 15536008 ↗
  • InjurySep 20, 2021Sharp Edges; Sharp Edges

    (B)(4) IS THE INITIAL IMPORTER OF THE DEVICE WHICH IS A ROLLATOR. THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. WE WILL SUBMIT A FOLLOW UP WHEN ADDITIONAL INFORMATION IS AVAILABLE. THE END-USER WALKED INTO THE MOTIONLESS DEVICE ON THE JETWAY. SHE GAUGED HER LEG ABOUT 6 INCHES LONG. AN AMBULANCE CAME TO TAKE HER TO THE ER. THE ER DOCTOR WAS UNABLE TO COMPLETELY SEW UP THE INJURY. HE GAVE HER 4 STITCHES AND INSTRUCTED HER NOT TO GET IT WET. SHE WAS SENT TO A WOUND CARE SPECIALIST WHO ADDED ADDITIONAL STITCHES TO THE WOUND. HE DEFINED THE INJURY AS AN AVULSION LACERATION. ANTIBIOTICS WERE PRESCRIBED TO PREVENT INFECTION.

    Patient problems: Laceration(s)

    MAUDE report 12495609 ↗

Other WALKER, MECHANICAL devices

All WALKER, MECHANICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page