Cook G00040Recalls & reported problems
Cook Incorporated · Connecting Tube Female to Male Luer Lock · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002000407), received between Nov 5, 2019 and Feb 2, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Split, Cut or Torn4
- Fluid/Blood Leak1
- Leak/Splash1
Patient problems
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 2, 2021Fluid/Blood Leak; Leak/Splash; Material Split, Cut or Torn; Fluid/Blood Leak; Leak/Splash; Material Split, Cut or Torn
AS REPORTED, DURING A PULMONARY ANGIOGRAM A MEDIUM PRESSURE CONNECTING TUBE BEGAN TO LEAK AFTER THE PRESSURE INJECTION CAUSED A HOLE IN THE SIDE OF THE DEVICE. THE MACHINE PRESSURE WAS SET AT 800 (NO UNITS PROVIDED). IT STATED THE PRESSURE DURING THE EVENT WAS 797. OMNIPAQUE WAS BEING INJECTED AT 30ML/S AND 20ML/S. THERE WAS NO CHANGE OR IMPACT ON THE PATIENT CARE OR OUTCOME. NO UNINTENDED PART OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY. NO ADDITIONAL PROCEDURE WAS REQUIRED DUE TO THIS OCCURRENCE. NO ADVERSE EFFECT WAS REPORTED DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11267485 ↗MalfunctionNov 5, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
ADDITIONAL INFORMATION WAS RECEIVED 10NOV2019. THE CUSTOMER REPORTED THAT THEIR INJECTOR WAS MALFUNCTIONING, AND THE ISSUE WAS NOT WITH THE ORIGINALLY REPORTED COOK DEVICE ASSOCIATED WITH THIS COMPLAINT. THE CUSTOMER HAS HAD THE UNKNOWN MANUFACTURER'S INJECTOR SERVICED AND THERE HAVE BEEN NO FURTHER PROBLEMS WITH THE TUBING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9280650 ↗MalfunctionNov 5, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9280543 ↗MalfunctionNov 5, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
ADDITIONAL INFORMATION WAS RECEIVED 10NOV2019. THE CUSTOMER REPORTED THAT THEIR INJECTOR WAS MALFUNCTIONING, AND THE ISSUE WAS NOT WITH THE ORIGINALLY REPORTED COOK DEVICE ASSOCIATED WITH THIS COMPLAINT. THE CUSTOMER HAS HAD THE UNKNOWN MANUFACTURER'S INJECTOR SERVICED AND THERE HAVE BEEN NO FURTHER PROBLEMS WITH THE TUBING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9279970 ↗
Other Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass devices
- Indeflator 1003327Abbott Vascular Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass2 recalls722 reports
- Copilot 1003331Abbott Vascular Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass189 reports
- DLP® 11001GMedtronic, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass2 recalls0 reports
- RHV 23242Abbott Vascular Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass66 reports
- DLP® 14000Medtronic, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass1 recall2 reports
- PhD™ 00884450746745Merit Medical Systems, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass1 recall2 reports
- PhD™ 00884450746714Merit Medical Systems, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass1 recall1 report
- NAMIC 64038301Medline Industries, Inc. · Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page